Table 1 Baseline results for the TORCH panel tests evaluated with the sigma metric

From: Six-sigma and quality planning of TORCH tests in the Peruvian population: a single-center cross-sectional study

TORCH panel

Abs

Cutt-off

Quality requirementsa

Imprecisionb

Sigma

∆SE

Toxoplasma gondii

 IgM

0.007

0.005

28.6

5

6

4.06

 IgG

0.012

0.008

33.3

6.2

5

3.73

Rubella

 IgM

0.027

0.02

25.9

2.6

10

8.35

 IgG

0.219

0.177

19.2

2.4

8

6.35

Cytomegalovirus

 IgM

0.054

0.047

13

2.3

6

3.99

 IgG

0.022

0.019

13.6

2.6

5

3.59

Herpes simplex virus 1

 IgM

0.084

0.074

11.9

3

4

2.32

 IgG

0.023

0.019

17.4

2.7

6

4.79

Herpes simplex virus 2

 IgM

0.055

0.045

18.2

4.6

4

2.30

 IgG

0.015

0.011

26.7

5.5

5

3.20

  1. Abs absorbance (optical density δ)
  2. aIn %
  3. bIn %CV