Table 1 Means and standard deviations of temperatures reached in function of lesioning duration and power.

From: Predictability of thermo-lesions using electrodes for deep brain stimulation - an in vitro study

 

1.25 W

1.51 W

1.73 W

2.00 W

30 s

59.2 +/- 4.0°C

65.9 +/- 4.4°C

67.2 +/- 3.7°C

71.8 +/- 4.2°C

60 s

63.5 +/- 4.2°C

69.3 +/- 3.5°C

73.7 +/- 5.3°C

82.1 +/- 5.7°C

90 s

66.3 +/- 5.2°C

71.3 +/- 4.6°C

76.6 +/- 5.5°C

82.9 +/- 5.9°C

120 s

68.1 +/- 5.5°C

72.9 +/- 4.7°C

78.3 +/- 5.2°C

85.5 +/- 7.1°C