Skip to main content

Feasibility of a prototype newborn resuscitation monitor to study transition at birth, measuring heart rate and ventilator parameters, an animal experimental study

Abstract

Background

Every year, an estimated 10 million babies are born, non-breathing and in need of resuscitation. Advances in management have been made over the past decades, however, approximately 700.000 yearly deaths result from this global problem. A prototype newborn resuscitation monitor (NRM) (Laerdal Global Health, Stavanger, Norway) has been developed with the purpose of studying newborn resuscitation. The monitor has the ability to continuously display HR using dry electrode ECG technology, to measure tidal volume, pressure and end tidal CO2, and to store the results for later analysis. Such monitor could enhance the care providers performance, and hence survival of neonates, by displaying the quality and response of the given care. The aim of this preclinical study was to describe the abilities of the NRM to measure ventilation and heart rate parameters against pathophysiological responses to different induced conditions in a piglet i.e. increased deadspace, pressure and washout of surfactant.

Methods

Piglets were chosen for the study, as they have tidal volumes of approximately 6 ml/kg, resembling the human neonate. Five piglets were anesthetized and intubated before starting positive pressure ventilation (PPV). The dry electrode ECG sensor of the NRM was placed over the abdomen, and experiments performed: (1) inducing different ventilation scenarios and (2) lavage of surfactant.

Results

The NRM was capable of continuously displaying HR and detecting inflicted changes in ventilation and compliance of piglets. It could measure inflated and exhaled volume, the pressure of the ventilations and also the end tidal CO2.

Conclusions

The NRM provides objective feedback in anesthetized animals, and may be used in clinical studies and hopefully generate new knowledge on neonatal transition and resuscitation. The monitor may be further developed for use in both low and high-resource settings.

Background

Every year, an estimated 10 million babies are born non-breathing and in need of resuscitation [1]. Advances in management of newborn infants requiring delivery room resuscitation have been made over the past decades, however, approximately 700.000 yearly deaths result from this global problem [2]. In 2010, the International Liaison Committee on Resuscitation (ILCOR) raised the important question of whether improved monitoring of the newborn would be of benefit in facilitating effective resuscitation including positive pressure ventilation [2]. During positive pressure ventilation, pressure and volume may individually play important roles in the establishment of effective or harmful ventilations. Experimental data indicate that the ventilation volume does not correlate well with the peak ventilation pressure, and that the provider cannot correctly estimate the volume given when only pressure is displayed as feedback [3]. Moreover, a recent study has shown the value of end tidal carbon dioxide (EtCO2) as a useful measure of lung aeration and pulmonary blood flow [4].

Assessment of the heart rate (HR) is essential during neonatal resuscitation, as an indicator of effective ventilation and immediate feedback to the provider [5]. However, previous studies have shown that common methods for assessing HR are inaccurate, such as palpation of the umbilical cord and auscultation of the chest [6]. ECG seems to be the most reliable and accurate technology for immediate monitoring of the newborn HR and continuous feedback during resuscitation [7, 8]. Nevertheless, this technology is not commonly available in the delivery rooms. Monitors to guide ventilations are rarely available. A prototype newborn resuscitation monitor (NRM) (Laerdal Global Health, Stavanger, Norway) has been developed as a research tool with the purpose to study delivery room resuscitation and explore possibilities for monitoring resuscitation. However, there was a need to demonstrate the abilities of the monitor on live subjects in a controlled setting prior to the clinical studies on newborns. The NRM was constructed to record ventilation pressure and volume, hence compliance, as well as EtCO2 and HR during varying ventilation rates, inflating volumes and dead space. The monitor complies with current medical device technical standards, is CE-marked and approved for patient use.

Piglets were chosen for this purpose, partly because their required tidal volume approximates 6 ml/kg, resembling the human neonate [9].

The objective of this preclinical study was to demonstrate the functions of the NRM to record physiological changes in live subjects through exploring how (1) EtCO2 measurements were affected by changes in ventilation rate, inflated volume, and different dead space (simulating a face mask ventilation situation) and (2) lavage of surfactant affected the compliance and HR (simulating different clinical conditions in an infant).

Methods

The study was performed at SEARCH (Sandnes Educational and Research Centre Høyland) animal laboratory in Sandnes, Norway. The protocol was approved by The Research Animal Council for the Rogaland Region. We included five piglets with postnatal age 1.5 (n = 1) and 2.5 weeks (n = 4), with median weight 5.1 kg (range 3.7–5.2 kg). As the monitor was CE marked and approved for use in humans, we chose a small sample size to test the functions and the signal data management before the clinical studies. The piglets were handled in accordance with the European Guidelines for Use of Experimental Animals by certified Federation of European Laboratory Animal Science Association category C researchers (FELASA C).

Measurements

The NRM has sensors capable of synchronously recording HR, airway pressure, flow, volume, and EtCO2. Flow and volume are measured using a flow sensor (MIM Gmbh, Krugzell, Germany) placed between the endotracheal tube (ETT) and the resuscitator bag. The flow sensor and its electronics are the same as used in the Florian respiration monitor (Acutronic AG, Zürich, Switzerland) that has previously been used in similar studies [10, 11]. The sensor has negligible flow resistance and dead space (1 ml), and measures airflow using hot wire anemometer technology. In addition, an airway adapter connecting two plastic tubes to the main cabinet is placed between the ETT and the bag. The first tube draws a sample of exhaled air (50 ml/min) for side stream CO2 measurement (ISA CO2 sensor, Masimo/PhaseIn AB,Danderyd, Sweden) (Fig. 1). The second tube measures airway pressure using a piezoresistive pressure sensor (MPXV5010, Freescale Semiconductor Inc, Austin, Tx).

Fig. 1
figure 1

Recordings of ventilation and ECG signal data: examples of manual baseline ventilations and ECG from piglet B

HR is calculated from ECG measured by dry electrodes: Two stainless steel discs are mounted on each side of a flexible arch that is placed gently over thorax or abdomen of the subject (Fig. 2). The pressure exerted by the flexible arch is comparable to that of a stethoscope used for auscultation. The ECG is recorded and HR calculated using an proprietary algorithm based on a zero crossing count algorithm [12 ]. The dry-electrode ECG technology and HR algorithm was validated (by the manufacturer) on adult human subjects (n = 30, 396 data points) against HR from a Philips HeartStart MRx (Philips Healthcare, Andover, MA) and showed a rate equivalence within 1 bpm in 84.3% of the data points, 99.8% within 4 bpm. The HR sensor is connected to the main cabinet through a cable, and the HR is displayed on the monitor.

Fig. 2
figure 2

The newborn resuscitation monitor

The monitor thus uses reliable and proven technology put together into one unit. Safety for patient and operator is ensured by complying with current medical device technical standards, and is tested and approved for patient use according to the EU Medical Device Directive.

During the experiment, the HR sensor was placed over the abdomen of the piglet, and a Laerdal Neonatal Resuscitator size 1 (240 ml) was used for manual ventilations. Data from all sensors were stored in an internal memory card, downloaded to a computer through an USB connection, and analyzed using Matlab (MathWorks, Natick, MA).

Protocol

The piglets were anesthetized in accordance with Norwegian guidelines for animal experiments. An ETT with cuff was inserted into the trachea and secured, and the HR sensor was placed over the abdomen. Anesthetic maintenance was provided. No interventions took place before the piglets were fully anesthetized. An intra-arterial line was established in one femoral artery to enable measurements of arterial pressures. The instrumented resuscitator (Fig. 2) was connected to the ETT, and baseline patterns and measurements of the inspiratory and expiratory pressure, volume, flow and EtCO2 at rest were recorded in the NRM. Each piglet underwent several experiments, and was euthanized with pentobarbital when mean arterial pressure decreased to ≤15 mbar.

Three experiments are described in this report; (1) measurements of EtCO2 with various tidal volumes, ventilation rates and dead space, and (2) blood was drawn from the arterial catheter for blood gas analysis and comparison of PaCO2 with EtCO2 every 4th minute. EtCO2 was recorded continuously by the monitor. (3) Lavage of surfactant. See Table 1 for description of the timeline for each piglet.

Table 1 Overview: overview of experiments and ventilations performed on each piglet

Ventilations in room air were performed with different tidal volumes and ventilation rates (Table 2). Different dead space (0–30 ml) were provided and controlled by an expandable syringe connected between the tube and airway sensors. Each combination of volume and rate were kept for 1 min, whereas each dead space series lasted 4 min. EtCO2 was recorded continuously by the monitor.

Table 2 Experiment with different dead space and volumes: levels of EtCO2 for different tidal volumes and dead space, and then levels of EtCO2 with different rates of inflation per minute (ipm) and dead space

During the surfactant deficiency experiment, the piglets were ventilated using 100% Oxygen, 40 ventilations/min and a volume of 10 ml/kg. We measured dynamic compliance [total expired volume/(peak inspiratory pressure − positive end expiratory pressure)]. Surfactant was lavaged from the piglet’s lungs with 10 ml saline ten times in each series (=100 ml in one series). Between each series of lavage, the piglets were ventilated for 3 min to allow steady state to occur (i.e. HR, ventilation pressure, and volume stabilizing at a certain level). The HR was registered continuously during the whole experiment. The series were repeated until the mean arterial pressure was ≤15 mbar.

Data analysis

Difference between groups were analyzed by Wilkoxon signed rank test and results given as median (range). Changes in compliance during surfactant lavage were analyzed using a linear mixed model. A two sided p value of <0.05 was considered to be statistically significant. The levels of EtCO2 and PaCO2 and the difference in compliance was compared between groups by Wilkoxon signed rank test. A p value of <0.05 was considered to be statistically significant. Data were analyzed using the IBM-SPSS statistical package (IBM SPSS Statistics for Windows, Version 22.0. Armonk, NY: IBM Corp), and figures were made using Matlab R2014a (MathWorks, Natick, MA).

Results

Data can be viewed in detail in availability of data and materials.

One piglet was excluded in both experiments due to excessive leakage around the endotracheal tube (ETT). The ventilation and HR signals were stored in the monitor correctly aligned in time, and data from each ventilation could be analyzed individually (Fig. 1).

EtCO2 readings with different dead space, volumes and ventilation rates

Different clinical scenarios were simulated and Table 2 shows the levels of EtCO2 during different tidal volumes (inflated volumes) (6–15 ml/kg), dead space (different size of mask/tube in airway) (0 and 20 ml) and ventilation rates (the rate of inflations given) (20 and 40 ventilations per minute). There was a tendency of lower EtCO2 levels with increasing dead space, i.e. 0 ml versus 20 ml, for the two smallest tidal volumes, i.e. 6 and 10 ml/kg (p = 0.066 and p = 0.068, respectively). Ventilation rates of 20 compared to 40 ventilations per minute did not influence the levels of EtCO2 with 0 and 20 ml dead space.

EtCO2 and PaCO2 were comparable and stable with dead spaces of 10 and 20 ml However, at 30 ml dead space there was a tendency for lower EtCO2 compared to PaCO2 (Table 3).

Table 3 Dynamic compliance and HR before and after lavage of surfactant from the lungs

Surfactant lavage

The dynamic compliance was measured in each piglet before the lavage of surfactant. The dynamic compliance and HR before and after the lavage series for each piglet are shown in Table 4. There was an average decrease in compliance by 3% after each lavage series (p = 0.004) (Fig. 3, panel a). The piglets tolerated 3–13 series of lavage before the mean arterial pressure decreased to ≤15 mbar. Piglet E with the lowest body weight tolerated three lavage series until mean arterial pressure was ≤15 mbar. ECG was continuously recorded and successfully stored during all the experiments. The HR remained stable throughout the lavage (Fig. 3, panel b).

Table 4 Levels of EtCO2 and PaCO2 measured every 4 min with different dead space
Fig. 3
figure 3

Lavage of surfactant: measured HR and dynamic compliance over time for piglets B (green line), C (turquoise line), D (blue line) and E (red line) during lavage of surfactant with 10 × 10 ml of saline. a shows the dynamic compliance measured at the corresponding times. b shows HR measured after each series of 10 × 10 ml saline lavage

Discussion

The monitor has the ability to continuously display HR using dry electrode ECG technology, to measure tidal volume, pressure and EtCO2 and to store the results for later analysis. The planned clinical studies will hopefully provide new knowledge on neonatal transition and resuscitation; about ventilation pressure and volume and thereby compliance, as well as mask leakage, EtCO2, and HR during face mask ventilation. The NRM will be used to study if measurements and feedback to the provider will facilitate more objective assessments and thus improve decision making and quality of care during newborn resuscitation.

Accuracy level of volumes and pressures registered during the controlled ventilations was in accordance with our expectations. We further observed a steady and significant decrease in compliance when surfactant was lavaged from the piglets lungs. The continuous fall in compliance of mean 3% for each lavage, underlines the monitors ability to detect minute changes in volume and pressure over time. This observed ability of the NRM to register physiological signal data is reassuring.

Previous findings suggest that EtCO2 levels can indicate lung aeration, but is not an accurate measure of PaCO2 [4]. Our findings were in line with these previous findings although we had a small number of measurements. It seemed EtCO2 was varying according to dilution of expired gas in terms of increased dead space and change in tidal volumes. This, as well as the delays in side stream technology, is important to consider when moving to clinical studies, and suggests EtCO2 may only be useful as an indication of presence or change in aeration in the setting of bag/mask ventilation.

The dry electrode ECG sensor detected satisfactory ECG signals throughout the study. In the clinical setting, the skin of newborns is likely to be moister which would further increase the dry electrodes’ capability of detecting ECG signals. Importantly, HR is the recommended indicator of successful ventilations [13], and the dry electrode ECG technique can provide valuable real-time feedback to the provider. Validation to conventional ECG was satisfactory. However, it was performed on adults and it precision and accuracy in newborns is yet to be studied.

Conclusions

The prototype NRM was capable of continuously displaying HR and detecting inflicted changes in ventilation and compliance of piglets. It can record parameters such as given volume and pressure and the responses (changes in HR and EtCO2). This monitor may facilitate research during neonatal transition and resuscitation that can provide valuable new insight.

Abbreviations

NRM:

newborn resuscitation monitor

HR:

heart rate

EtCO2 :

end tidal carbon dioxide

PPV:

positive pressure ventilation

ETT:

endotracheal tube

mbar:

millibar

References

  1. Lawn JE, Blencowe H, Oza S, You D, Lee AC, Waiswa P, Lalli M, Bhutta Z, Barros AJ, Christian P, Mathers C, Cousens SN, Lancet Every Newborn Study G. Every newborn: progress, priorities, and potential beyond survival. Lancet. 2014;384:189–205.

    Article  PubMed  Google Scholar 

  2. Perlman J, Kattwinkel J, Wyllie J, Guinsburg R, Velaphi S, Nalini Singhal for the Neonatal ITFG. Neonatal resuscitation: in pursuit of evidence gaps in knowledge. Resuscitation. 2012;83:545–50.

    Article  PubMed  Google Scholar 

  3. Kattwinkel J, Stewart C, Walsh B, Gurka M, Paget-Brown A. Responding to compliance changes in a lung model during manual ventilation: perhaps volume, rather than pressure, should be displayed. Pediatrics. 2009;123:e465–70.

    Article  PubMed  Google Scholar 

  4. Hooper SB, Fouras A, Siew ML, Wallace MJ, Kitchen MJ, te Pas AB, Klingenberg C, Lewis RA, Davis PG, Morley CJ, Schmolzer GM. Expired CO2 levels indicate degree of lung aeration at birth. PLoS ONE. 2013;8:e70895.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  5. Dawes G. Birth asphyxia, resuscitation and brain damage. Foetal and neonatal physiology. Year Book; 1968. p. 141–159.

  6. Kamlin CO, O’Donnell CP, Everest NJ, Davis PG, Morley CJ. Accuracy of clinical assessment of infant heart rate in the delivery room. Resuscitation. 2006;71:319–21.

    Article  PubMed  Google Scholar 

  7. van Vonderen JJ, Hooper SB, Kroese JK, Roest AA, Narayen IC, van Zwet EW, Te Pas AB. Pulse oximetry measures a lower heart rate at birth compared with electrocardiography. J Pediatr. 2015;166:49–53.

    Article  PubMed  Google Scholar 

  8. Wyckoff MH, Aziz K, Escobedo MB, Kapadia VS, Kattwinkel J, Perlman JM, Simon WM, Weiner GM, Zaichkin JG. Part 13: neonatal resuscitation: 2015 american heart association guidelines update for cardiopulmonary resuscitation and emergency cardiovascular care. Circulation. 2015;132:S543–60.

    Article  PubMed  Google Scholar 

  9. te Pas AB, Wong C, Kamlin CO, Dawson JA, Morley CJ, Davis PG. Breathing patterns in preterm and term infants immediately after birth. Pediatr Res. 2009;65:352–6.

    Article  Google Scholar 

  10. McCallion N, Davis PG, Morley CJ. Volume-targeted versus pressure-limited ventilation in the neonate. Cochrane Database Syst Rev. 2005;11:CD003666.

    Google Scholar 

  11. McCallion N, Lau R, Dargaville PA, Morley CJ. Volume guarantee ventilation, interrupted expiration, and expiratory braking. Arch Dis Child. 2005;90:865–70.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  12. Köhler BUHC, Orglmeister R. Qrs detection using zero crossing counts. Prog Biomed Res. 2003;2003(8):138–45.

    Google Scholar 

  13. Perlman JM, Wyllie J, Kattwinkel J, Atkins DL, Chameides L, Goldsmith JP, Guinsburg R, Hazinski MF, Morley C, Richmond S, Simon WM, Singhal N, Szyld E, Tamura M, Velaphi S. Neonatal resuscitation chapter C: part 11: neonatal resuscitation: 2010 international consensus on cardiopulmonary resuscitation and emergency cardiovascular care science with treatment recommendations. Circulation. 2010;122:S516–38.

    Article  PubMed  Google Scholar 

Download references

Authors’ contributions

JL participated in planning and carried out the piglet experiment, analyzing the data and drafting the manuscript. JE participated in planning and carried out the piglet experiment, analyzing the data and drafting the manuscript. KO participated in planning and the piglet experiment, analyzing the data and drafting the manuscript. HE participated in planning and carried out the piglet experiment, analyzing the data and drafting the manuscript. All authors read and approved the final manuscript.

Acknowledgements

The authors wish to extend their gratitude to Sandnes Educational and Research Centre Høyland (SEARCH) and Dr. Nils Petter Oveland for providing facilities, expertise and valuable advice and assistance in performing the experiments. We also like to thank Dr. Ronnaug Solberg for valuable inputs regarding piglet studies, and Dr. Jörn Shulz for statistical expertise.

Competing interests

Joar Eilevstjønn is an employee of Laerdal Global Health/Medical. Jørgen Linde is married to an employee of Laerdal Global Health. All authors had no other financial relationships with any organizations that might have an interest in the submitted work in the previous 5 years or no other relationships or activities that could appear to have influenced the submitted work.

Availability of data and materials

Request for data from this study will be released upon request to data steering comity of Safer Births. For more information: http://www.saferbirths.com.

Ethics approval and consent to participate

The protocol was approved by The Research Animal Council for the Rogaland Region. The piglets were handled in accordance with the European Guidelines for Use of Experimental Animals by certified Federation of European Laboratory Animal Science Association category C researchers (FELASA C).

Funding

JE and HE received a PhD grants from the Laerdal foundation.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Jørgen E. Linde.

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Linde, J.E., Eilevstjønn, J., Øymar, K. et al. Feasibility of a prototype newborn resuscitation monitor to study transition at birth, measuring heart rate and ventilator parameters, an animal experimental study. BMC Res Notes 10, 235 (2017). https://doi.org/10.1186/s13104-017-2530-z

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s13104-017-2530-z

Keywords