From: A novel experimental setup for evaluating the stiffness of ankle foot orthoses
PL | PU | DL | DU | |||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|
ICC (%) | SEM (Nm/°) | SDD (Nm/°) | ICC (%) | SEM (Nm/°) | SDD (Nm/°) | ICC (%) | SEM (Nm/°) | SDD (Nm/°) | ICC (%) | SEM (Nm/°) | SDD (Nm/°) | |
Test–retest | 1 | 0 | 0 | 1 | 0.25 | 0.69 | 1 | 0.06 | 0.16 | 0.99 | 0.09 | 0.24 |
Intra-tester | 0.99 | 0.29 | 0.81 | 0.97 | 0.40 | 1.10 | 0.97 | 0.19 | 0.53 | 0.94 | 0.25 | 0.70 |
Inter-tester | 0.99 | 0.19 | 0.54 | 0.98 | 0.29 | 0.80 | 0.94 | 0.25 | 0.69 | 0.98 | 0.12 | 0.32 |