Evaluation of three sample preparation methods for the direct identification of bacteria in positive blood cultures by MALDI-TOF

Abstract

Background

Patient mortality is significantly reduced by rapid identification of bacteria from sterile sites. MALDI-TOF can identify bacteria directly from positive blood cultures and multiple sample preparation methods are available. We evaluated three sample preparation methods and two MALDI-TOF score cut-off values. Positive blood culture bottles with organisms present in Gram stains were prospectively analysed by MALDI-TOF. Three lysis reagents (Saponin, SDS, and SepsiTyper lysis bufer) were applied to each positive culture followed by centrifugation, washing and protein extraction steps. Methods were compared using the McNemar test and 16S rDNA sequencing was used to assess discordant results.

Results

In 144 monomicrobial cultures, using ≥2.000 as the cut-off value, species level identifications were obtained from 69/144 (48%) samples using Saponin, 86/144 (60%) using SDS, and 91/144 (63%) using SepsiTyper. The difference between SDS and SepsiTyper was not statistically significant (P = 0.228). Differences between Saponin and the other two reagents were significant (P < 0.01). Using ≥1.700 plus top three results matching as the cut-off value, species level identifications were obtained from 100/144 (69%) samples using Saponin, 103/144 (72%) using SDS, and 106/144 (74%) using SepsiTyper and there was no statistical difference between the methods. No true discordances between culture and direct MALDI-TOF identification were observed in monomicrobial cultures. In 32 polymicrobial cultures, MALDI-TOF identified one organism in 34–75% of samples depending on the method.

Conclusions

This study demonstrates two inexpensive in-house detergent lysis methods are non-inferior to a commercial kit for analysis of positive blood cultures by direct MALDI-TOF in a clinical diagnostic microbiology laboratory.

Background

Bacteraemia and sepsis have a high mortality rate and it is crucial to start appropriate antimicrobial treatment as promptly as possible [1, 2]. Conventional identification of organisms causing bacteraemia relies on Gram staining of a positive blood culture followed by solid media subculture for 18–24 h and then confirmatory tests to identify the causative organism. Whilst empirical antimicrobial treatment will have been started at the point of blood culture collection, earlier identification of bacteria in blood culture would enable the medical microbiologist to: more specifically target antimicrobial treatment (based on intrinsic and local patterns of antimicrobial susceptibilities of certain genera or species); instigate further relevant investigations; and inform infection prevention and control measures at an earlier stage which could potentially reduce mortality rates.

Matrix-Assisted Laser Desorption/Ionization Time-of-Flight (MALDI-TOF) is an inexpensive method that can identify a broad range of bacterial species directly from positive blood cultures [3]. Numerous methodologies for preparation of positive blood culture samples for MALDI-TOF have been described with a central aim of producing a bacterial pellet that is free from human blood cells, proteins, and culture medium. This can be achieved by detergent lysis of the human cells followed by centrifugation and washing steps of the bacterial pellet using a commercial kit or various other reagents (e.g. saponin and sodium dodecyl sulphate (SDS) solution) [37]. Other methods have utilised serum separator tubes [810], differential centrifugation or sedimentation [5, 1114] and ammonium chloride lysis [15]. A subsequent protein extraction step on the pellet is usually required to obtain high yields and high quality results.

In this study we assessed the performance of a commercial kit (Bruker SepsiTyper), two different detergent lysis reagents (5% saponin solution, 10% SDS solution) and two different MALDI-TOF score cut-off values for the identification of bacteria in positive blood culture bottles.

Methods

Study setting

The Public Health Laboratory, Birmingham, UK provides a diagnostic microbiology service to the Heart of England NHS Foundation Trust in Birmingham from whom it receives about 27,000 blood culture samples per year. All blood culture bottles are incubated and microbial growth is detected using the BacT/ALERT® system (bioMérieux). Blood culture media used are BacT/ALERT® SA Standard Aerobic, BacT/ALERT® SN Standard Anaerobic and BacT/ALERT® FA Plus (antimicrobial neutralization) media (bioMérieux). No charcoal containing bottles are used. When blood culture bottles flag positive on the system (indicating microbial growth), aliquots of the blood culture medium are used to prepare Gram stains and agar plate subcultures for identification of cultured organisms. The laboratory routinely uses MALDI-TOF (Bruker MALDI Biotyper system) for identification of bacterial isolates from agar plates. Some bacterial species are confirmed with biochemical tests in addition to MALDI-TOF according to ISO 15189:2012 United Kingdom Accreditation Service accredited methods used for routine clinical diagnostics. Routine clinical diagnostic results reported by the laboratory were considered the definitive result with which direct MALDI-TOF results (see below) were compared.

Study sample collection

On weekdays, over a four week period in December 2013, all blood cultures from patients ≥18 years old which flagged positive on the BacT/ALERT® (bioMérieux) system and had micro-organisms seen in the Gram stain were selected. All selected blood cultures were prepared for direct MALDI-TOF using three different lysis reagents in parallel (as described below) in addition to routine Gram stain and sub-culture methods.

Direct identification of micro-organisms in positive blood cultures by MALDI-TOF

Two hundred microliters of each of the three lysis reagents (5% Saponin solution, 10% SDS solution, Bruker SepsiTyper Lysis Buffer) were aliquoted into a new screw-capped microcentrifuge tubes and 1 ml of each positive blood culture was added to each tube. All tubes were vortex mixed for 10 s and then incubated at room temperature for 5 min and centrifuged for 1 min at 13,000g. Supernatant was removed and each pellet was re-suspended in either 1 ml MALDI grade water for Saponin or SDS or 1 ml SepsiTyper Washing Buffer and centrifuged for 1 min at 13,000g. Supernatant was again removed and each pellet re-suspended in 300 μl MALDI grade water and 900 μl 100% Ethanol followed by centrifugation for 2 min at 13,000g. The supernatant was then removed and each sample centrifuged again for 2 min at 13,000g. Residual ethanol was then removed and the pellets dried for 10 min at room temperature. Pellet size was then estimated. For small or non-visible pellets, 5 µl of 70% formic acid and 5 µl of 100% Acetonitrile was added. For medium-sized pellets, 20 µl of 70% formic acid and 20 µl of 100% Acetonitrile was added. For large pellets, 50 µl of 70% formic acid and 50 µl of 100% Acetonitrile was added. Pellets were re-suspended thoroughly then each sample was centrifuged for 2 min at 13,000g. One microliter of each supernatant was then spotted onto a sample spot of a MALDI target plate and allowed to dry. Two spots were prepared for each sample. One microliter of α-Cyano-4-hydroxycinnamic acid (HCCA) matrix solution (Bruker) was then added to each spotted sample and allowed to dry. Genus and species identification was then obtained using the Bruker MALDI Biotyper system.

MALDI-TOF data analysis

Spectra were automatically captured and analysed by the Bruker MALDI Biotyper system and database (version 2.0, IVD MALDI Biotyper database 4613). No manual acquisition was carried out. Two MALDI spots were analysed for each positive bottle and the highest scoring spot was selected for data analysis. Two different MALDI scoring methods were used: acceptable genus and species identification if a score ≥2.000 was obtained or acceptable genus and species identification if a score ≥1.700 and the top three matches were concordant. Samples giving scores below the cut off values evaluated were recorded as having no identification by direct MALDI-TOF.

16S rDNA PCR and capillary electrophoresis

Samples with discordant results between direct MALDI-TOF and the diagnostic laboratory’s reported identification from agar plates were analysed by 16S rDNA PCR and DNA sequence analysis. DNA sequences were queried against NCBI BLAST [16].

Data analysis

Within the study period, the first positive aerobic and anaerobic blood culture bottle from each patient was selected for data analysis. The McNemar test was used to evaluate if the differences observed between the three different lysis methods were statistically significant.

All MALDI-TOF testing was carried out without investigators seeing the Gram stain or conventional culture and identification results. At the conclusion of the study period, reported results of culture and identification by the clinical diagnostic laboratory for all positive blood cultures were downloaded from the laboratory information system for comparison with the direct MALDI-TOF results obtained.

Results

During the four week study period, a total of 199 blood culture bottles from 122 patients flagged positive and organisms were seen on the Gram stain. Of the 199 bottles, 10 were antimicrobial neutralization, 99 were standard aerobic and 90 were standard anaerobic media. Species identified from the individual bottles by direct MALDI-TOF are shown in Table 1. The first positive aerobic and/or anaerobic bottle(s) from each patient was included for further study giving 181 blood cultures (including 63 sets of two bottles) from 122 patients.

Table 1 Direct MALDI-TOF results from 199 blood culture bottles by lysis method and cut off value used

Of the 181 positive cultures, five failed to grow on sub-culture. Of the 15 direct MALDI-TOF results from these bottles, 14 gave no identification. In one bottle, Corynebacterium mucifaciens was identified with a score of 1.729 using SDS. Gram positive bacilli had been seen in the Gram stain.

Thirty-two of the 181 positive blood cultures showed polymicrobial growth (≥2 different species) on subculture. Two or more organisms were seen on the Gram stain in only 3/32 (9%) bottles with at least one organism seen in all 32 bottles. In 12/32 (38%) samples both organisms grown would have had the same morphology on Gram stain. Where direct MALDI-TOF identifications were obtained for polymicrobial culture bottles, MALDI-TOF only ever identified one organism from a bottle. Using Saponin, MALDI-TOF identified a single organism in 11/32 (34%) polymicrobial bottles using ≥2.000 and 20/32 (63%) bottles using ≥1.700 and the top three matches were concordant as a cut-off score respectively. Using SDS, a single organism was identified in 21/32 (66%) or 24/32 (75%) bottles respectively. Using SepsiTyper, a single organism was identified in 22/32 (69%) or 24/32 (75%) bottles respectively. Where obtained, the direct MALDI-TOF identifications were always concordant with the single organisms identified in the Gram stain and at least one of the organisms seen in Gram stains with two organisms seen.

One hundred and forty-four positive blood cultures from 100 patients had one organism identified from conventional subculture (monomicrobial). These were selected for further detailed analysis to identify the yield of identifications with direct MALDI-TOF and the concordance of these direct identifications with conventional identifications from sub-cultured organisms. Direct MALDI-TOF identifications using both cut off strategies for monomicrobial bottles were compared with organism identifications from subcultures. Each direct MALDI-TOF result was thus determined to be concordant with subculture, discordant with subculture or “No ID” where no reliable identification using the cut off was obtained. These results, broken down by bacterial groups and cut off score are shown in Table 2.

Table 2 Direct MALDI-TOF results in monomicrobial cultures

In the 144 monomicrobial cultures, using ≥2.000 as the cut-off value, the lowest yield of identification to species level was obtained with Saponin and the highest yield was obtained using SepsiTyper (Table 2). No discordances between the organisms identified by direct MALDI-TOF and the corresponding report from routine diagnostic culture were observed using ≥2.000 as the cut-off value.

The proportion of samples with identification to species level was increased by amending the MALDI-TOF cut-off value to ≥1.700 and top three results matching. However, two discordances from two samples were observed (Table 2). These discordances were further investigated; in the first sample Corynebacterium tuberculostearicum was identified by MALDI-TOF using the SepsiTyper method only (also confirmed using PCR and sequencing of the 16S gene from the blood culture) but the culture results had been reported as “Propionibacterium sp.”. Further examination of laboratory notes indicated that the identification of a Propionibacterium sp. was made on colony and Gram morphology only indicating that the reported culture result may have been imprecise and that there was no true discordance. In the second discrepant sample the three MALDI-TOF preparation methods and 16S PCR with sequencing all consistently detected Staphylococcus epidermidis in the aerobic bottle and Actinomyces oris in the anaerobic bottle but the culture results had been reported as “S. epidermidis”. Further examination of laboratory notes indicated that a faint growth of another organism had been noted on anaerobic plates but identification was not pursued as the isolate was not thought to be clinically significant, again indicating that the reported culture result was incomplete and that there was no true discordance.

The difference in yield of culture concordant identifications by direct MALDI-TOF using both cut-off strategies was evaluated with the McNemar test using a significance level of P < 0.05. None of the differences in yield between methods using ≥1.700 and top three results matching were statistically significantly different in this evaluation (Saponin vs. SDS P = 0.8445; Saponin vs. SepsiTyper P = 0.4414; SDS vs. SepsiTyper P = 0.5465). Using the ≥2.000 scoring system, the difference in yield between SDS and SepsiTyper was not statistically significant (P = 0.2278). Differences between Saponin versus SDS or Saponin versus SepsiTyper were significant (P = 0.0125 and P = 0.0015 respectively).

Discussion

This study is the first to perform a three way comparison of the SepsiTyper kit with two detergent lysis methods for preparing positive blood cultures for direct MALDI-TOF analysis prospectively in a clinical setting.

Using a cut off score of ≥2.000 (i.e. the manufacturer’s recommended cut off value for identification of organisms from culture plates) there was no statistical difference found between the rates of obtaining sub-culture concordant species level identifications using 10% SDS or SepsiTyper to prepare monomicrobial positive blood cultures for direct MALDI-TOF. Rates of identification were 60% and 63% respectively. The use of 5% Saponin with a cut off score of ≥2.000 was found to give significantly fewer reliable species level MALDI-TOF identifications from direct blood cultures that either of the other methodologies.

Using a cut off score of ≥1.700 and top three results matching, there was no statistical difference found between the rates of obtaining sub-culture concordant species level identifications using 10% SDS, 5% saponin or SepsiTyper to prepare monomicrobial positive blood cultures for direct MALDI-TOF. The rate of identification for all three methods ranged from 69 to 74% using this lower cut off. With all three lysis methods, the lower score cut off value of ≥1.700 and top three results matching increased the proportion of species level identifications by at least 12% more when compared to the more stringent ≥2.000 cut off value but did however yield two apparent discordances between MALDI-TOF and conventional culture. Both apparent discordances were subsequently found to be due to incomplete identification of cultured organisms and the results obtained by MALDI-TOF were correct and corroborated by 16S PCR and sequencing.

We have demonstrated that up to 72% of bacteria in monomicrobial blood cultures can be identified on day one by MALDI-TOF. However polymicrobial cultures remain problematic for MALDI-TOF directly from blood culture as previously reported [11, 17]. In this study, where a direct identification was obtained from a culture that was polymicrobial on sub-culture, the identification was always consistent with one of the cultured organisms. This suggests that the direct identifications from polymicrobial cultures, when obtained with the methods described, can be considered accurate for the organism identified (despite not giving identification of all organisms) when making empirical clinical decisions, whilst awaiting technical confirmation. In common with several other studies [5, 6] we found a higher proportion of identification by MALDI-TOF to species level for blood cultures containing Gram negative bacteria than Gram positive bacteria and reduced rates of successful identification for streptococci.

The performance of the SepsiTyper kit has been evaluated in at least 21 reported studies as recently reviewed by Morgenthaler and Kostrzewa [6]. The reported percentage identification to species level in monomicrobial cultures is variable across the studies from 56 to 100% and probably reflects the different methodologies and gold standards with which it was compared as well as possible differences in the proportions of different bacteria isolated from bacteraemic patients at different sites. When compared with alternative methods, SepsiTyper has been found to be superior to gel separator tubes [18], differential centrifugation [5, 13, 19] and another commercial preparation kit [19]. In agreement with this study, SepsisTyper has previously been found to comparable with methods using lysis with saponin [4, 20] and SDS [5]. There are no previous reports comparing both Saponin and SDS.

For a clinical diagnostic laboratory, it is attractive to select the convenience of a commercially available, validated kit. However, the performance of any methodology will always need to be locally verified and operating costs are an increasingly important consideration. At the time of this evaluation, the cost per sample of the ready-made SDS or saponin reagent was 100 times lower than the cost of the Sepsityper kit with identical consumption of other consumables and staff time.

Identification of bacteria in the clinical diagnostic laboratory by MALDI-TOF is still a developing technology. As MALDI-TOF is used increasingly as a direct identification method from blood cultures the databases and algorithms that support the identifications are likely to improve. During this study, the investigators interpreted the MALDI-TOF results in isolation from the Gram results and any clinical information. If this method were implemented, the results would be interpreted in the light of laboratory expertise, the Gram stain and clinical information which have been shown to improve correct identifications and minimise misidentifications [5, 7, 21, 22]. Results from this study also suggest that the cut off score value to be selected for clinical use needs to balance the benefits of a greater number of identifications on day one against the risks of possible misidentification, especially in the case of polymicrobial cultures. A number of studies have demonstrated clinical benefits from earlier direct identification of organisms in blood cultures by using MALDI-TOF [17, 23, 24]. One of our aims for the development of direct MALDI-TOF on positive cultures within our laboratory is to assess the potential clinical impact on patient care.

Conclusions

This study has demonstrated that two relatively inexpensive in-house detergent lysis methods are non-inferior when compared with a commercially available kit for preparing positive blood cultures for direct MALDI-TOF in a large clinical diagnostic microbiology laboratory. Using a MALDI-TOF score cut off value of ≥1.700 and top three results matching for species level identification gave reliable organism identifications from 72% of monomicrobial blood cultures and identified one of the organisms from polymicrobial cultures in 34–75% of samples depending on the method. No organisms were identified by MALDI-TOF in any samples that were truly discordant with the culture results.

Abbreviations

MALDI-TOF:

matrix-assisted laser desorption/ionization time-of-flight mass spectrometry

SDS:

sodium dodecyl sulfate

HCCA:

α-cyano-4-hydroxycinnamic acid

PHE:

Public Health England

References

  1. Dellinger RP, Levy MM, Rhodes A, Annane D, Gerlach H, Opal SM, Sevransky JE, Sprung CL, Douglas IS, Jaeschke R, Osborn TM, Nunnally ME, Townsend SR, Reinhart K, Kleinpell RM, Angus DC, Deutschman CS, Machado FR, Rubenfeld GD, Webb SA, Beale RJ, Vincent JL, Moreno R. Surviving sepsis campaign: international guidelines for management of severe sepsis and septic shock: 2012. Crit Care Med. 2013;41:580–637.

    Article  PubMed  Google Scholar 

  2. Seifert H. The clinical importance of microbiological findings in the diagnosis and management of bloodstream infections. Clin Infect Dis. 2009;48(Suppl 4):S238–45. doi:10.1086/598188.:S238-S245.

    Article  PubMed  Google Scholar 

  3. Kok J, Thomas LC, Olma T, Chen SC, Iredell JR. Identification of bacteria in blood culture broths using matrix-assisted laser desorption-ionization Sepsityper and time of flight mass spectrometry. PLoS ONE. 2011;6:e23285.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  4. Meex C, Neuville F, Descy J, Huynen P, Hayette MP, De MP, Melin P. Direct identification of bacteria from BacT/ALERT anaerobic positive blood cultures by MALDI-TOF MS: MALDI Sepsityper kit versus an in-house saponin method for bacterial extraction. J Med Microbiol. 2012;61:1511–6.

    Article  CAS  PubMed  Google Scholar 

  5. Saffert RT, Cunningham SA, Mandrekar J, Patel R. Comparison of three preparatory methods for detection of bacteremia by MALDI-TOF mass spectrometry. Diagn Microbiol Infect Dis. 2012;73:21–6.

    Article  CAS  PubMed  Google Scholar 

  6. Morgenthaler NG, Kostrzewa M. Rapid identification of pathogens in positive blood culture of patients with sepsis: review and meta-analysis of the performance of the sepsityper kit. Int J Microbiol. 2015;. doi:10.1155/2015/827416.

    PubMed  PubMed Central  Google Scholar 

  7. Ferroni A, Suarez S, Beretti JL, Dauphin B, Bille E, Meyer J, Bougnoux ME, Alanio A, Berche P, Nassif X. Real-time identification of bacteria and Candida species in positive blood culture broths by matrix-assisted laser desorption ionization-time of flight mass spectrometry. J Clin Microbiol. 2010;48:1542–8.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  8. Wimmer JL, Long SW, Cernoch P, Land GA, Davis JR, Musser JM, Olsen RJ. Strategy for rapid identification and antibiotic susceptibility testing of gram-negative bacteria directly recovered from positive blood cultures using the Bruker MALDI Biotyper and the BD Phoenix system. J Clin Microbiol. 2012;50:2452–4.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  9. Moussaoui W, Jaulhac B, Hoffmann AM, Ludes B, Kostrzewa M, Riegel P, Prevost G. Matrix-assisted laser desorption ionization time-of-flight mass spectrometry identifies 90% of bacteria directly from blood culture vials. Clin Microbiol Infect. 2010;16:1631–8.

    Article  CAS  PubMed  Google Scholar 

  10. Stevenson LG, Drake SK, Murray PR. Rapid identification of bacteria in positive blood culture broths by matrix-assisted laser desorption ionization-time of flight mass spectrometry. J Clin Microbiol. 2010;48:444–7.

    Article  CAS  PubMed  Google Scholar 

  11. La Scola B, Raoult D. Direct identification of bacteria in positive blood culture bottles by matrix-assisted laser desorption ionisation time-of-flight mass spectrometry. PLoS ONE. 2009;4:e8041.

    Article  PubMed  PubMed Central  Google Scholar 

  12. Christner M, Rohde H, Wolters M, Sobottka I, Wegscheider K, Aepfelbacher M. Rapid identification of bacteria from positive blood culture bottles by use of matrix-assisted laser desorption-ionization time of flight mass spectrometry fingerprinting. J Clin Microbiol. 2010;48:1584–91.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  13. Juiz PM, Almela M, Melcion C, Campo I, Esteban C, Pitart C, Marco F, Vila J. A comparative study of two different methods of sample preparation for positive blood cultures for the rapid identification of bacteria using MALDI-TOF MS. Eur J Clin Microbiol Infect Dis. 2012;31:1353–8.

    Article  CAS  PubMed  Google Scholar 

  14. Ferreira L, Sanchez-Juanes F, Porras-Guerra I, Garcia-Garcia MI, Garcia-Sanchez JE, Gonzalez-Buitrago JM, Munoz-Bellido JL. Microorganisms direct identification from blood culture by matrix-assisted laser desorption/ionization time-of-flight mass spectrometry. Clin Microbiol Infect. 2011;17:546–51.

    Article  CAS  PubMed  Google Scholar 

  15. Prod’hom G, Bizzini A, Durussel C, Bille J, Greub G. Matrix-assisted laser desorption ionization-time of flight mass spectrometry for direct bacterial identification from positive blood culture pellets. J Clin Microbiol. 2010;48:1481–3.

    Article  PubMed  PubMed Central  Google Scholar 

  16. Altschul SF, Gish W, Miller W, Myers EW, Lipman DJ. Basic local alignment search tool. J Mol Biol. 1990;215:403–10.

    Article  CAS  PubMed  Google Scholar 

  17. Clerc O, Prod’hom G, Vogne C, Bizzini A, Calandra T, Greub G. Impact of matrix-assisted laser desorption ionization time-of-flight mass spectrometry on the clinical management of patients with Gram-negative bacteremia: a prospective observational study. Clin Infect Dis. 2013;56:1101–7.

    Article  CAS  PubMed  Google Scholar 

  18. Klein S, Zimmermann S, Kohler C, Mischnik A, Alle W, Bode KA. Integration of matrix-assisted laser desorption/ionization time-of-flight mass spectrometry in blood culture diagnostics: a fast and effective approach. J Med Microbiol. 2012;61:323–31.

    Article  CAS  PubMed  Google Scholar 

  19. Loonen AJ, Jansz AR, Stalpers J, Wolffs PF, van den Brule AJ. An evaluation of three processing methods and the effect of reduced culture times for faster direct identification of pathogens from BacT/ALERT blood cultures by MALDI-TOF MS. Eur J Clin Microbiol Infect Dis. 2012;31:1575–83.

    Article  CAS  PubMed  Google Scholar 

  20. Martiny D, Dediste A, Vandenberg O. Comparison of an in-house method and the commercial Sepsityper kit for bacterial identification directly from positive blood culture broths by matrix-assisted laser desorption-ionisation time-of-flight mass spectrometry. Eur J Clin Microbiol Infect Dis. 2012;31:2269–81.

    Article  CAS  PubMed  Google Scholar 

  21. Nonnemann B, Tvede M, Bjarnsholt T. Identification of pathogenic microorganisms directly from positive blood vials by matrix-assisted laser desorption/ionization time of flight mass spectrometry. APMIS. 2013;121:871–7.

    Article  CAS  PubMed  Google Scholar 

  22. Schieffer KM, Tan KE, Stamper PD, Somogyi A, Andrea SB, Wakefield T, Romagnoli M, Chapin KC, Wolk DM, Carroll KC. Multicenter evaluation of the Sepsityper extraction kit and MALDI-TOF MS for direct identification of positive blood culture isolates using the BD BACTEC FX and VersaTREK® diagnostic blood culture systems. J Appl Microbiol. 2014;116:934–41.

    Article  CAS  PubMed  Google Scholar 

  23. Perez KK, Olsen RJ, Musick WL, Cernoch PL, Davis JR, Land GA, Peterson LE, Musser JM. Integrating rapid pathogen identification and antimicrobial stewardship significantly decreases hospital costs. Arch Pathol Lab Med. 2013;137:1247–54.

    Article  PubMed  Google Scholar 

  24. Vlek AL, Bonten MJ, Boel CH. Direct matrix-assisted laser desorption ionization time-of-flight mass spectrometry improves appropriateness of antibiotic treatment of bacteremia. PLoS ONE. 2012;7:e32589.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

Download references

Authors’ contributions

HT: Conducted experiments, collected data and prepared manuscript. JE: Study design, conducted experiments, collected data, analysed data, revised manuscript. AH: Study conception and design and revised manuscript. SG: Study conception and design and revised manuscript. All authors read and approved the final manuscript.

Acknowledgements

The authors would like to thank all staff in the PHE Public Health Laboratory Birmingham who helped in this evaluation.

Competing interests

The authors declare that they have no competing interests.

Availability of data and material

The datasets used and/or analysed during the current study available from the corresponding author on reasonable request.

Ethics approval and consent to participate

This study was an evaluation of different analytical methods to establish the most effective method of disease diagnosis. This work was undertaken by the patient’s care team, and looked only at bacteria in blood samples. It did not involve the analysis or study of relevant material as defined by the Human Tissue Authority and so ethical approval was not required. This is in accordance with the revised guidance in the Governance Arrangements for Research Ethics Committees (GAfREC) that was released in September 2011. This project was subject to an internal review at PHE to ensure that it was fully compliant the NHS Research Governance Framework for Health and Social Care (April 2005), and with all other current regulatory requirements.

Funding

This study was internally funded by the PHE Public Health Laboratory Birmingham. The study was carried out as part of normal employment duties of all authors who were all employed by the PHE Public Health Laboratory Birmingham during design of the study and collection, analysis, and interpretation of data and writing the manuscript.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Hannah Tanner.

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Tanner, H., Evans, J.T., Gossain, S. et al. Evaluation of three sample preparation methods for the direct identification of bacteria in positive blood cultures by MALDI-TOF. BMC Res Notes 10, 48 (2017). https://doi.org/10.1186/s13104-016-2366-y

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s13104-016-2366-y

Keywords