From: Developing a high-performance liquid chromatography fast and accurate method for quantification of silibinin
Parameters
Value
Slope
3110.6
Intercept
− 17,877
Standard deviation of calibration curve
7515.2
Linear range (µg/ml)
10–100
Coefficient of determination (R2)
0.996
LLOQ (µg/ml)
10
ULOQ (µg/ml)
100
LOD (µg/ml)
7.25
LOQ (µg/ml)
24.2