Reproducibility of left ventricular blood flow kinetic energy measured by four-dimensional flow CMR

Abstract

Objectives

Four-dimensional flow CMR allows for a comprehensive assessment of the blood flow kinetic energy of the ventricles of the heart. In comparison to standard two-dimensional image acquisition, 4D flow CMR is felt to offer superior reproducibility, which is important when repeated examinations may be required. The objective was to evaluate the inter-observer and intra-observer reproducibility of blood flow kinetic energy assessment using 4D flow of the left ventricle in 20 healthy volunteers across two centres in the United Kingdom and the Netherlands.

Data description

This dataset contains 4D flow CMR blood flow kinetic energy data for 20 healthy volunteers with no known cardiovascular disease. Presented is kinetic energy data for the entire cardiac cycle (global), the systolic and diastolic components, in addition to blood flow kinetic energy for both early and late diastolic filling. This data is available for reuse and would be valuable in supporting other research, such as allowing for larger sample sizes with more statistical power for further analysis of these variables.

Objective

Quantitative cardiovascular magnetic resonance (CMR) imaging can provide a wealth of information to distinguish health from disease [1]. Four-dimensional flow (4D flow) CMR allows for a comprehensive assessment of the blood flow kinetic energy (KE) of the left ventricle (LV) [2]. Assessment of LV KE by 4D flow CMR is thought to offer superior reproducibility compared to standard two-dimensional phase contrast acquisition [3]. In our publication, for which this dataset corresponds to [4], we set out to answer questions we felt to be fundamental in evaluating the diagnostic utility of 4D flow CMR. First, we determined the normal ranges of LV KE values across the spectrum of age, which is an important step in differentiating between healthy and diseased states. Second, given the established association between standard 2D parameters of diastolic function and the myocardial stiffening accompanying the ageing process [5], we investigated the association of LV blood flow KE using 4D flow CMR with 2D mitral inflow and myocardial tissue velocities by CMR. Third, we assessed inter-observer and intra-observer reproducibility of 4D flow CMR LV KE assessments. Reproducibility is a prerequisite for any investigatory technique where repeated examinations may be required.

Reproducibility in cardiac MRI research is a pervasive issue across the field, which impacts the translational pathway [6]. Given the relative infancy of research within 4D flow CMR and the accepted potential for this imaging technique to propel our understanding of cardiovascular disease, it becomes of utmost importance to establish repeatability of the various aspects of the techniques, ideally across multiple centres. We hope the publication of our data will offer further opportunities to generate research which contributes to our growing understanding of the diagnostic utility of 4D flow CMR techniques. Our data will be of particular use to research groups incorporating reproducibility analysis within validation studies, and is available for reuse and assimilation within quantitative syntheses.

Data description

The dataset includes the LV blood flow KE parameters of twenty healthy adult volunteers void of any known cardiovascular disease, collected from two sites in the United Kingdom (Leeds) and the Netherlands (Leiden). CMR was performed on a dedicated cardiovascular 1.5 Tesla Phillips Ingenia system equipped with a 28-channel coil and Phillips dStream digital broadband MR architecture technology. The data was acquired following analysis of CMR images using dedicated 4D flow research software (Mass; Version 2017EXP, Leiden University Medical Center, Leiden, The Netherlands). After acquiring images, advanced 4D flow CMR techniques were applied to determine the kinetic energy (KE) for each phase of the cardiac cycle, following which all KE parameters were normalized to the LV end-diastolic volume, and reported in standard units (μJ/ml).

The dataset is composed of two parts. First, inter-observer reproducibility. This refers to two separate operators analysing the twenty cases independent of each other. Second, intra-observer reproducibility. This refers to the same operator analysing the same ten cases on two separate occasions, with a three-month interval between. The data presented is the endpoint of a range of software-delivered processing of the images, which is beyond the scope of this report. Further detail on the 4D LV KE mapping can be found within our primary publication on this topic [4]. The reproducibility aspect presented within the dataset is the result of assessor contouring of the short-axis LV cine volumetric stacks prior to automation and manual readjustment. The number of cases for both inter- and intra-observer reproducibility exceeds the minimum number required as per power calculations based upon previous KE studies. This dataset should serve as a valuable benchmark for other research groups in establishing the required sample sizes to ensure adequate statistical power within CMR trials, based on the reproducibility outcomes identified from this cohort. This dataset, when analysed using the appropriate correlation statistical tests, demonstrated excellent concordance in global KE parameters for both inter and intra-observer reproducibility analyses.

For each case within the two reproducibility analyses, we include data for: global LV KE (an average KE of the LV flow for the complete cardiac cycle), systolic KE (the average KE of the LV flow during systole), diastolic KE (the average KE of the LV flow during diastole), peak E-wave KE (the peak KE of the LV flow during early diastolic filling), and peak A-wave KE (the peak KE of the LV flow during late diastolic filling).

Limitations

The temporal resolution of 4D flow is 40 ms (ms), which may impact the precision of KE assessments. This is an issue characteristic of 4D flow analyses, rather than a limitation unique to our dataset. 4D flow was acquired during free-breathing, for which issues relating to heart rate may have impacted on the time-varying flow characteristics, which could not be corrected for. There are no other immediate limitations of this dataset.

Availability of data and materials

The data described in this Data note can be freely and openly accessed on Harvard Dataverse under https://doi.org/10.7910/DVN/JMTN02. Please see Table

Table 1 Overview of data files/data sets

1 and references for details and links to the data [7].

Abbreviations

2D:

2 Dimensional

4D-Flow:

4-Dimensional flow

CMR:

Cardiovascular magnetic resonance

KE:

Kinetic energy

LV:

Left ventricle

References

  1. Kawel-Boehm N, Hetzel SJ, Ambale-Venkatesh B, Captur G, Francois CJ, Jerosch-Herold M, et al. Reference ranges (“normal values”) for cardiovascular magnetic resonance (CMR) in adults and children: 2020 update. J Cardiovasc Magn Reson. 2020;22(1):87.

    Article  Google Scholar 

  2. Dyverfeldt P, Bissell M, Barker AJ, Bolger AF, Carlhäll C-J, Ebbers T, et al. 4D flow cardiovascular magnetic resonance consensus statement. J Cardiovasc Magn Reson. 2015;17(1):72.

    Article  Google Scholar 

  3. Stoll VM, Loudon M, Eriksson J, Bissell MM, Dyverfeldt P, Ebbers T, et al. Test-retest variability of left ventricular 4D flow cardiovascular magnetic resonance measurements in healthy subjects. J Cardiovasc Magn Reson. 2018;20(1):15.

    Article  Google Scholar 

  4. Crandon S, Westenberg JJM, Swoboda PP, Fent GJ, Foley JRJ, Chew PG, et al. Impact of age and diastolic function on novel, 4D flow CMR biomarkers of left ventricular blood flow kinetic energy. Sci Rep. 2018;8(1):14436.

    Article  Google Scholar 

  5. Wong J, Chabiniok R, deVecchi A, Dedieu N, Sammut E, Schaeffter T, et al. Age-related changes in intraventricular kinetic energy: a physiological or pathological adaptation? Am J Physiol-Heart Circ Physiol. 2016;310(6):H747-55.

    Article  Google Scholar 

  6. Bøtker HE, Hausenloy D, Andreadou I, Antonucci S, Boengler K, Davidson SM, et al. Practical guidelines for rigor and reproducibility in preclinical and clinical studies on cardioprotection. Basic Res Cardiol. 2018;113(5):39.

    Article  Google Scholar 

  7. Grafton-Clarke C, Garg P. Reproducibility of left ventricular blood flow kinetic energy measured by four-dimensional flow CMR. V4 ed. 2021. Harv Dataverse. https://doi.org/10.7910/DVN/JMTN02.

Download references

Acknowledgements

We thank radiographers from the primary sites for their assistance in recruiting and collecting data for this study.

Funding

This work was supported by the British Heart Foundation [FS/10/62/28409 to S.P.], Dutch ZonMw [Project Number: 104003001 to J.W]. This research was funded in whole, or in part, by the Wellcome Trust [220703/Z/20/Z; 215799/Z/19/Z]. For the purpose of open access, the author has applied a CC BY public copyright licence to any Author Accepted Manuscript version arising from this submission. 

Author information

Authors and Affiliations

Authors

Contributions

CG-C and PG collated the datasets, drafted the manuscript, and submitted the data to the repository. All authors approve the final manuscript. The contributions of each author towards the initial published study is as follows: SC was involved in study conception. PPS was involved in data acquisition, CMR analysis and provided critical input. JJMW was involved in acquisition and analysis, as well as data interpretation., RJvdG assisted with the conception and design of the study, in addition to developing the methods for the study and proving the core lab for analysis. JPG was involved in data collection, data interpretation and writing the manuscript. AJS was involved in study conception, data collection and writing the manuscript. VSS and SP were involved in study design, data collection, data interpretation, supervision and writing the manuscript. PG contributed significantly to the conception, analysis, statistics and writing of this study. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Ciaran Grafton-Clarke.

Ethics declarations

Ethics approval and consent to participate

The study protocol was approved by the National Research Ethics Service (12/YH/0169) in the UK and Institutional Medical Ethics Committee (P11.136) in Leiden, Netherlands). The study complied with the Declaration of Helsinki and all volunteers gave written informed consent.

Consent for publication

Written informed consent provided by all participants for publication.

Competing interests

The authors declare that they have no competing interests.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Grafton-Clarke, C., Crandon, S., Westenberg, J.J.M. et al. Reproducibility of left ventricular blood flow kinetic energy measured by four-dimensional flow CMR. BMC Res Notes 14, 289 (2021). https://doi.org/10.1186/s13104-021-05697-3