Bacterial contamination and antimicrobial susceptibility patterns of intensive care units medical equipment and inanimate surfaces at Ayder Comprehensive Specialized Hospital, Mekelle, Northern Ethiopia
BMC Research Notes volume 12, Article number: 621 (2019)
Abstract
Objective
To determine bacterial contaminants and their antimicrobial susceptibility patterns from medical equipment and inanimate surfaces.
Results
Of 130 swabs, 115 (88.5%) swabs were culture positive, of which contaminated medical equipment and inanimate surfaces account 70 (83.3%) and 45 (97.8%), respectively. All the swabs collected from sphygmomanometer, bedside table, computer and computer standing tables were 100% contaminated with bacteria. From the culture-positive swabs, a total of 171 bacterial isolates were identified, out of which 117 (68.4%) and 54 (31.6%) isolates were gram-positive and gram-negative, respectively. Most isolates (82%) were resistant to ampicillin and 13%, 8.6%, and 14% was observed in ciprofloxacin, gentamicin, and tetracycline respectively. Multi-drug resistant was observed in Escherichia coli (72.7%) and Staphylococcus aureus (58.7%).
Introduction
Intensive care unit (ICU) acquired infections are global public health concern [1]. Bacterial contamination of the medical equipment and inanimate surfaces used in the ICU put ICU admitted patients (especially those with an underline disease, impaired immunity and with invasive procedures) at higher risk for hospital-acquired infection [2]. Non-critical medical equipment and inanimate surfaces have a capacity to harbor bacteria for a long period of time and can become into contact with patients and medical personnel during disease management [3,4,5,6].
The prevalence of ICU acquired infections in developed countries is indicated to be 5–10%, while their prevalence is shown to exceed 2–20 times higher in developing countries [7]. A high incidence (> 50%) of ICU acquired infections are reported as compared with other wards (10%) [1]. This might be due to invasive procedures, presence of debilitating patients, prolonged stays and emerging of multidrug-resistant bacteria [8,9,10].
Studies have reported that contaminated medical equipment and inanimate surfaces were highly associated with ICU acquired infections [11,12,13,14,15,16]. Although few studies were carried out in the Hospitals in Ethiopia, most studies were focusing on other wards and those performed in intensive care units were done on single or few medical equipment and inanimate surfaces [17,18,19]. Moreover, there is no study done in the study setting related to contamination of ICU medical equipment and inanimate surfaces. Therefore, this study aimed at determining bacterial contaminates and antimicrobial susceptibility patterns from medical equipment and inanimate surfaces in the ICUs of Ayder Comprehensive Specialized Hospital, Northern Ethiopia.
Main text
Materials and methods
Study area, design and period
A hospital-based cross-sectional study was conducted from October 2016 to June 2017 at Ayder Specialized Comprehensive Hospital, Mekelle, Northern Ethiopia. Mekelle city is situated 783Â km North of Addis Ababa, the capital of Ethiopia. It is the largest hospital serving for more than 4.1 million people of the region and the neighboring regions. The hospital has adult, neonatal, and pediatric ICU health services.
Sampling technique
All non-critical medical equipment and inanimate surfaces that had frequent contact with patients and/or healthcare providers during the study period were included. A total of 130 swabs from medical equipment and inanimate surfaces (55 swabs from Adult, 44 swabs from pediatric and 31 swabs from neonatal) were collected using purposive sampling technique.
Sample collection, handling, and transport
Sterile cotton-tipped applicator sticks, moistened with sterile normal saline, was used to collect swab specimen. Swab samples were taken using aseptic techniques with the presence of a spirit lamp. All swab samples were inserted in a separate sterile test tube, labeled and transported to the Medical Microbiology and Immunology Laboratory, Mekelle University, using icebox within 1Â h and processed immediately [20].
Laboratory processing
Bacterial culture and identification
Swabs were inoculated on to Blood agar, MacConkey agar and Mannitol salt agar (Oxoid Ltd, UK). The inoculated plates were incubated at 37 °C for 24 h. The inoculated plates were inspected after 24 h of incubation and bacterial isolates from culture-positive plates were identified at the species level by their colony morphology, gram-staining and biochemical characteristics (catalase, coagulase, urease, indole, oxidase, citrate utilization, glucose and lactose fermentation, gas and hydrogen sulfide production) [20].
Antimicrobial susceptibility testing
Antimicrobial susceptibility testing was performed using modified Kirby–Bauer disk diffusion according to the Clinical and Laboratory Standards Institute (CLSI), guideline [21]. Bacterial isolates were tested for commonly used antibiotics including: amoxicillin–clavulanic acid (AMC, 30 mcg), ampicillin (AMP, 10 mcg), ceftriaxone (CTR, 30 mcg), ciprofloxacin (CIP, 5 mcg), chloramphenicol (C, 30 mcg), gentamicin (GEN, 10 mcg) and tetracycline (TE, 30 mcg), cefoxitin (CX, 30 mcg), erythromycin (E, 15 mcg), penicillin G (P, 10 units), amikacin (AK, 30 mcg), cefotaxime (CTX, 30 mcg), nalidixic acid (NA, 30 mcg) (HIMEDIA, Company).
Data analysis
Data were entered into Microsoft Excel. Data were imported and analyzed using Statistical Package for Social Sciences (SPSS) software version 21.0. Descriptive statistics were computed and results were summarized by using tables.
Quality assurance
Aseptic techniques were used in all the steps of specimen collection and inoculation to minimize contamination. Specimens were collected in the presence of sprit lamp to prevent bacterial contamination from air. Reagents and antimicrobial discs were checked for expiry date. Sterility of culture media was carried out by incubating 5% of the prepared media prior to inoculation. Escherichia coli (ATCC 25922), Staphylococcus aureus (ATCC 25923) and Pseudomonas aeruginosa (ATCC 27853) reference strains were used to control the performance of culture media and antibiotic discs. All results were checked accuracy and cleaned after entering to SPSS.
Operational definition
Non-critical medical equipment is equipment used for the diagnosis and management of the patient in ICU such as sphygmomanometer, stethoscope, and thermometer.
The inanimate surface is a surface of the material used in ICU during providing patient care such as bedside tables, mattress, computers, and computer standing tables.
Results
Bacterial contamination of medical equipment and inanimate
A total of 130 swabs (84 from medical equipment and 46 from inanimate surfaces) were collected and inoculated to culture media and all the specimens were analysed. From the total analyzed swabs, bacterial growth was identified in 115 (88.5%) of the swab specimens. Bacterial contaminants were isolated in 70 (83.3%) and 45 (97.8%) from medical equipment and inanimate surfaces, respectively. Bacterial contaminations were statistically significant with the medical and inanimate surfaces bacterial contamination (P-value = 0.022, X2 = 14.843). All the swabs collected from sphygmomanometer, bedside table, computer and computer standing tables were 100% contaminated with bacteria. From the culture-positive swabs, a total of 171 bacterial isolates were identified, out of which 117 (68.4%) were gram-positive and 54 (31.6%) were gram-negative bacteria. From the total bacterial isolates in medical equipment and inanimate surfaces, coagulase-negative staphylococcus 53 (34.9%), Staphylococcus aureus 40 (26.3%), Citrobacter freundii 14 (9.2%) and Klebsiella pneumoniae 10 (8%) were the most commonly isolated bacteria. From the contaminated medical equipment samples, 32 (34.8%) CoNS, 23 (25%) S. aureus and 13 (14.1%) C. fraundii, whereas from the inanimate surface swab samples, 28 (35.4%) CoNS, 23 (29.1%) S. aureus and 5 (6.3%) K. pneumonia was the most commonly isolated bacteria (Table 1).
Antimicrobial susceptibility patterns of the bacterial isolates
The antimicrobial susceptibility tests of the isolates revealed varying ranges of resistance to the tested antimicrobial agents. More than 50% of Staphylococcus species were found to be resistant to penicillin G, erythromycin, amoxicillin–clavulanic acid, ampicillin, cefoxitin. Methicillin-resistant Staphylococcus aureus was observed in 34 (73.9%) of the isolates. More than 74% of gram-negative rods were resistant to amoxicillin–clavulanic acid, ampicillin, and nalidixic acid. Moreover, above 63% of E. coli isolates were resistant to amoxicillin–clavulanic acid, ceftriaxone, ampicillin, cefotaxime, and nalidixic acid (Table 2). Multidrug resistance was observed in 17 (28.3%) of CoNS, 27 (58.7%) of S. aureus, 3 (20%) of C. freundii and 8 (72.7%) of E. coli (Table 3).
Discussions
In this study, the overall bacterial contamination of noncritical medical equipment and inanimate surfaces were high. In addition to this, in this study, similar bacteria were isolated from swab samples taken from parts of some non-critical medical equipment and inanimate surfaces indicating cross-contamination. The predominant isolate was CoNS. With regard to the antimicrobial susceptibility pattern, the isolates showed a significant resistance pattern.
The findings of this study were compared with other similar studies. This finding is in line with the studies conducted from Ethiopia [17] and Asia [15, 22]. On the other hand, our finding is higher than another study from Ethiopia [19], sub-Saharan Africa [5, 16, 22, 23], Asia [13, 14, 24,25,26], Brazil [27], Italy [28] and UK [12]. On the contrary, this finding is lower than the other study from Brazil [29]. The difference might be attributed to the frequency of medical equipment and inanimate surfaces decontamination, types of decontaminants used, improper health professional practice, nature of the medical equipment and inanimate surfaces to harbor bacteria [12,13,14, 22, 24,25,26,27,28]. This might be through bacterial cross-contamination of surfaces. This was similarly observed in a study conducted in the UK [12].
The predominant bacteria isolated were CoNS. Bacterial contamination of stethoscope in this study was in-line with the study results reported from sub-Saharan Africa [5, 16, 23] and Asia [13, 24]. On the other hand, this finding is lower than the studies from Ethiopia [17], Asia [15, 22] and Brazil [29], but higher than studies from Ethiopia [19], Asia [14, 25, 26], Brazil [27, 29], Italy [28] and UK [12]. The bacterial contamination in a sphygmomanometer, was in-line with a study done in the UK [12] but higher than the studies from Nigeria [16], New York [30], Australia [31], France [32] and Iraq [33]. Considering inanimate surfaces, the contamination profile of bedside table was in-line with studies conducted in Brazil [29]. However, our findings from the computer table were higher than the studies done at sub-Saharan Africa [34] and Iraq [35]. Bacterial contamination of computer in this study was in-line with the studies conducted in Ethiopia [36], Egypt [37], and Pakistan [38] and higher than from sub-Saharan Africa [39] and Germany [40]. This might be attributed to the number of user’s and prolong the survival time of bacteria on plastic surfaces [39, 40], and the proximity of the equipment or surface to the patient [29, 35], the differences in sample size and method [16, 17].
With regard to the antimicrobial susceptibility pattern of the isolated bacteria to commonly prescribed antibiotics in our study area, the resistance of CoNS to erythromycin and penicillin G was higher than the studies conducted from Ethiopia [17, 19, 36], but lower than a study from sub-Saharan Africa [34]. In addition to this, in this study, S. aureus has shown resistant to ciprofloxacin and gentamycin, which is higher than the studies conducted from Ethiopia [17, 36], but, lower from sub-Saharan Africa [23, 34] and Asia [22, 26]. The characteristic multidrug resistance observed in this study is in line a study from Ethiopia [36], sub-Saharan Africa [34] and Asia [22]. On the other hand, lower multidrug resistance was observed in other studies from Ethiopia [12, 17], sub-Saharan Africa [23] and Asia [26]. The inconsistent drug-resistant and multidrug-resistant patterns observed might be due to variations in geographic areas, hospital environmental conditions, inappropriate administration of antimicrobial drugs, self-medication practice [17, 22, 23, 26, 32, 34, 36].
Conclusion
In this study, high bacterial contamination was observed in medical equipment and inanimate surfaces. CoNS, S. aureus, C. freundii, K. pneumonia, and E. coli were the most commonly isolated bacteria. Resistance to amoxicillin–clavulanic acid, ampicillin, and nalidixic acid was observed among the gram-negative bacteria. Resistant to penicillin G, erythromycin, amoxicillin–clavulanic acid, and ampicillin, cefoxitin was shown in gram-positive bacteria. Moreover, a significant multi-drug resistance was observed in the isolated bacteria. These calls for strengthening the existing infection prevention and antibiotic stewardship program with the application of strict follow up to minimize bacterial contamination of medical equipment and inanimate surfaces.
Limitation
This study limited to carry out the extended beta-lactam spectrum.
Availability of data and materials
Not applicable.
Abbreviations
- CoNS:
-
coagulase-negative staphylococcus
- CLSI:
-
Clinical and Laboratory Standards Institute
- ICU:
-
intensive care unit
- IP:
-
infection prevention
- MDR:
-
multidrug resistance
- SPSS:
-
Statistical Package for Social Sciences
References
Vincent J-L, Rello J, Marshall J, Silva E, Anzueto A, Martin CD, et al. International study of the prevalence and outcomes of infection in intensive care units. JAMA. 2009;302(21):2323–9.
Pekka Y, Ala-Kokko TI, Jouko L, Pasi H, Hannu S. Intensive care acquired infection is an independent risk factor for hospital mortality: a prospective cohort study. Crit Care. 2006;10(2):1–6.
Kathleen MA. Contamination and cross contamination on hospital surfaces and medical equipment. www.initiatives-patientsafety.org. Accessed June 2019.
Chidi O, Allison A, Wale A, Roger H, Clyde C, Ralph E-C, et al. Contamination of equipment in emergency settings: an exploratory study with targeted automated intervention. Ann Surg Innov R. 2009;3:8.
Chigozie JU, Chinwendu DN, Kingsley ON, Richard CN, Cletus DU, Nittita P-P. Stethoscope disinfection campaign in a Nigerian teaching hospital: results of a before-and-after study. J Infect Dev Ctries. 2014;8(1):86–93.
Nseir S, Blazejewski C, Lubret R, Wallet F, Courcol R, Durocher A. Risk of acquiring multidrug-resistant Gram-negative bacilli from prior room occupants in the intensive care unit. Clin Microbiol Infect. 2011;17(8):1201–8.
Pittet D, Allegranzi B, Storr J, Bagheri NS, Dziekan G, Donaldson L, et al. Infection control as a major World Health Organization priority for developing countries. J Hosp Infect. 2008;68:285–92.
Keshni N, Ilisapeci N, Sharan R, Kerri V, Stephen MG, Karen B. A descriptive study of nosocomial infections in an adult intensive care unit in Fiji: 2011–12. J Trop Med. 2014;1:5.
Bereket W, Hemalatha K, Getenet B, Wondwossen T, Solomon A, Zeynudin A, et al. Update on bacterial nosocomial infections. Eur Rev Med Pharmacol Sci. 2012;16:1039–44.
Sah MK, Mishra SK, Ohora H, Kirikae T, Sherchan JB, Rijal BP, et al. Nosocomial bacterial infection and antimicrobial resistant pattern in a Tertiary care Hospital in Nepal. J Inst Med. 2014;36(3):38–48.
Ding JG, Sun QF, Li KC. Retrospective analysis of nosocomial infections in the intensive care unit of a tertiary hospital in China during 2003 and 2007. BMC Infect Dis. 2009;9:115.
Whittington AM, Whitlow G, Hewson D, Thomas C, Brett SJ. Bacterial contamination of stethoscopes on the intensive care unit. Anaesthesia. 2009;64:620–4.
Ibrahim HK, Mehmet O, Isik DK, Yasemin Z, Eda S, Vedat D. The role of stethoscopes in the transmission of hospital infections. Afr J Biotechnol. 2011;10(30):5769–72.
Abrar-Nasser DM, Afra SA, Khulood YF, Khulood AA-J, Mawaddah HS, Waad SA, et al. The prevalence of stethoscopes contamination at Tertiary Center in AL Madinah Al monawwarah Saudi Arabia 2015. Int J Acad Sci Res. 2016;4(1):103–10.
Bharathi M, Lakshmi N, Kalyani CS, Padmaja IJ. The prevalence of bacterial contamination of stethoscope diaphragms: a cross-sectional study, among health care workers of a tertiary care hospital. Int J Med Microbiol. 2014;32(2):201–2.
Yushau M, Bukar A, Aliyu BS, Abdulkaree A. Bacterial contamination of some hospital equipment in Kano, Nigeria. Hamdard Medicus. 2012;55(3):1–4.
Teklu S, Getenet B, Tesfaye K, Tsegaye S. Bacterial contamination, bacterial profile and antimicrobial susceptibility pattern of isolates from stethoscopes at Jimma University Specialized Hospital. Ann Clin Microbiol Antimicrob. 2013;12:39.
Tewelde T, Yibrah B, Kibrom G-S. Microbial contamination of operating Theatre at Ayder Referral Hospital, Northern Ethiopia. IJPSR. 2015;6(10):1–4.
Dabsu R, Woldeamanuel Y, Asrat D. Otoscope and stethoscope: vehicles for microbial colonization at Tikur Anbessa Specialized Referral Hospital, Addis Ababa, Ethiopia. Ethiop J Health Dev. 2014;28(1):35–9.
Monica C. District laboratory practice in tropical countries, Part 2. 2nd ed. Cambridge: Cambridge University Press; 2006. p. 105–15.
Clinical and Laboratory Standards Institute (CLSI). Performance standards for antimicrobial susceptibility testing, vol. 36., Twenty-Six Informational Supplement, No. 1Wayne: CLSI; 2016.
Gurjeet S, Urhekar AD, Anahita VH, Neha S, Bhaskar D. Bacterial contamination of stethoscopes used by health care workers in a Tertiary Care Hospital in Navimumbai. Int J Pharm Biol Sci. 2013;3(1):186–93.
Chigozie JU, Annayo O, Patrick GO, Christian MO. Bacterial contamination of stethoscopes used by health workers: public health implications. J Infect Dev Ctries. 2010;4(7):436–41.
Youngster I, Berkovitch M, Heyman E, Lazarovitch Z, Goldman M. The stethoscope as a vector of infectious diseases in the paediatric division. Acta Paediatr. 2008;97:1253–5.
Alothman A, Bukhari A, Aljohani S, Muhanaa A. Should we recommend stethoscope disinfection before daily usage as an infection control rule? Open Infect Dis J. 2009;3:80–2.
Kuhu P, Ranadeep C, Amrita B, Arnab KS. Bacterial contamination and disinfection of stethoscopes: a knowledge gap among health care personnel of a Tertiary Care Hospital of Rural Bengal. IOSR J Dent Med Sci. 2015;14(7):44–9.
Vanessa MS, Elizandra O, Regina C, Fernando RG, Melo C. Microbiological analysis of inanimate surfaces in an intensive care unit and patient safety. Revista de Enfermagem Referência. 2014;3:45–52.
Gabriele M, Emma C, Daniele L, Sandra B, Elena A, Pietro M. Environmental contaminants in hospital settings and progress in disinfecting techniques. Biomed Res Int. 2013;1:8.
Quesia SD, Robert I, Adriana CO. Comparing resistant microorganisms isolated from patients and environment in an intensive care unit. Adv Infect Dis. 2014;4:30–5.
Havill LN, Havill LH, Elise M, Diane GD, John MB. Cleanliness of portable medical equipment disinfected by nursing staff. Am J Infect Control. 2011;39:602–4.
Carol D. Blood pressure cuffs and pulse oximeter sensors: a potential source of cross-contamination. Australas Emerg Nurs J. 2009;12:104–9.
De Gialluly C, Morange V, De Gialluly E, Loulergue J, Van der Mee N, Quentin R. Blood pressure cuff as a potential vector of pathogenic microorganisms: a prospective study in a teaching hospital. Infect Control Hosp Epidemiol. 2006;27:940–3.
Nazar EN, Ali TA, Saad LH. Bacterial contamination in the intensive care unit at Al-Imam Al-Hussein Hospital in Thiqar Province in Iraq. Glob J Health Sci. 2013;5(1):143–9.
Aminu M, Usman-Sani H, Usman MA. Characterization and determination of antibiotic susceptibility pattern of bacteria isolated from some fomites in a teaching hospital in northern Nigeria. Afr J Microbiol Res. 2014;8(8):814–8.
Carvalho KS, Melo MC, Melo GB, Gontijo-Filho PP. Hospital surface contamination in wards occupied by patients infected with MRSA or MSSA in a Brazilian university hospital. Rev Cienc Farm Basica APL. 2007;28(2):159–63.
Alemu A, Misganaw D, Wondimeneh Y. Bacterial profile and their antimicrobial susceptibility patterns of computer keyboards and mice at Gondar University Hospital, Northwest Ethiopia. Biomed Biotechnol. 2015;3(1):1–7.
Eltablawy SY, Elhifnawi HN. Microbial contamination of some computer keyboards and mice in National Center for Radiation Research and Technology (NCRRT). World Appl Sci J. 2009;6(2):162–7.
Kausar M, Nabiha N. Study of bacteria on computers. Int J Curr Microbiol Appl Sci. 2014;3(4):813–23.
Onochie CC, Anyim C, Alo MN, Onwa NC, Okonkwo EC, Afiukwa FN. Bacteriological examination of computer keyboards and mouse devices and their susceptibility patterns to disinfectants. Am J Microbiol. 2013;4(1):9–19.
Bernd H, Matthias B, Habil A, Lorenzo Q, Rainer R, Bernhard F, et al. Computer keyboard and mouse as a reservoir of pathogens in an intensive care unit. J Clin Monit. 2004;18:7–12.
Acknowledgements
We would like to thank the College of Health Sciences, Mekelle University and Medical Microbiology and Immunology Department for their material support during laboratory work. We would also like to extend our appreciation to staffs of Ayder Comprehensive Specialized Hospital for their technical and administrative support.
Funding
Not applicable.
Author information
Authors and Affiliations
Contributions
AD conceived the study. AD, AGK, HH, SN designed, carried out the coordination of the study. AD and MA carried out the statistical analysis. SN and AGK draft the manuscript. All authors read and approved the final manuscript.
Corresponding author
Ethics declarations
Ethics approval and consent to participate
The study protocol was evaluated and approved by the Research Ethics Review Committee (RERC) of College of Health Sciences; Mekelle University was obtained (REC, 0951/2017). Permission to collect data was obtained from administrative bodies of Ayder comprehensive specialized Hospital to conduct the study. The Ethics involves the permission to disseminate the findings of the study through the scientific workshop and publish in reputable journals.
Consent for publication
Not applicable.
Competing interests
The authors declare that they have no competing interests.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.
About this article
Cite this article
Darge, A., Kahsay, A.G., Hailekiros, H. et al. Bacterial contamination and antimicrobial susceptibility patterns of intensive care units medical equipment and inanimate surfaces at Ayder Comprehensive Specialized Hospital, Mekelle, Northern Ethiopia. BMC Res Notes 12, 621 (2019). https://doi.org/10.1186/s13104-019-4658-5
Received:
Accepted:
Published:
DOI: https://doi.org/10.1186/s13104-019-4658-5