Skip to main content

Table 1 Comparison between the old Actigraph model 7164 and the new Actigraph GT1M depending on wear time threshold (i.e. 10, 30 or 60 minutes wear time), values are median ± IQR (interquartile range)

From: Accelerometer measured daily physical activity and sedentary pursuits–comparison between two models of the Actigraph and the importance of data reduction

10 min wear time

 

Mod 7164

GT1M

P-value

7164/ GT1M

Sedentary (<100 cpm)

184 ± 50

159 ± 36

0.12

1.16

MVPA (>1952 cpm)

61 ± 20

56 ± 23

0.01

1.09

VPA (>5725 cpm)

12 ± 8

9 ± 3

0.002

1.33

VVPA (>9499 cpm)

3.2 ± 3.0

0.3 ± 1.1

<0.001

10.67

30 min wear time

 

Mod 7164

GT1M

P-value

7164/ GT1M

Sedentary (<100 cpm)

337 ± 76

331 ± 64

1.0

1.02

MVPA (>1952 cpm)

61 ± 21

56 ± 23

0.01

1.09

VPA (>5725 cpm)

12 ± 8

9 ± 3

0.002

1.33

VVPA (>9499 cpm)

3.2 ± 3.0

0.4 ± 1.1

<0.001

8.00

60 min wear time

 

Mod 7164

GT1M

P-value

7164/ GT1M

Sedentary(<100 cpm)

429 ± 64

438 ± 53

0.01

0.98

MVPA (>1952 cpm)

61 ± 21

56 ± 23

0.01

1.09

VPA (>5725 cpm)

12 ± 8

9 ± 3

0.002

1.33

VVPA (>9499 cpm)

3.2 ± 3.0

0.4 ± 1.1

<0.001

8.00

  1. Moderate and vigorous physical activity (MVPA), vigorous physical activity (VPA) and very vigorous physical activity (VVPA). The ratio between median data output of 7164 and GT1M is also expressed.