Skip to main content

Table 1 The compression system performed with a high level of accuracy which is indicated by the linearity, repeatability, and percent error

From: Potential regenerative rehabilitation technology: implications of mechanical stimuli to tissue health

Pressure (kPa)

Linearity (%)

Repeatability (%FS)

Error (%FS)

138

3.79

0.54

0.51

207

1.75

0.57

0.42

276

1.26

0.61

0.44

345

0.83

0.69

0.38

414

0.58

0.54

0.32

  1. These metrics were calculated using the 10 cycles at each air pressure (%FS = percent full scale).