Table 2 A comparison between Lactate Pro meter results and the automated lactate analyser (Abx Pentra 400) of amniotic fluid collected under sterile conditions and spiked with a known quantity of lactate

From: A pilot study to determine the feasibility of collecting amniotic fluid samples from women during labour and measuring amniotic fluid lactate at point of care

Lactate concentration [Lactate] (mmol/L) Lactate Pro [Lactate] (mmol/L) AbX Pentra 400
Baseline 9.5 ± 0.1 10.0 ± 0.2
0.5 mM* 10.0 ± 0.1 10.6 ± 0.5
1.0 mM 10.5 ± 0.5 10.5 ± 0.1
5 mM 14.6 ± 0.6 14.2 ± 0.3
10 mM 21 ± 0.8 18.4 ± 0.1
20 mM Hi 26.3 ± 0.7
  1. * Lactate concentration added (in addition to baseline levels of lactate in amniotic fluid).