Table 2 The sensitivity and specificity of SSf-COVID19 kit with saliva compared with RT-PCR with nasopharyngeal swabs

From: Clinical application of the Panbio™ COVID-19 Ag rapid test device and SSf-COVID19 kit for the detection of SARS-CoV-2 infection

Overall patients (n = 157)

 

Nasopharyngeal swab

Total

Positive

Negative

SSf-COVID19 kit

Positive

32

0

32

Negative

2

123

125

 

Total

34

123

157

Sensitivity = 94.1%; Specificity = 100%; Cohen’s weighted kappa index = 0.96

Patients with cycle threshold values ≤ 25 (n = 157)

 

Nasopharyngeal swab

Total

Positive

Negative

SSf-COVID19 kit

Positive

28

4

32

 

Negative

1

124

125

 

Total

29

128

157

Sensitivity = 96.6%; Specificity = 97.0%; Cohen’s weighted kappa index = 0.90

Symptomatic patients (n = 43)

 

Nasopharyngeal swab

Total

Positive

Negative

SSf-COVID19 kit

Positive

27

0

27

 

Negative

1

15

16

 

Total

28

15

43

Sensitivity = 96.4%; Specificity = 100%; Cohen’s weighted kappa index = 0.95

  1. RT-PCR, reverse transcription-polymerase chain reaction