Dental root elevator embedded into a subgingival caries: a case report
BMC Research Notes volume 8, Article number: 60 (2015)
Abstract
Background
Breakage of surgical instruments is a rare complication. A mistake in operator technique or sub-standard/aged tools could lead to this type of accident. A tooth elevator is an instrument used in minor oral surgical procedures to luxate the tooth or fractured root from its socket. The authors have not found any previously published cases reporting the breakage of a tooth elevator tip which then remained as a foreign body in a hidden caries cavity.
Case presentation
A 28-year-old African black male was referred to a hospital in Tanzania for an intraoral radiography. The patient explained that six months previously his mandibular left third molar had been extracted. Whilst the healing process had been satisfactory, he had recently experienced acute oral pain in this region. The dental X-ray showed an image consistent with a piece of broken metal embedded in a distal subgingival caries at the mandibular left second molar.
Conclusion
Oral and dental surgeons should take particular care when employing metal instruments with strong force in poorly visible areas. A radiographic study should be carried out when instrument breakage occurs. If an unexpected accident takes place during a surgical procedure, the patient must be informed in accordance with ethical codes, and suitable measures adopted to resolve the issue.
Background
Foreign bodies in the oropharyngeal region are a problem frequently faced by otolaryngologists and dental surgeons. The accidental ingestion of sharp elements such as fish bones or toothpick pieces is quite common in the oral cavity and removal can be easily performed through simple manipulation. However, other foreign bodies in this region may require a surgical approach and/or endoscopic techniques under general anaesthesia [1,2].
Accidents can take place during surgery due to a number of factors including operator technique and sub-standard or aged instruments. Manufacture is strictly controlled, particularly in the case of dental, medical and surgical instruments which could cause serious injury to patients if they proved to be faulty. Occasionally, however, alterations in manufacturing technique or ineffective quality control occur and they are employed unknowingly [3].
A tooth elevator is an instrument used in minor oral surgical procedures to luxate the tooth or fractured root from its socket. There is a vast range of elevator designs, including right and left sides for several types (Figure 1). Alveolar bone fracture and fracture/luxation of the adjacent tooth are two common problems associated with their use.
It is well known that any tooth extraction requires a preoperative radiograph to obtain information about root morphology and possible concomitant dental pathology such as interproximal caries or apical lesions. The aim of the present report is to describe an incidental radiographic finding that was consistent with a tooth elevator tip which had remained as a foreign body in a hidden caries cavity for approximately six months.
Case presentation
A 28-year-old African black male was referred to a hospital in Tanzania for an intraoral radiography. His major complaint was an acute, diffuse-patterned oral pain in the left mandibular area. The patient had no significant previous medical history. A clinical examination confirmed satisfactory dental hygiene without any apparently relevant findings. The patient explained that his mandibular left third molar had been extracted six month previously and that the surgery had been quite laborious and lengthy. Although the healing process had been satisfactory, considerable pain had recently begun in this area. The dental X-ray showed an image consistent with a piece of metal embedded in a subgingival distal caries at the mandibular left second molar (37), which was probably related to a previous tooth impaction process of the mandibular left third molar (38). Two periapical lesions detected at the mandibular left second molar roots might have justified his acute tooth ache due to acute exacerbation (Figure 2).
Initially, the differential diagnosis of this radiographic finding was that of a possibly loosened or released piece of amalgam. Nevertheless, the patient had never previously undergone any dental fillings. After repeating the X-ray twice in order to rule out a possible image artefact, it was concluded that during the third molar extraction a curved blade elevator tip had fractured and impacted into the subgingival caries cavity. As the patient had only been referred for a dental X-ray he was given a written note explaining the finding. His dentist was informed about the need to perform an occlusal access cavity in order to initiate a root canal treatment and remove the foreign body from the second molar (37).
Discussion
The breakage of some instruments, such as endodontic files and dental burs, due to a number of factors including defective manufacturing, stress, fatigue, rust, and poor handling is not unknown in dentistry [4]. Few papers, however, in the literature have dealt with the breakage of instruments used for exodontias. Yasuhara et al., registered various medical accidents caused by defective surgical instruments over two years. In the maxillofacial speciality the authors reported 7 incidences out of 548 operations [5]. According to Kluess et al., it is important that any incident with orthopaedic surgical instruments should be reported to the manufacturer and the health authorities for sufficient processing and risk assessment of the accident [6]. In the case of some reusable metal instruments both titanium alloys and stainless steel are in the high performance range. The latter is the most widely used material for instruments and, according to surgical requirements, its alloys vary: the most frequent being martensitic and austenitic stainless steels. Biomedical cutting instruments are often made of martensitic stainless steel due to its pronounced durability coupled with acceptable corrosion resistance. Surgical instruments that may be subject to high pressure forces, such as a dental elevator, are composed of austenitic stainless steel as it is less brittle [7]. A safe and effective elevator should have extreme values for torque, and high stress values [8].
Surgical instruments manufactures should carry out strict quality controls and have their instruments bear a visible mark as a sign of guarantee. Various authors have suggested that the inferior quality of some surgical instruments may be a reflection of poor working conditions and low standards, particularly in the developing world. Responsibility lies with the suppliers from developed countries manufacturing in the developing world who behave in an unethical manner, maximising profits and minimising the remuneration of the people who actually produce the goods [9-11].
The location and retrieval of broken fragments during a tooth extraction procedure should not be a serious problem, in most cases the fragment is immediately identified. Any instrument breakage implies the obligation to search for the fractured fragment and remove it in order to avoid possible infection and prevent complications due to swallowing or aspiration of the fragment [3]. Some of the metallic pieces of the surgical instrument could end up in a fibrous issue capsule and gain access to the adjacent spaces [12]. The case we report here is unusual in that the elevator tip was embedded in a subgingival caries and remained there asymptomatically for approximately six months.
At present there are a number of radiological explorations to identify metallic foreign objects. Cone beam computerized tomography (CBCT) is an excellent tool to locate metallic foreign objects [13]. However, a single periapical radiograph or using more than one radiograph to apply a tube-shift technique, may be sufficient. Whenever possible, simpler techniques should be first applied. In addition, occlusal radiography can also be employed as necessary. If an occlusal film is not available a periapical radiograph can be put on the surface of the tooth, or on the edentulous crest, and this may reveal an embedded foreign object. If with using these techniques, the foreign object is not detected, then other more sophisticated techniques such as CBCT should be applied. In our case, only conventional intraoral periapical radiographic imaging was available, due to limited hospital facilities, to inform the referral doctor about our findings.
The care of medical and surgical instruments has a decisive effect on their efficiency and durability. Damage due to breakage and scratches is usually caused by their being incorrectly stored and secured during sterilization, or being carelessly positioned in the treatment area thus favouring falls. Additionally, metal instruments used in clinical practice may be subjected to fatigue due to the effects of sterilization processes. Manufacturers recommend that all instruments made of metal should be checked regularly before packaging in order to diminish the risk of possible incidences [14].
The concept of honest and correct communication between physician and patient is a crucial issue. It is linked to the disclosure of adverse events and errors, a complex topic covering medical, psychological, legal, and ethical aspects [15-17]. Current socio-cultural trends, which condition the medical profession in a variety of ways, and a greater focus on the dignity and rights of the patient, have led to a growing tendency to fully inform patients with regard to their illness, progress, and therapy. Nevertheless, it is still common for doctors to choose to remain silent or manipulate the truth in some way, especially when the prognosis or accident is serious or negative and the patient seems to be having real difficulties accepting it [18]. As clinicians, this reported case should be helpful by reminding us of the importance of our ethical code with respect to the patients, especially when an adverse eventuality appears during a surgical procedure.
Conclusion
In dentistry is always advisable to use good quality and reliable brands for any instrument. Whenever any retention of a broken metal instrument is suspected an imaging radiological study will indicate its position and help avoid potential surgical complications. Preoperative and postoperative check-ups of instruments are also essential. Dental and oral surgeons should be particularly careful when metal instruments deployed with strong force are used in poorly visible areas such as the third molar region. If an unexpected accident takes place during a surgical procedure the patient should be informed in accordance with ethical codes, and suitable measures adopted to resolve the issue.
Consent
Written informed consent was obtained from the patient for publication of this Case Report and any accompanying images. A copy of the written consent is available for review by the Editor-in-Chief of this journal.
Abbreviations
- CBCT:
-
Cone beam computerized tomography
References
Kim YH, Cho SI, Do NY, Park JH. A case of pharyngeal injury in a patient with swallowed toothbrush: a case report. BMC Res Notes. 2014;7:788.
Brunet-Llobet LL, Miranda-Rius J, Lahor-Soler E, Mrina O, Nadal A. A gray-purple mass on the floor of the mouth: gigantic mucogingival pyogenic granuloma in a teenage patient. Open Dent J. 2014;8:125–8.
Ruprecht A, Ross A. Location of broken instrument fragments. J Can Dent Assoc. 1981;47:245.
Pierro VS, de Morais AP, Granado L, Maia LC. An unusual accident during a primary molar extraction. J Clin Pediatr Dent. 2010;34:193–5.
Yasuhara H, Fukatsu K, Komatsu T, Obayashi T, Saito Y, Uetera Y. Prevention of medical accidents caused by defective surgical instruments. Surgery. 2012;151:153–61.
Kluess D, Zenk K, Mittelmeier W. Reportable incidents with surgical instruments in orthopedic surgery. Orthopade. 2014;43:561–7.
Nelson CA. Material selection indices for design of surgical instruments with long tubular shafts. J Med Eng Technol. 2013;37:102–8.
Kandler HJ. The design and construction of dental elevators. J Dent. 1982;10:317–22.
Brophy T, Srodon PD, Briggs C, Barry P, Steatham J, Birch MJ. Quality of surgical instruments. Ann R Coll Surg Engl. 2006;88:390–3.
Bhutta MF. Fair trade for surgical instruments. BMJ. 2006;333:297–9.
Buttha MF. Quality of surgical instruments. Ann R Coll Surg Engl. 2007;89:190–2.
Balaji SM. Burried broken extraction instrument fragment. Ann Maxillofac Surg. 2013;3:93–4.
Eggers G, Welzel T, Mukhamadiev D, Wörtche R, Hassfeld S, Mühling J. X-ray-based volumetric imaging of foreign bodies: a comparison of computed tomography and digital volume tomography. J Oral Maxillofac Surg. 2007;65:1880–5.
Moore UJ, Fanibunda K, Gross MJ. The use of a metal detector for localisation of a metallic foreign body in the floor of the mouth. Br J Oral Maxillofac Surg. 1993;31:191–2.
Wu AW, McCay L, Levinson W, Iedema R, Wallace G, Boyle DJ, et al. Disclosing Adverse Events to Patients: International Norms and Trends. J Patient Saf. 2014. [Epub ahead of print].
Hammami MM, Al-Gaai EA, Al-Jawarneh Y, Amer H, Hammami MB, Eissa A, et al. Patients’ perceived purpose of clinical informed consent: Mill’s individual autonomy model is preferred. BMC Med Ethics. 2014;15:2.
Hammami MM, Al-Jawarneh Y, Hammami MB, Al Qadire M. Information disclosure in clinical informed consent: “reasonable” patient’s perception of norm in high-context communication culture. BMC Med Ethics. 2014;15:3.
Turillazzi E, Neri M. Medical error disclosure: from the therapeutic alliance to risk management: the vision of the new Italian code of medical ethics. BMC Med Ethics. 2014;15:57.
Acknowledgement
Clínics Associats, a well-established Professional Partnership in Terrassa (Barcelona), deserves a special mention for leading and supporting the KiliBarnaDental project at Soweto General Hospital in Arusha, United Republic of Tanzania. The authors want to express their gratitude to Miguel Torres Foundation, The Spanish Society of Periodontology (SEPA) and Straumann, SA (Spain) for their contributions. The authors also thank Dolors Icart-Rovira, nurse practitioner, for her assistance.
Author information
Authors and Affiliations
Corresponding author
Additional information
Competing interests
The authors declare that they have no competing interests.
Authors’ contributions
JMR and LBL were the consultants responsible for diagnosing this case report. JMR and LBL provided the information to ELS, OM and ARR who co-wrote the paper. All authors have read and approved the final version of this manuscript.
Rights and permissions
This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly credited. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.
About this article
Cite this article
Miranda-Rius, J., Brunet-Llobet, L., Lahor-Soler, E. et al. Dental root elevator embedded into a subgingival caries: a case report. BMC Res Notes 8, 60 (2015). https://doi.org/10.1186/s13104-015-1011-5
Received:
Accepted:
Published:
DOI: https://doi.org/10.1186/s13104-015-1011-5