Skip to main content

Table 2 Average concentrations found, accuracy and within-day assay precision for naltrexone and 6-β-naltrexol

From: Analysis of naltrexone and its metabolite 6-beta-naltrexol in serum with high-performance liquid chromatography

Compound Concentration added (ng/ml) Concentration found (ng/ml) Accuracy (%) Precision (%)
Naltrexone 5 4.97 0.7 5.7
  150 146.6 2.3 0.9
6-β-naltrexol 5 4.87 2.6 4.2
  150 146.1 2.6 0.8