Table 4 % difference between the planned dose and measured dose taken with 0.9% isotonic NaCl solution

From: Calculated and measured radiation dose for the low energy xoft axxent eBT X-ray source

 

Aplicator Volume

(cc)

Measured Dose

(Gy)

Planned Dose vs Measured Dose

(%)

20 Gy

30 cc

22.92 ± 2.11

14.60

35 cc

22.88 ± 1.55

14.40

40 cc

22.14 ± 1.39

10.70

16 Gy

30 cc

18.18 ± 2.77

13.62

35 cc

18.32 ± 3.52

14.50

40 cc

18.06 ± 3.78

12.87

10 Gy

30 cc

10.90 ± 0.88

9.00

35 cc

10.79 ± 0.59

7.90

40 cc

10.41 ± 0.68

4.10