Table 2 Showing the diagnostic performance of HIV–Syphilis SD Bioline Duo Kit

From: Laboratory assessment of SD Bioline HIV/Syphilis Duo Kit among pregnant women attending antenatal clinic Mayuge Health Center III, East central Uganda

RDT Infection RDT result Positivea Negativea Total
Diagnostic performance of HIV–Syphilis SD Bioline Duo Kit
 SD HIV/Syphilis Duo HIV Positive 7 0 7
Negative 0 375 375
Total 7 375 382
Syphilis Positive 8 0 8
Negative 0 374 374
Total 8 374 382
RDT Infection % Sensitivity (95% CI) % Specificity (95% CI)   
Showing the operation performance of the HIV–Syphilis SD Bioline Duo Kit   
SD Bioline HIV–Syphilis Duo assay HIV 100.0 (99.5 to 100.0) 100.0 (98.6 to 100.0)   
Syphilis 100 (98.3 to 100.0) 100 (98.6 to 100)   
  1. aThe gold standard method for HIV was the Ministry of Health (Uganda) HIV testing Algorithm; while for Syphilis, it was the Treponema pallidum Hemagglutination Assay (TPHA)
\