Table 1 Means and standard deviations of temperatures reached in function of lesioning duration and power.

From: Predictability of thermo-lesions using electrodes for deep brain stimulation - an in vitro study

  1.25 W 1.51 W 1.73 W 2.00 W
30 s 59.2 +/- 4.0°C 65.9 +/- 4.4°C 67.2 +/- 3.7°C 71.8 +/- 4.2°C
60 s 63.5 +/- 4.2°C 69.3 +/- 3.5°C 73.7 +/- 5.3°C 82.1 +/- 5.7°C
90 s 66.3 +/- 5.2°C 71.3 +/- 4.6°C 76.6 +/- 5.5°C 82.9 +/- 5.9°C
120 s 68.1 +/- 5.5°C 72.9 +/- 4.7°C 78.3 +/- 5.2°C 85.5 +/- 7.1°C