Coherence and reliability of a wearable inertial measurement unit for measuring postural sway

Abstract

Objective

The aim of this study was to test the coherence of a wearable device, an inertial measurement unit (IMU) against the gold standard; also, to test the intra-trial reliability. This study has a cross-sectional design, where measurement of postural sway in the medio-lateral and anterior–posterior directions was performed simultaneously on a force plate and with a IMU called the Snubblometer (“snubbla” is stumble in Swedish). Thirty-two healthy volunteers participated in the tests.

Results

The coherence between the IMU and the force plate was 0.84 (strong) in the medio-lateral direction with eyes open (EO) and 0.88 (strong) with eyes closed (EC). The ICC for intra-trial reliability for the IMU varied between 0.50 and 0.67 (moderate to good) with a CV between 17.8 and 22.1% and ICC varied between 0.75 and 0.86 (good) for inter-trial reliability, with an SEM of 0.98 to 1.96 mm/s. We have demonstrated that the IMU was both reliable and highly coherent with golden standard, although the two assessment methods were not interchangeable. The ability to move the balance lab out into real life in the form of a wearable device will provide opportunities to perform research that has not been possible before.

Introduction

Balance is defined as “sensing the position of the body’s centre of mass and moving the body to adjust the position of the centre of mass over the base of support provided by the feet” [1]. Balance can be measured in many different ways and situations, depending on the task and on the objective of measurement; efforts have been made to determine adequate and clinically useful measures of balance for different populations specifically in terms of identifying people at risk of falling [2,3,4]. However, these efforts do not seem to be sufficient to determine who actually sustains a fall-related fracture [5]. Postural sway is one measure of balance that has been suggested as appropriate for identifying individuals who are at high risk of falling [6]. In particular, the medio-lateral displacement of the centre of pressure in postural sway seems to be able to predict falls [7]. Postural sway is the movement caused by the active feedback control mechanisms responding to the forces of gravity on the body while standing upright [8]. Postural sway is often measured using a force plate [9]; this is suitable for laboratory use but harder to apply in clinical practice. The use of a wearable device as an alternative method of measurement may increase the usefulness of postural sway as measure of balance and of fall risk in clinical practice. However, it is important to first explore the validity of such a device, in terms of coherence and reliability.

The aim of this study was therefore to test the coherence of an inertial measurement unit against the gold standard—a force plate. We also aimed to test the intra-trial reliability of the IMU.

Main text

Method

The study is an experimental cross-sectional study.

Measurements

The force plate used in the study, representing the gold standard for measuring postural sway (Good Balance™, Metitur Ltd, Finland). The force plates were tested for test–retest and intra-session reliability as well as validity [9,10,11].

The inertial measurement unit (IMU) used comprises a nine-axis Inertial Measurement Unit (IMU), called the Snubblometer® (‘snubbla’ is ‘stumble’ in Swedish), http://www.Infonomy.com).

Participants

Healthy participants were recruited among students and staff at the Health Science Centre, Lund University. Inclusion criteria were: no dizziness or balance problems, no neck pain, no newly acquired injury to the hip, knee, or foot for the last 2 months, and corrected visual impairment, if any. None of the participants had any hearing problems. The group consisted of 32 people: 22 women and 10 men between 20 and 66 years of age (mean 29.8, SD 13.3).

Procedure

Postural sway was measured with the subject standing on a force plate while wearing the IMU attached at the lumbar back, in line with L4. Measurements were taken from the force plate and from the IMU simultaneously. The starting position on the force plate was a normal standing position with the arms hanging by the sides and feet in the standardised position with toes facing 30o outwards. The subjects were told to hold the head in a neutral position and to stand as still as possible. Two different measurements were performed for 30 s each, one with the eyes open and another with the eyes closed. The tests were performed three times on each subject, with a resting time of 5 min between them.

Statistics

All data were tested for normal distribution. Since the data were normally distributed, parametric statistics was applied. Mean and standard deviations (SD) are used when describing the data. Paired student’s t-test was used when comparing postural sway assessed with the balance plate against the algorithms used to assess postural sway with the IMU. Pearson’s correlation (rp) was used for analysis of the coherence between the balance plate and the IMU. A Bland–Altman analysis was performed to detect any systematic variance [12]. Limits of agreement (LOA) and bias were calculated for the Bland–Altman plots to show upper and lower LOA [13]. The p < 0.05 criterion was used for establishing statistical significance.

Test–retest correlations were calculated both for intra-session and inter-session relations. Intra-class correlation coefficient (ICC) and coefficient of variance (CV) were used to analyse intra-session and inter-session reliability. ICC ranges from negative 1 (negative agreement) to 1 (perfect agreement) [14]. All statistical analyses were performed using SPSS for Windows (version 25.0, SPSS Inc., Chicago, Illinois, USA).

Ethics

All subjects participated voluntarily and consent was obtained before the tests were carried out. The measures were performed according to guidelines set out by the Helsinki Declaration of 1974. No physical operation was performed and the participants were not influenced by the procedure, either physically or psychologically. The study was approved by the Regional Ethical Review Board in Lund Dnr 2016/585.

Results

Before the statistical analysis was performed, normal distribution of the data was confirmed. Measures of postural sway on the force plate and with the IMU are displayed in Table 1.

Table 1 Postural sway assessed with a force plate (FP) and with an inertial measurement unit (IMU)

Correlation between balance plate and IMU

Analysis showed very high and statistically significant relations between the IMU and the force plate (Table 2).

Table 2 Correlations of assessments of postural sway assessed with a force plate (FP) and with an inertial measurement unit (IMU)

Bias and limits of agreement

LOA and the values for 2SD is presented in Additional file 1: Figure S1. The Bland–Altman plots for postural sway with eyes closed in the medio-lateral direction show a bias between the assessment on the FP and the IMU of − 0.93 mm/s and LOA − 4.33 to 2.47 mm/s and in the anterior–posterior direction with a bias of − 1.29 mm/s and LOA − 3.59 to 1.01 mm/s. With eyes open, the bias between FP and IMU was − 3.20 mm/s and LOA − 8.94 to 2.54 mm/s in the medio-lateral direction. In the anterior–posterior direction with eyes open, a bias of − 2.74 mm/s was revealed and LOA − 7.02 to 1.54 mm/s.

Reliability analysis

No statistical differences were found between the three trials (Table 3) for either of the two methods or with eyes closed or opened. When analysing intra-trial reliability with ICC, the ICC varied between 0.74 and 0.84 for the balance plate and between 0.50 and 0.67 for the IMU. For inter-trial reliability, the ICC spread between 0.89 and 0.94 for the force plate and between 0.75 and 0.86 for the IMU.

Table 3 Intra- and inter-trail reliability for the inertial measurement unit and the force plate

Discussion

This study outlines the coherence and reliability of a wearable device. The main finding of this study was that the IMU displayed strong coherence with measures of postural sway on a force plate, both in the medio-lateral and anterio-posterior directions and both with eyes open and eyes closed. The bias and LOA demonstrated however that the two methods are not interchangeable. The intra-trial reliability for the IMU was high.

Coherence

In this study, coherence between 0.71 and 0.88 is to be considered as high. The high coherence found in eyes-closed tests and in the anterior–posterior direction is promising, since this has been difficult to achieve in earlier research [15, 16].

Reliability

The ICC in our study ranges from 0.50 to 0.86 and the reliability for the IMU is therefore considered moderate to good. Measures of gait using wearable sensors have been tested, showing good reliability [17] but differences between different models of accelerometers [18].

Strengths

A strength of our study was that the assessments were performed in a standardised manner on the same population. The assessments with the IMU and force plate were made simultaneously for the same individual.

Implications for future research

A wearable device that can measure postural sway in a valid and reliable way will be beneficial for use both in clinic and in research. The ability to move the balance lab out into real life in the form of a wearable device will provide opportunities to perform research that has not been possible before. However, our findings on healthy people needs to be confirmed on people with different balance disorders.

Our research group is currently performing tests with the IMU attached to the thigh in order to measure gait parameters as well as near falls.

Conclusions

The wearable device called the Snubblometer® tested in this study showed a strong coherence with measures of postural sway on a force plate, both in the medio-lateral and anterio-posterior directions and both with eyes open and eyes closed, although the two methods were found not to be interchangeable. The intra-trial reliability for the wearable device was moderate to good and the inter-trial reliability was good.

Limitations

  • The repeated measurements were made at one occasion.

  • Making repeated measurement at different day would have made an inter-trail analysis possible.

  • The study population in this study is healthy volunteers, who can be assumed have less postural sway than populations with diseases affecting balance.

Abbreviations

FP:

force plate

Med-Lat:

medio-lateral displacement

Ant–Post:

anterior–posterior displacement

SD:

standard deviation

CV:

coefficient of variance

rp :

Pearson’s correlation coefficient

LOA:

limits of agreement

ICC:

intraclass correlation coefficient

IMU:

inertial measurement unit

References

  1. Nashner L, Shupert C, Horak F. Head-trunk movement coordination in the standing posture. Prog Brain Res. 1988;76:243–51.

    Article  CAS  Google Scholar 

  2. Albertsson DM, Mellstrom D, Petersson C, Eggertsen R. Validation of a 4-item score predicting hip fracture and mortality risk among elderly women. Ann Fam Med. 2007;5(1):48–56.

    Article  Google Scholar 

  3. Ekvall Hansson E, Magnusson M. Vestibular asymmetry predicts falls among elderly patients with multi-sensory dizziness. BMC Geriatr. 2013;13(1):77.

    Article  Google Scholar 

  4. Hansson EE, Mansson NO, Ringsberg KA, Hakansson A. Falls among dizzy patients in primary healthcare: an intervention study with control group. Int J Rehabil Res. 2008;31(1):51–7.

    Article  Google Scholar 

  5. Beckman A, Hansson EE. Fractures in people with dizziness: 5-year follow-up. J Am Geriatr Soc. 2011;59(9):1767–9.

    Article  Google Scholar 

  6. Muir JW, Kiel DP, Hannan M, Magaziner J, Rubin CT. Dynamic parameters of balance which correlate to elderly persons with a history of falls. PLoS ONE. 2013;8(8):e70566.

    Article  CAS  Google Scholar 

  7. Piirtola M, Era P. Force platform measurements as predictors of falls among older people—a review. Gerontology. 2006;52(1):1–16.

    Article  Google Scholar 

  8. Maurer C, Peterka RJ. A new interpretation of spontaneous sway measures based on a simple model of human postural control. J Neurophysiol. 2005;93(1):189–200.

    Article  Google Scholar 

  9. Era P, Sainio P, Koskinen S, Haavisto P, Vaara M, Aromaa A. Postural balance in a random sample of 7,979 subjects aged 30 years and over. Gerontology. 2006;52(4):204–13.

    Article  CAS  Google Scholar 

  10. Bauer C, Groger I, Rupprecht R, Gassmann KG. Intrasession reliability of force platform parameters in community-dwelling older adults. Arch Phys Med Rehabil. 2008;89(10):1977–82.

    Article  Google Scholar 

  11. Goldie PA, Bach TM, Evans OM. Force platform measures for evaluating postural control: reliability and validity. Arch Phys Med Rehabil. 1989;70(7):510–7.

    CAS  PubMed  Google Scholar 

  12. Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1(8476):307–10.

    Article  CAS  Google Scholar 

  13. Weir JP. Quantifying test-retest reliability using the intraclass correlation coefficient and the SEM. J Strength Cond Res. 2005;19(1):231–40.

    PubMed  Google Scholar 

  14. Kirkwood BR, Sterne JAC. Essential medical statistics. 2nd ed. Bodmin: Blackwell Publishing; 2003. p. 501.

    Google Scholar 

  15. Heebner NR, Akins JS, Lephart SM, Sell TC. Reliability and validity of an accelerometry based measure of static and dynamic postural stability in healthy and active individuals. Gait Posture. 2015;41(2):535–9.

    Article  Google Scholar 

  16. Rouis A, Rezzoug N, Gorce P. Validity of a low-cost wearable device for body sway parameter evaluation. Comput Methods Biomech Biomed Eng. 2014;17(Suppl 1):182–3.

    Article  Google Scholar 

  17. Salarian A, Horak FB, Zampieri C, Carlson-Kuhta P, Nutt JG, Aminian K. iTUG, a sensitive and reliable measure of mobility. IEEE Trans Neural Syst Rehabil Eng. 2010;18(3):303–10.

    Article  Google Scholar 

  18. Tanha T, Tornberg AB, Wollmer P, Dencker M. Head-to-head comparison between actigraph 7164 and GT1M accelerometers in adolescents. Clin Physiol Funct Imaging. 2013;33(2):162–5.

    Article  Google Scholar 

Download references

Authors’ contributions

Both authors read and approved the final manuscript.

Acknowledgements

Thanks to students and staff at the Health Science Centre for their participation. Thanks to Emil Karlsson, David Nilsson, Martin Johansson, Nils Annertz, Hanna Hallström and Alice Jönsson for help with data collection. Thanks to Infonomy for providing the mobile device.

Competing interests

The authors declare that they have no competing interests.

Availability of data and materials

The datasets generated and analysed during the current study are not publicly available due to the General Data Protection Regulation but are available from the corresponding author on reasonable request.

Consent for publication

No individual data is reported. Not applicable.

Ethics approval and consent to participate

All subjects participated voluntarily and written consent was obtained before the tests were carried out. The measures were performed according to guidelines set out by the Helsinki Declaration of 1974. No physical operation was performed and the participants were not influenced by the procedure, either physically or psychologically. The study was approved by the Regional Ethical Review Board in Lund Dnr 2016/585.

Funding

This study was funded by Vinnova (the Swedish Innovation Agency). Vinnova had no part in designing the study or in analysis or interpretation of collected data. Vinnova did not participate in writing the manuscript.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Eva Ekvall Hansson.

Additional file

Additional file 1: Figure S1.

Postural sway assessed with the force plate (FP) Inertial Measurement Unit (IMU) revealed Bland–Altman plots for average postural sway (mm/s) for Med-Lat sway with (A) EC (Bias: − 0.93 mm/s; LOA − 4.33 to 2.47 mm/s) and (B) EO (Bias: − 1.29 mm/s; LOA − 3.59 to 1.01 mm/s) and for Ant–Post sway with (C) EC (Bias: − 3.20 mm/s; LOA − 8.94 to 2.54 mm/s), (D) EO (Bias: − 2.74 mm/s; LOA − 7.02 to 1.54 mm/s).

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Ekvall Hansson, E., Tornberg, Å. Coherence and reliability of a wearable inertial measurement unit for measuring postural sway. BMC Res Notes 12, 201 (2019). https://doi.org/10.1186/s13104-019-4238-8

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s13104-019-4238-8

Keywords