Skip to main content

Table 3 Review summary of SLN biopsy in gynaecological cancers using NIR fluorescence

From: A prospective pilot study of detection of sentinel lymph nodes in gynaecological cancers using a novel near infrared fluorescence imaging system

Reference

Cancer type

Surgical procedure

Imaging device

No. of patients

Key results

Furukawa [26]

Cervical

Open

Photodynamic eye (Hamamatsu)

12

At least 1 SLN identified in 83 % of the patients

Crane [27]

Cervical

Open

Custom made prototype

10

Intraoperative detection rate = 64 % relative to total number of ex vivo fluorescent SLNs

Crane [11]

Vulvar

Open

Custom made prototype

10

Detection rates relative to gamma probe: fluorescence = 89.7 %, patent blue = 72.4 %

Van der Vorst [10]

Cervical

Open

Mini-FLARE

9

At least 1 SLN identified in each patient. Optimal ICG:HSA dose: 500 μM

Rossi [28]

Endometrial/cervical

Robotic-assisted laparoscopy

SPY scope (Novadaq)

20

At least 1 SNL identified in 85 % of the cases

Holloway [8]

Endometrial

Robotic-assisted laparoscopy

Da Vinci NIR fluorescence imaging system

35

At least 1 SLN identified in 97 % of the patients with fluorescence and 77 % with calorimetric analysis

Hutteman [9]

Vulvar

Open

Mini-FLARE

9

Detection rates relative to gamma probe: fluorescence = 100 %, patent blue = 71 %

Optimal ICG:HSA dose: 750 μM

Schaafsma [29]

Cervical

Open

Mini-FLARE

18

At least 1 SLN identified in 97 % of the patients. No significant difference in signal-to-background ratio between ICG alone and ICG:HSA

Schaafsma [13]

Vulval

Open

Mini-FLARE

24

Detection rates relative to gamma probe: fluorescence = 100 %, patent blue = 77 %

No significant difference in detection rate between ICG alone and ICG:HSA

Rossi [12]

Endometrial

Robotic-assisted laparoscopy

SPY® scope (Novadaq)

29

SLN detection rate: 82 % for cervical injection, 33 % for hysteroscopic endometrial injection

Mathéron [18]

Vulval

Open

Photodynamic eye (Hamamatsu)

15

SLN detection rate: 98 % radioactive, 96 % fluorescence, 65 % blue dye

Jewell [21]

Endometrial/cervical

Robotic-assisted laparoscopy

PINPOINT® (Novadaq)

227

SLN detection rate: 95 %. Combined use of ICG and blue dye proved unnecessary

Sinno [30]

Endometrial

Robotic-assisted laparoscopy

PINPOINT® (Novadaq)

71

Fluorescence ICG superior than blue dye (78.9 % vs. 42.4 % bilateral detection rate)