Carotid artery velocity time integral and corrected flow time measured by a wearable Doppler ultrasound detect stroke volume rise from simulated hemorrhage to transfusion

Abstract

Objective

Doppler ultrasonography of the common carotid artery is used to infer stroke volume change and a wearable Doppler ultrasound has been designed to improve this workflow. Previously, in a human model of hemorrhage and resuscitation comprising approximately 50,000 cardiac cycles, we found a strong, linear correlation between changing stroke volume, and measures from the carotid Doppler signal, however, optimal Doppler thresholds for detecting a 10% stroke volume change were not reported. In this Research Note, we present these thresholds, their sensitivities, specificities and areas under their receiver operator curves (AUROC).

Results

Augmentation of carotid artery maximum velocity time integral and corrected flowtime by 18% and 4%, respectively, accurately captured 10% stroke volume rise. The sensitivity and specificity for these thresholds were identical at 89% and 100%. These data are similar to previous investigations in healthy volunteers monitored by the wearable ultrasound.

Introduction

Inferring change in stroke volume (SV) is the bedrock of functional hemodynamic monitoring [1, 2]. Yet measuring SV change (SV) is clinically challenging, so surrogates like the common carotid artery Doppler pulse have been proposed [3, 4] with some conflicting data [5]. We contend that statistically-inadequate beat sampling coupled with variation introduced by the respiratory cycle are important arbiters of inconsistent clinical research [6]. To rectify the shortcomings of handheld examinations, we developed a wireless, wearable Doppler ultrasound [7, 8]. This device adheres to the neck, maintains a constant insonation angle and accurately measures beat-to-beat changes across multiple cardiorespiratory cycles [6, 7].

In early, proof-of-concept investigations, the carotid artery maximum velocity time integral (VTI) and corrected flow time (ccFT) accurately identified a + 10% SV with thresholds of + 15% and + 2–4%, respectively [9, 10]. More recently, in a human model of hemorrhage and resuscitation comprising approximately 50,000 cardiac cycles, we found a strong, linear correlation between SV, and both changing maximum carotid VTI (VTI) and ccFT (ccFT) [8]. While 10% SV associated with 18% VTI and 4.3% ccFT, the optimal VTI and ccFT thresholds for detecting a + 10% SV were not investigated. In this Research Note, we report these thresholds, their sensitivities, specificities and areas under their receiver operator curves (AUROC).

Main text

Methods

Clinical setting

11 healthy, adult volunteers with no cardiovascular history and who provided written, informed consent were recruited. The study was approved by the Research Ethics Board of the Mayo Clinic (IRB number 19–010,136).

Adherent Doppler system

The U.S. Food and Drug Administration (FDA) cleared, 4 MHz Doppler ultrasound (Fig. 1) (Flosonics Medical, Sudbury, Canada) was placed and the ccFT, maximum and power-weighted, i.e., centroid, VTIs were captured [6, 8,9,10].

Fig. 1
figure 1

Picture of wireless, wearable Doppler ultrasound device

Lower body negative pressure (LBNP) and stroke volume

As previously reported [6], all subjects underwent a 7-stage protocol in duplicate; a non-invasive SV monitor, the Nexfin® (Edwards Lifesciences, Irvine, California), was synchronized with the Doppler monitor for each cardiac cycle.

Statistical analysis

All data was averaged over 10-s windows. SV and carotid Doppler were referenced to resting baseline to model hemorrhage. Cardiac cycles with artifact or during LBNP stage transition were excluded, as described [8]. SV change from the lowest-tolerated LBNP stage back to atmosphere modeled rapid blood transfusion, data from this transition was included.

Each 10-s data point was dichotomized as ≥  + 10% SV or <  + 10% SV based on the non-invasive SV monitor. An equal number of data points were randomly sampled from negative pressure stages without replacement to match the sample size of the positive cases (≥ + 10% SV) for 1000 iterations. Within each iteration, the optimal threshold for each metric was selected using the Youden index. Averages of optimal thresholds and the corresponding sensitivities and specificities were calculated from all iterations and reported as the best final estimations. AUROC was calculated using the subsamples that produced the same optimal thresholds as the final estimation after 1000 iterations.

Results

The median and interquartile ranges for age and BMI were 27 (23—38) years and 23 (21.9–25.6) kg/m2, respectively; 39% were female. Vital signs and their change are previously reported [6]. Figure 2 A-D illustrates the progression of mean arterial pressure (MAP), SV, maximum VTI, ccFT and SV for all protocols. The optimal thresholds of % VTI, % ccFT and their calculation are shown in Fig. 2 E and F. Both + 18% VTI and + 4% ccFT were 89% sensitive and 100% specific at detecting ≥  + 10% SV∆. The areas under their receiver operator curves were 0.97 and 0.98, respectively. Though not illustrated, a + 20% change in the centroid VTI had identical diagnostic characteristics.

Fig. 2
figure 2

Hemodynamic data captured during lower body negative pressure (LBNP) and release. Measures from AD are synchronously captured. Each faint line represents a single protocol, while the emboldened line represents the average of all protocols. A Stroke volume (SV) percent change during progressively severe LBNP (i.e., hemorrhage model) and release of LBNP (i.e., rapid transfusion model). B Mean arterial pressure (MAP) percent change. C velocity time integral (VTI) from the wearable Doppler percent change. D corrected flow time (ccFT) percent change. E The optimal carotid artery maximum VTI threshold for distinguishing ≥  + 10% SV. Each data point represents a 10-s average. Prior to subsampling, there were 3596 data points categorized as <  + 10% SV and 598 data points categorized as ≥  + 10% SV. The data categorized as <  + 10% SV were randomly subsampled, iteratively 1000 times, to 598 data points (see methods). The sensitivity of maximum VTI is 532/598 = 89% and specificity is 598/598 = 100%. F The optimal ccFT threshold for distinguishing + 10% SV. Each data point represents a 10-s average. The sensitivity is 532/598 = 89% and specificity is 598/598 = 100%

Discussion

In this model of hemorrhage and transfusion, the % VTI and % ccFT thresholds for detecting a ≥  + 10% SV are consistent with earlier proof-of-concept data using a separate paradigm, also in healthy volunteers [9, 10]. With simulated blood transfusion, the 4% ccFT approximates an absolute ccFT of 10 ms, comparable to the 7 ms threshold observed in critically-ill patients with undifferentiated shock [4]. The + 18% VTI threshold in this investigation is consistent with the slope of the SV- VTI regression line, described previously [8]. Curiously, the + 20% threshold for the power-weighted (i.e., centroid) VTI matches the threshold of changing carotid blood flow in septic patients [3]. Assuming static vessel diameter, changing centroid velocity best approximates volumetric flow change [8].

Limitations

Our findings have a number of limitations deserving of elaboration. First, from a broad hemodynamic perspective, common carotid blood flow relates to cardiac output (i.e., total blood flow) by the following, general relationship:

$$carotid \ blood \ flow \ = \ cardiac \ output \ x \ \frac{{Z_{whole\, body} }}{{Z_{carotid\, artery} }}$$

where Z is the impedance to flow for the listed vascular beds. Thus, using the carotid artery as a surrogate for SV implies a constant ratio of whole body-to-downstream carotid impedance. That HR increased and MAP remained constant during the protocol, Zwhole body (e.g., total vascular resistance) likely rose relative to Zcarotid artery, especially if downstream internal carotid artery impedance fell by auto-regulation. This could partly explain the relatively large change in carotid VTI (i.e., 18%) observed to detect a 10% SV. Second, though we did not measure carotid diameter to calculate carotid artery flow, the normal carotid artery pressure-diameter relationship is relatively flat above a MAP of 80 mmHg [11, 12]. More specifically, the carotid diameter changes by only 0.2 mm when MAP rises from 80 to 110 mmHg [12]; the average MAP for our subjects during the lowest tolerated LBNP stage was 99 mmHg. Nevertheless, even a 0.2 mm diameter change in a 7 mm carotid artery affects total flow by ± 6% [13]. On the other hand, in hypotensive patients, diameter assessment is likely more important; for example, measuring the diameter of the descending aorta improves the sensitivity of Doppler ultrasonographic flow assessment by approximately one-third in the critically-ill [14]. Third, we chose non-invasive pulse contour analysis as a gold standard because it is continuous, user-independent and accurately trends % SV, which is important for functional hemodynamic monitoring. However, absolute SV measures by non-invasive pulse contour analysis are less adequate, especially in the critically-ill [15, 16]. Still, we believe that the pattern of SV that we observed is valid because it replicates the SV measured in other LBNP investigations [17, 18] and is consistent with the LBNP SV measured by other gold standards including left ventricular outflow tract VTI [19], bioimpedance [20] and bioreactance [21].

In summary, in this large data set of continuously-monitored SV synchronous with carotid Doppler, multiple measures from a wearable ultrasound patch identified + 10% SV with high accuracy.

Data availability

The datasets used and/or analysed during the current study are available from the corresponding author on reasonable request.

Abbreviations

AUROC:

Area under the receiver operator curve

SV:

Stroke volume

SV :

Stroke volume change

VTI:

Velocity time integral

ccFT:

Carotid corrected flow time

VTI :

Velocity time integral change

ccFT :

Carotid corrected flow time change

FDA:

Food and Drug Administration

LBNP:

Lower body negative pressure

BMI:

Body mass index

MAP:

Mean arterial pressure

ms:

Milliseconds

Zwhole body :

Whole body impedance

Zcarotidy artery :

Carotid artery impedance

mmHg:

Millimeters of mercury

References

  1. Kenny J-ES, Barjaktarevic I. Letter to the editor: stroke volume is the key measure of fluid responsiveness. Crit Care. 2021;25(1):104.

    Article  Google Scholar 

  2. Kenny J-ÉS. Functional hemodynamic monitoring with a wireless ultrasound patch. J Cardiothorac Vasc Anesth. 2021;35(5):1509–15.

    Article  Google Scholar 

  3. Marik PE, Levitov A, Young A, Andrews L. The use of bioreactance and carotid Doppler to determine volume responsiveness and blood flow redistribution following passive leg raising in hemodynamically unstable patients. Chest. 2013;143(2):364–70.

    Article  Google Scholar 

  4. Barjaktarevic I, Toppen WE, Hu S, Montoya EA, Ong S, Buhr R, et al. Ultrasound assessment of the change in carotid corrected flow time in fluid responsiveness in undifferentiated shock. Crit Care Med. 2018;11:1040–6.

    Article  Google Scholar 

  5. Beier L, Davis J, Esener D, Grant C, Fields JM. Carotid ultrasound to predict fluid responsiveness: a systematic review. J Ultrasound Med. 2020;39(10):1965–76.

    Article  Google Scholar 

  6. Kenny J-ÉS, Barjaktarevic I, Mackenzie DC, Elfarnawany M, Math ZYB, Eibl AM, et al. Carotid Doppler measurement variability in functional hemodynamic monitoring: an analysis of 17,822 cardiac cycles. Crit Care Explor. 2021;3(6):e0439.

    Article  Google Scholar 

  7. Kenny J-ÉS, Munding CE, Eibl JK, Eibl AM, Long BF, Boyes A, et al. A novel, hands-free ultrasound patch for continuous monitoring of quantitative Doppler in the carotid artery. Sci Rep. 2021;11(1):1–11.

    Article  Google Scholar 

  8. Kenny J-ÉS, Barjaktarevic I, Mackenzie DC, Elfarnawany M, Yang Z, Eibl AM, et al. Carotid Doppler ultrasonography correlates with stroke volume in a human model of hypovolaemia and resuscitation: analysis of 48570 cardiac cycles. Br J Anaesth. 2021;127(2):e60–3.

    Article  Google Scholar 

  9. Kenny J-ÉS, Barjaktarevic I, Mackenzie DC, Eibl AM, Parrotta M, Long BF, et al. Diagnostic characteristics of 11 formulae for calculating corrected flow time as measured by a wearable Doppler patch. Intensiv Care Med Exp. 2020;8(1):1–11.

    Article  Google Scholar 

  10. Kenny JÉS, Barjaktarevic I, Eibl AM, Parrotta M, Long BF, Eibl JK, et al. A carotid Doppler patch accurately tracks stroke volume changes during a preload-modifying maneuver in healthy volunteers. Crit Care Explor. 2020;2(1):e0072.

    Article  Google Scholar 

  11. Magder S. The meaning of blood pressure. Crit Care. 2018;22(1):257.

    Article  CAS  Google Scholar 

  12. Hansen F, Mangell P, Sonesson B, Länne T. Diameter and compliance in the human common carotid artery—variations with age and sex. Ultrasound Med Biol. 1995;21(1):1–9.

    Article  CAS  Google Scholar 

  13. Taylor KJW, Burns PN, Wells PNT. Clinical Applications of Doppler Ultrasound, Ch 4. 2nd ed. Raven Press; 1995. p. 94.

    Google Scholar 

  14. Monnet X, Chemla D, Osman D, Anguel N, Richard C, Pinsky MR, et al. Measuring aortic diameter improves accuracy of esophageal Doppler in assessing fluid responsiveness. Crit Care Med. 2007;35(2):477–82.

    Article  Google Scholar 

  15. Ameloot K, Palmers P-J, Malbrain ML. The accuracy of noninvasive cardiac output and pressure measurements with finger cuff: a concise review. Curr Opin Crit Care. 2015;21(3):232–9.

    Article  Google Scholar 

  16. Saugel B, Hoppe P, Nicklas JY, Kouz K, Körner A, Hempel JC, et al. Continuous noninvasive pulse wave analysis using finger cuff technologies for arterial blood pressure and cardiac output monitoring in perioperative and intensive care medicine: a systematic review and meta-analysis. Br J Anaesth. 2020;125(1):25–37.

    Article  Google Scholar 

  17. Convertino VA, Ludwig DA, Cooke WH. Stroke volume and sympathetic responses to lower-body negative pressure reveal new insight into circulatory shock in humans. Auton Neurosci. 2004;111(2):127–34.

    Article  Google Scholar 

  18. Skytioti M, Søvik S, Elstad M. Respiratory pump maintains cardiac stroke volume during hypovolemia in young, healthy volunteers. J Appl Physiol. 2018;124(5):1319–25.

    Article  CAS  Google Scholar 

  19. Vettorello M, Sher S, Santambrogio S, Calini A, Tardini F, Lippi M, et al. Validation of a novel index of hemorrhage using a lower body negative pressure shock model. Minerva Anestesiol. 2016;82(8):839–49.

    PubMed  Google Scholar 

  20. Convertino VA, Cooke WH, Holcomb JB. Arterial pulse pressure and its association with reduced stroke volume during progressive central hypovolemia. J Trauma Acute Care Surg. 2006;61(3):629–34.

    Article  Google Scholar 

  21. Bartels SA, Bezemer R, de Vries FJW, Milstein DM, Lima A, Cherpanath TG, et al. Multi-site and multi-depth near-infrared spectroscopy in a model of simulated (central) hypovolemia: lower body negative pressure. Intensive Care Med. 2011;37(4):671–7.

    Article  Google Scholar 

Download references

Acknowledgements

We wish to acknowledge Briana Ziegler, Alex Carlson and Brad Cierzan for their technical assistance, recruitment, scheduling and data management. Additionally, Matt Myers for his assistance in data analysis.

Funding

This work was supported by the Office of the Assistant Secretary of Defense for Health Affairs, through the Congressionally Directed Medical Research Program under Award No. W81XWH1910591. Opinions, interpretations, conclusions and recommendations are those of the authors and are not necessarily endorsed by the Department of Defense.

Author information

Authors and Affiliations

Authors

Contributions

JESK: conception, study design, analysis and interpretation, drafting. IB: analysis and interpretation, critical revisions. DCM: analysis and interpretation, critical revisions. ME: data acquisition, analysis and interpretation, critical revisions. ZY: data acquisition, analysis and interpretation, critical revisions. AME: conception, data acquisition, analysis and interpretation, critical revisions. JKE: conception, data acquisition, analysis and interpretation, critical revisions. CHK: data acquisition, study design, analysis and interpretation, critical revisions. BDJ: data acquisition, study design, analysis and interpretation, critical revisions. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Jon-Émile S. Kenny.

Ethics declarations

Ethics approval and consent to participate

The study was approved by the Research Ethics Board of the Mayo Clinic (IRB number 19-010136); written and informed consent was obtained for all subjects.

Consent for publication

Not applicable.

Competing interests

JESK, ME, ZY, AME, JKE are employees of Flosonics Medical, a start-up building the wearable ultrasound, IB reports consulting fees for GE Healthcare, DCM, CHK and BDJ report no conflicts.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Kenny, JÉ.S., Barjaktarevic, I., Mackenzie, D.C. et al. Carotid artery velocity time integral and corrected flow time measured by a wearable Doppler ultrasound detect stroke volume rise from simulated hemorrhage to transfusion. BMC Res Notes 15, 7 (2022). https://doi.org/10.1186/s13104-021-05896-y

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s13104-021-05896-y

Keywords