From: Calculated and measured radiation dose for the low energy xoft axxent eBT X-ray source
35 cc | ||||||
---|---|---|---|---|---|---|
20 Gy | 16 Gy | 10 Gy | ||||
Dwell Position (cm) | Calculated Time (s) | Calibrated Time (s) | Calculated Time (s) | Calibrated Time (s) | Calculated Time (s) | Calibrated Time (s) |
24.5 | 7.4 | 6.5 | 5.9 | 5.2 | 3.7 | 3.2 |
24 | 90.6 | 79.1 | 72.5 | 63.4 | 45.3 | 39.4 |
23.5 | 161.2 | 140.7 | 128.9 | 112.7 | 80.6 | 70.1 |
23 | 184 | 160.6 | 147.2 | 128.7 | 92 | 80.0 |
22.5 | 163.1 | 142.4 | 130.5 | 114.1 | 81.6 | 71.0 |
22 | 11.1 | 9.7 | 8.9 | 7.8 | 5.6 | 4.9 |