Skip to main content

Table 4 Expected and observed concentrations for Rosetta spike-ins 11 and 12, including %CV and %Bias.

From: Multiplexed measurements of gene signatures in different analytes using the Nanostring nCounter™ Assay System

Rosetta spike-in #11

Transcript name

Intended conc, AU

Best Fit Expected

Observed

StdDev

CV, %

Bias, %

r60_a20:50:rp

100

32.09

28.62

1.47

5.2

10.8

r60_a22_rp

10

3.21

4.71

0.28

5.9

46.9

r60_a104_rp

10

3.21

2.62

0.14

5.2

18.5

r60_1_rp

10

3.21

4.29

0.26

6.1

33.6

r60_a107_rp

30

9.63

10.56

0.70

6.6

9.7

r60_3_rp

3

0.96

1.02

0.09

9.1

6.1

r60_a135_rp

9

2.89

3.13

0.16

5.0

8.3

r60_a97_rp

0.5

0.16

0.14

0.03

21.8

15.5

r60_n11:30:rp

1.5

0.48

0.34

0.07

21.3

30.1

Rosetta spike-in #12

r60_a20:50:rp

100

44.18

37.56

2.25

6.0

15.0

r60_a22_rp

100

44.18

81.54

5.51

6.8

84.5

r60_a104_rp

30

13.25

12.16

0.70

5.8

8.2

r60_1_rp

10

4.42

5.85

0.30

5.1

32.3

r60_a107_rp

10

4.42

4.49

0.36

8.0

1.6

r60_3_rp

9

3.98

4.28

0.25

5.8

7.7

r60_a135_rp

3

1.33

1.30

0.08

6.2

1.6

r60_a97_rp

1.5

0.66

0.61

0.07

11.3

7.9

r60_n11:30:rp

0.5

0.22

0.12

0.04

29.7

43.6

  1. Calculations based on a per-sample basis.