CPAP pressure and flow data at 2 positive pressure levels and multiple controlled breathing rates from a trial of 30 adults

Abstract

Objectives

A unique dataset of airway flow/pressure from healthy subjects on Continuous Positive Airway Pressure (CPAP) ventilation was collected. This data can be used to develop or validate models of pulmonary mechanics, and/or to develop methods to identify patient-specific parameters which cannot be measured non-invasively, during CPAP therapy. These models and values, particularly if available breath-to-breath in real-time, could assist clinicians in the prescription or optimisation of CPAP therapy, including optimising PEEP settings.

Data description

Data was obtained from 30 subjects for model-based identification of patient-specific lung mechanics using a specially designed venturi sensor system comprising an array of differential and gauge pressure sensors. Relevant medical information was collected using a questionnaire, including: sex; age; weight; height; smoking history; and history of asthma. Subjects were tasked with breathing at five different rates (including passive), matched to an online pacing sound and video, at two different levels of PEEP (4 and 7 cmH2O) for between 50 and 180 s. Each data set comprises ~ 17 breaths of data, including rest periods between breathing rates and CPAP levels.

Objective

CPAP is a form of non-invasive ventilation used in both hospital and home settings to treat respiratory conditions and wean patients off invasive ventilation [1, 2]. Positive end-expiratory pressure (PEEP) is the key CPAP setting. Insufficient PEEP can result in under-oxygenation or airway collapse [3, 4]. Conversely, excessive PEEP causes pulmonary barotrauma [3, 4]. Rupture of tissue (emphysema) and air trapping can lead to pneumothorax [3, 4]. Excessive PEEP can also cause increases in intracranial pressure due to intrathoracic pressure increases, as well as increased fluid retention [3, 5]. Hence, the setting is critical care and outcomes, as well as carrying risk with sub-optimal settings.

Protocols for PEEP settings vary significantly and are predominantly based on clinical judgment of the comparative foreseeable risks of hyperinflation and under oxygenation [6,7,8]. Thus, variability of results and inequity of care can lead to issues of unconscious bias [9]. In CPAP ventilation, the “CPAP Titration Protocol” [6, 10, 11] is commonly used to titrate PEEP based on patient blood oxygenation or symptoms of airway collapse.

The described dataset [12] has been used to develop a method of extrapolating patient breathing effort and ventilator unloading non-invasively in CPAP therapy [13]. Objective, model-based quantification of ventilator unloading in CPAP creates novel real-time monitoring. This metric provides further, novel feedback to clinicians on the efficacy of the CPAP therapy at given PEEP settings with the potential to guide clinical decision support and care. More specifically, the goal of CPAP is to support the work of breathing, so a metric quantifying this support in real-time would offer potential clinical value. Future work will involve a larger trial with a secondary measurement method to further validate the model using a larger subject pool.

Data description

Data was collected using a customised sensor system consisting of a Venturi tube on either side of the CPAP expiration hole (Fig. 1). The dual-Venturi apparatus (Fig. 2) with a central expiration hole was 3D printed to fit standard CPAP masks (FreeMotion RT041, Fisher and Paykel Healthcare, New Zealand), with an inner diameter of 15 mm (D1) and a venturi diameter of 12 mm (D2). The data set [12] thus captures flow delivered from the CPAP machine and flow delivered to the CPAP mask.

Differential pressure sensors connected in both directions over the Venturi restrictions enable flow measurement (Fig. 3), and static pressure was also measured at both locations. As flow sensors were uni-directional, both inspiratory and expiratory flow were measured at each Venturi tube separately, using two sets of differential pressure (flow) sensors.

Data was recorded at a frequency of 83.33 Hz (resampled to 100 Hz) from the sensors via a central Arduino unit (Nano V3, Baite Electronics, China) in analogue-to-digital converter (ADC) counts, which was serially interfaced with a laptop for data acquisition. The sensor system design approach is open access with details in [14]. Sensors were calibrated against a known flow profile obtained from a hospital-grade mechanical ventilator (PB840, Puritan Bennet, USA, using CURESoft [15]).

The trial protocol consisted of a sequence of breaths at resting, 6, 9, 30, and 60 breaths per minute, repeated at nominal PEEP levels of 4 and 7 cmH2O. At each cued breathing rate, 17 breaths were cued for inhalation and exhalation by an audio-visual pacer.

Raw data was processed in Matlab (Matlab 2020a, The Mathworks Inc, Natick, MA, USA), with pressure values calculated based on datasheet information for each of the sensors [16, 17]. The flow was calculated from differential pressure (∆P = P1− P2) across each Venturi as a function of its relationship to the ratio decrease in cross-sectional tube area at the venturi restriction (\(\frac{{\mathrm{A}}_{1}}{{\mathrm{A}}_{2}}\)) by derivation of Bernoulli and continuity equations (Eq. 1), yielding:

$$ {\text{Q}} = {\text{c}}_{{\text{d}}} {\text{A}}_{2} \sqrt {\frac{{2\left( {\Delta {\text{P}}} \right)}}{{{\uprho }(1 - \left( {\frac{{{\text{A}}_{1} }}{{{\text{A}}_{2} }}} \right)^{2} }}} $$
(1)

Given a drag coefficient (cd) of 0.97 and the density of air (ρ) as 1.225 kg/m3, and assuming total flow into the Venturi restriction equals the total flow out of the same Venturi.

The inspiratory and expiratory flow were merged into a single multidirectional flow dataset for each Venturi by a minimum inspiration length and volume based on the cued breath rate. The final data set thus has flow in [L/s] and pressure in [cmH2O] at two locations: at the entry to the CPAP mask (Q1, P2) and downstream of the expiration hole (Q2, P2) as illustrated in Fig. 1.

Data (Table 1) [12] is collated into Excel files and organised into raw (“RAW_CSV_Data_Files”) and processed data (“Processed_CSV_Data_Files”) folders. Both Folders contain subfolders of the 10 PEEP and breath rate combinations, which in turn contain an excel file per subject. Raw data is arranged in columns of time, pressure, and flow (at both venturis and in both directions). Processed data is arranged in columns of time, pressure, and flow (at both venturis). Corresponding subject demographic data is provided as an excel file organised in columns by subject, sex, age, weight, height, BMI, smoking history, and vaping history. A “README” text document is also included outlining how the data a stored.

Table 1 Overview of data files/data sets

Limitations

Data with significant observed sensor error was removed during processing. Significant error was considered when no distinguishable breaths (inspiration and expiration) were captured. These breaths showed no flow from the two differential pressure sensors. A potential source of this error was moisture in the pressure tubing causing pressure sensor failure for some smaller breaths due to moisture blocking the sensor tube. The dataset remains unedited, and these breaths were simply not processed. These breaths comprised less than 20% of the data in the entire data set. 2% of the data in the set was not processed due to human error in recording trial data. A more optimally fitted mask would be expected to reduce the error in fully capturing expiration, by reducing leaks around the mask’s seal with the face.

The inclusion of subjects with significant respiratory abnormalities and/or induced respiratory distress would provide more comprehensive data for the development of clinical metrics, decision-support systems, and/or clinical CPAP protocols. The preliminary patient WOB and ventilator unloading findings [13] establish the value of this data, and hence an extended trial is scheduled to collect information from a larger and more diverse (in age, ethnicity, and respiratory condition) subject pool.

Availability of data and materials

The datasets generated and/or analysed during the current study are available on Physionet [12].

Abbreviations

ARDS:

Acute respiratory distress syndrome

CPAP:

Continuous positive airway pressure

E:

Elastance

P:

Pressure

PEEP:

Positive end expiratory pressure

Q:

Flow

WOB:

Work of breathing

References

  1. Popat B, Jones AT. Invasive and non-invasive mechanical ventilation. Medicine. 2012;40(6):298–304. https://doi.org/10.1016/j.mpmed.2012.03.010.

    Article  PubMed  Google Scholar 

  2. Burns KEA, Adhikari NKJ, Keenan SP, Meade M. Use of non-invasive ventilation to wean critically ill adults off invasive ventilation: meta-analysis and systematic review. BMJ. 2009;338:b1574. https://doi.org/10.1136/bmj.b1574.

    Article  PubMed  PubMed Central  Google Scholar 

  3. Duncan AW, Oh TE, Hillman DR. PEEP and CPAP. Anaesth Intensive Care. 1986;14(3):236–50.

    Article  CAS  Google Scholar 

  4. Rajdev K, Idiculla PS, Sharma S, Von Essen SG, Murphy PJ, Bista S. Recurrent pneumothorax with CPAP therapy for obstructive sleep apnea. Case Rep Pulmonol. 2020;2020:8898621. https://doi.org/10.1155/2020/8898621.

    Article  PubMed  PubMed Central  Google Scholar 

  5. Yiallourou TI, Schmid Daners M, Kurtcuoglu V, et al. Continuous positive airway pressure alters cranial blood flow and cerebrospinal fluid dynamics at the craniovertebral junction. Interdiscip Neurosurg. 2015;2(3):152–9. https://doi.org/10.1016/j.inat.2015.06.004.

    Article  Google Scholar 

  6. Heinzer RC, Stanchina ML, Malhotra A, et al. Lung volume and continuous positive airway pressure requirements in obstructive sleep apnea. Am J Respir Crit Care Med. 2005;172(1):114–7. https://doi.org/10.1164/rccm.200404-552OC.

    Article  PubMed  PubMed Central  Google Scholar 

  7. Mercat A, Richard J-CM, Vielle B, et al. Positive end-expiratory pressure setting in adults with acute lung injury and acute respiratory distress syndrome: a randomized controlled trial. JAMA. 2008;299(6):646–55. https://doi.org/10.1001/jama.299.6.646.

    Article  CAS  PubMed  Google Scholar 

  8. Oba Y, Thameem DM, Zaza T. High levels of PEEP may improve survival in acute respiratory distress syndrome: a meta-analysis. Respir Med. 2009;103(8):1174–81. https://doi.org/10.1016/j.rmed.2009.02.008.

    Article  PubMed  Google Scholar 

  9. FitzGerald C, Hurst S. Implicit bias in healthcare professionals: a systematic review. BMC Med Ethics. 2017;18(1):19. https://doi.org/10.1186/s12910-017-0179-8.

    Article  PubMed  PubMed Central  Google Scholar 

  10. Policy: CPAP Titration Protocol. In: Disorders UoTMBCfS, ed. Galveston and Webster. 2019.

  11. Titration protocol reference guide. In. Koininklijke Philips N.V: Philips Respironics. 2017.

  12. Guy E., Knopp J., Chase G. CPAP pressure and flow data from a local trial of 30 adults at the University of Canterbury (version 1.0.1). Journal. 2022. https://doi.org/10.13026/xfae-vv63.

  13. Guy EF, Chase JG, Knopp JL, Shaw GM. Quantifying ventilator unloading in CPAP ventilation. Comput Biol Med. 2022. https://doi.org/10.1016/j.compbiomed.2022.105225.

    Article  PubMed  Google Scholar 

  14. Holder-Pearson L, Chase G. Physiologic-range flow and pressure sensor for respiratory systems. HardwareX. 2021. https://doi.org/10.1016/j.ohx.2021.e00227.

    Article  PubMed  PubMed Central  Google Scholar 

  15. Szlavecz A, Chiew YS, Redmond D, et al. The Clinical utilisation of respiratory elastance software (cure soft): a bedside software for real-time respiratory mechanics monitoring and mechanical ventilation management. BioMed Eng Online. 2014;13(1):140. https://doi.org/10.1186/1475-925X-13-140.

    Article  PubMed  PubMed Central  Google Scholar 

  16. Sensirion. Datasheet SDP8xx-Analog. Journal. 2018.

  17. NXP. Integrated silicon pressure sensor on-chip signal conditioned, temperature compensated and calibrated. 2008.

Download references

Acknowledgements

Not applicable.

Funding

The data resulted from research funded by the University of Canterbury.

Author information

Authors and Affiliations

Authors

Contributions

EFSG: data collection system design, experimental design, data collection and article drafting. JGC: project supervisor, experimental design, and article drafting. JLK: project supervisor and article drafting. OG: data collection system design and data collection. SB: data collection system design and data collection. LH-P: data collection system design. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Ella F. S. Guy.

Ethics declarations

Ethics approval and consent to participate

Ethical consent for the trial was Granted on 16 JULY 2020 by the Human Ethics Committee at the University of Canterbury (Ref: HEC 2020/14/LR). An amendment on 19 MARCH 2021 confirmed the ability to publish the de-identified dataset.

Consent for publication

Not applicable.

Competing interests

The authors declare that they have no competing interests.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Guy, E.F.S., Knopp, J.L., Gilbertson, O. et al. CPAP pressure and flow data at 2 positive pressure levels and multiple controlled breathing rates from a trial of 30 adults. BMC Res Notes 15, 257 (2022). https://doi.org/10.1186/s13104-022-06133-w

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s13104-022-06133-w

Keywords