Skip to main content

Table 2 Dwell times and dwell positions calculated at different dose values for 35cc balloon applicator volume

From: Calculated and measured radiation dose for the low energy xoft axxent eBT X-ray source

35 cc

 

20 Gy

16 Gy

10 Gy

Dwell Position (cm)

Calculated Time (s)

Calibrated Time

(s)

Calculated Time (s)

Calibrated Time

(s)

Calculated Time (s)

Calibrated Time

(s)

24.5

7.4

6.5

5.9

5.2

3.7

3.2

24

90.6

79.1

72.5

63.4

45.3

39.4

23.5

161.2

140.7

128.9

112.7

80.6

70.1

23

184

160.6

147.2

128.7

92

80.0

22.5

163.1

142.4

130.5

114.1

81.6

71.0

22

11.1

9.7

8.9

7.8

5.6

4.9