From: Accuracy of tympanic temperature measurement using an infrared tympanic membrane thermometer
Axillary vs. tympanic temperature
Correlation Coefficient
Bias ± SD
95% CI
Limits of agreement
0.697
−0.093 ± 0.72
−0.20; 0.02
−1.54 to 1.36