Analysis of reciprocating system real-time torque variation during root canal preparation

Abstract

Objective

This study monitored real-time torque variation of the WaveOne Gold (WOG) and Reciproc Blue (RB) during root canal preparation of mandibular molars. Thirty-six mandibular molars were prepared with WOG Primary 25.07 (WOGP, n = 36) and the RB R25 25.08 (RBR25, n = 36) for the mesial canals, whereas WOG Large 45.05 (WOGL, n = 18) and RB R40 40.06 (RBR40, n = 18) for the distal. Canal preparation was divided into thirds and the torque, maximum torque and time, were recorded.

Results

The RBR25 instruments exhibited higher maximum torque in the apical third in contrast to the WOGP instruments (p < 0.05). The intragroup analysis found a significant difference in maximum torque between the cervical and apical thirds, and the middle and apical thirds (p < 0.05) for both instruments (RBR25 and WOGP). The WOGP group had the shortest preparation time (p < 0.05). There were no significant differences between the WOGL and RBR40 for any of the parameters evaluated (p > 0.05). The RBR25 had the highest torque when compared to the WOGP. Both instruments exhibited higher torque in the apical third and there were no significant differences between the instruments in the distal canal.

Introduction

In recent years the use of instrumentation systems has gained prominence in endodontics, as they require only a single instrument for root canal mechanical preparation. The WaveOne Gold (WOG) (Dentsply Sirona, Ballaigues, Switzerland) and Reciproc Blue (RB) (VDW, Munich, Germany) files are manufactured with M-Wire alloys and have specific heat treatments that provide greater flexibility and fatigue resistance [1,2,3].

However, instrument fracture during root canal preparation remains a constant concern for endodontists. rotational speed, torque, alloy type, instrument design, instrumentation technique, and operator experience may be directly related to endodontic instrument fractures [4,5,6,7,8].

Electric endodontic motors with torque control, which only release the amount of mechanical energy required for instrument operation, can control these factors [9]. Among the electric models, the X-Smart IQ (Dentsply Sirona, Ballaigues, Switzerland) enables real-time torque monitoring (RTTM), which displays the torque as the instrument is used to prepare the canal; thus, it increases the confidence and safety of professionals during biomechanical preparation, especially when using single-use instruments.

Although these instruments have been widely studied, torque variation during root canal preparations using WOG and RB systems has not been investigated, to our knowledge. Thus, the aim of this study was to compare, ex vivo, the torque of these two reciprocating systems during mandibular molar canal preparations.

Main text

Methods

Selection of teeth

Thirty-six recently-extracted mandibular molars were used, following approval by the Universidade Positivo Institutional Research Ethics Committee (opinion 2,077,489). All donors signed an informed consent form and a donation term. The teeth were radiographed and analyzed under operative microscopy. Radiographically all teeth had similar canal diameter and morphology. The inclusion criteria were: complete rhizogenesis; presence of three independent canals; and the absence of endodontic treatment and caries, resorptions, perforations, or signs of fractures and cracks. Teeth with radiographic evidence of calcifications and aberrant canal morphology were excluded.

Biomechanical preparation of root canals

The biomechanical preparation of root canals was performed by a single operator, an Endodontist with 5 years of experience, aiming to minimize the operator bias.

After coronal access, a #10K-file (Dentsply Sirona, Ballaigues, Switzerland) was used to confirm patency. Once the actual tooth length was determined by visualizing the instrument in the foramen, the working length (WL) was considered 1 mm smaller than this measurement. The canal preparations were performed by thirds to record the torque used in each region.

The mesial canals were prepared using WaveOne Gold Primary (WOGP, Dentsply Sirona, Ballaigues, Switzerland) and Reciproc Blue R25 (RBR25, VDW, Munich, Germany). Thus, the mesial roots were prepared using both systems, alternating their distribution in the mesiobuccal and mesiolingual canals, resulting in 36 samples per group. When necessary, the mesial reference edge was adjusted to ensure the same WL was used for both mesial canals.

The distal canals were prepared using WaveOne Gold Large (WOGL) and Reciproc Blue R40 (RBR40). For this root, the distribution of the systems was alternated between teeth, resulting in 18 samples per group.

The instruments were powered by an X-Smart IQ motor (Dentsply Sirona, Ballaigues, Switzerland), enabling RTTM. The WaveOne instruments were used in the “WaveOne” programming mode and the Reciproc Blue instruments in the “Reciproc Blue” mode. Before the use of each instrument, each system was calibrated using the X-Smart IQ program. One operator performed the instrumentation, and the instruments were discarded after one use.

Irrigation was performed 3 mm below the WL during instrument kinematics with 2.5% sodium hypochlorite using a NaviTip #20 needle (Ultradent, South Jordan, Utah, USA) connected to a 5 mL disposable syringe, followed by a final irrigation with 5 mL of 17% EDTA. The canals were then irrigated with 5 mL of distilled water and dried with absorbent paper tips.

The X-Smart IQ motor enables RTTM. After treatment completion, the motor generates a final report displaying the torque of each movement during modeling. The torque values for each movement, the maximum torque value for each instrument, and the time (s) for each canal preparation were recorded. To eliminate interference or induction of the results, the operator did not have access to the iPad Mini screen (Apple Inc., Cupertino, California, USA) that displayed the real-time torque values.

Statistical analysis

The Kolmogorov–Smirnov test was performed to verify the normality of the sample. An independent analysis of the mesial and distal canals by non-parametric testing was performed using the SPSS® (IBM® SPSS® Inc., Chicago, USA), version 24, with a 5% significance level. The Kruskal–Wallis test was used for intragroup analysis to determine significant differences in torque and maximum torque reached in each third of the canal with the same instrument. The Mann–Whitney U test was used for intergroup analysis to determine significant differences in torque and maximum torque among the instruments used within the same third of the canal. The Mann–Whitney U test was also used to determine significant differences in root canal preparation times between instruments.

Results

The sample was not normally distributed (p < 0.05). The results of the comparisons are described below.

Mesial canal (WOGP vs. RBR25)

The mesial canal results are shown in Table 1.

Table 1 Median values for torque and maximum torque reached during mesial root canal preparation using WOGP and RBR25 files

Intragroup torque analysis by thirds (cervical, middle, apical)

The intragroup analysis by thirds found no significant differences (p > 0.05). Maximum torque was significantly different between regions for both instruments (p < 0.05).

Intergroup torque analysis by thirds (WOGP vs. RBR25)

There were no significant differences in torque between groups (p > 0.05). However, maximum torque was significantly different in the apical third between groups (p < 0.05).

Preparation time analysis (WOGP vs. RBR25)

Preparation time for the WOGP was significantly shorter when compared to that of the RBR25 (p < 0.05).

In the torque analysis, there were no significant intragroup differences for either the WOGP or RBR25 (p > 0.05). There were also no significant intergroup differences for the cervical, middle, and apical thirds (p > 0.05).

In the maximum torque analysis, there were significant intragroup differences for both the WOGP and RBR25 (p < 0.05) groups. There were also significant intergroup differences for the apical third (p < 0.05).

In the time analysis, there was a significant difference between instruments (p < 0.05). Figure 1 shows the RTTM during the mesial canal preparations with WOGP and RBR25.

Fig. 1
figure 1

Torque peaks recorded during mesial root canal preparation. RTTM during preparation with WOGP (a) and RBR25 (b)

Distal canal (WOGL vs. RBR40)

The distal canal results are shown in Table 2.

Table 2 Median values for torque and maximum torque reached during distal root canal preparation using WOGL and RBR40 files

Intragroup torque analysis by thirds (cervical, middle, apical)

There were no significant intragroup differences in torque (p > 0.05) or maximum torque (p > 0.05).

Intergroup torque analysis by thirds (WOGL vs. RBR40)

There were no significant differences between instruments for the middle (p > 0.05) and apical thirds (p > 0.05). The WOL did not generate torque during preparation of the cervical portions; therefore, comparisons for the cervical third could not be made. The RBR40 touched the cervical third in only four samples.

Preparation time analysis (WOGL vs. RBR40)

There were no significant differences in preparation time between instruments (p > 0.05).

No torque or maximum torque was recorded for the cervical third while using the WOGL.

In the torque analysis, there were no significant intragroup differences for either the WOGP (p > 0.05) or RBR40 (p > 0.05) group. There were also no significant intergroup differences for the middle or apical thirds (p > 0.05).

In the maximum torque analysis, there were no significant intragroup differences for either the WOGL or RBR40 (p > 0.05) group. There were also no significant intergroup differences for the middle or apical thirds (p > 0.05).

In the time analysis, there was no significant difference between instruments (p > 0.05).

Discussion

With the development of the X-Smart IQ endodontic motor, torque monitoring for rotating and reciprocating systems has been improved to real time. This allows the clinician to assess variations that occur during root canal preparations and modify the instruments actions.

In vitro methods are typically used for mechanical property analysis of endodontic instruments [3, 10, 11]. However, this ex vivo study better approximates clinical procedures, as it utilizes recently-extracted human. This study compared the torque of two reciprocating M-Wire files, WOG and RB, which were chosen based on similarities in manufacturing, tip size, taper, and kinematics. Although the root canal anatomy may make sample homogenization difficult, the bias of this variable was reduced by using the WOG and RB instruments in the same root and alternating the canals; therefore, both instruments performed similar actions on the same root length and curvature.

The median torque used during canal preparation was not significantly different between the instruments tested, meaning that both instrument systems exhibited similar behavior. However, there was a significant difference between instruments for the torque used during the apical third preparation. The RBR25 had the highest maximum torque during this preparation, which may suggest a risk for instrument deformation or fracture. Laboratory studies have shown that RB have greater resistance to cyclic fatigue when compared to that of WOG [1, 12,13,14]. However, the greater torque variation exhibited by the RBR25 in this study did not interfere with the instruments’ safety, as none deformed or fractured during their first use. This may justify the manufacturers’ indications that both instruments are single-use only.

A clinical study by Bueno et al. [15] found that the fracture indexes of the conventional WaveOne and Reciproc were low, and the instruments did not show any significant differences after reuse; however, the WOP fractured in one case, and the R25 had two cases. The highest torque for RBR25 found here cannot be associated with the RB fractures of the aforementioned clinical study, because they are two different instruments with distinctive characteristics.

Kim et al. [16] and Plotino et al. [17] demonstrated that Reciproc R25 instruments have a lower metallic mass and lower torsional resistance in contrast to WOP. In addition, Silva et al. [18] stated that although blue heat treatment increases the flexibility of this instrument, it lowers the torsional resistance and lowers the torque required for failure. This suggests that the maximum torque reached by the instrument may be related more to torsional resistance than to cyclic fatigue.

Reciproc Blue instruments have an “S” cross-section, while WOG files are parallelogram-shaped. Due to this instrument design, it is expected that WOG have a lower cutting power and, consequently, generate lower debris production [19]. Plotino et al. [20] and Özyürek et al. [21] stated that Reciproc files have a higher cutting efficiency than WaveOne files, possibly due to the differences in cross-sectional design and cutting angle, and the cross-sectional design is a more important indicator of reciprocating instrument cutting power. However, the results of this study indicated that, although the RB has greater cutting efficiency, its maximum torque was higher in the apical third when compared to that of the WOG, which can be explained by the difference in taper and cross-sectional design. The RBR25 file has a larger taper, which may remove more dentin during preparation, but requires greater torque to do so.

When comparing the canal thirds in the same instrument group, the WOGP and RBR25 had similar results. The maximum torque of both instruments was significantly higher in the apical third but was not significantly different when compared between the middle and cervical thirds. This result agrees with the reportedly higher number of instrument fractures that occur in the apical third [22] due to the anatomical configuration of the root canal, which has a smaller diameter and greater curvature in this region [23].

The root canal preparation time for the RBR25 was higher than that of the WOGP. This was possibly due to its greater conicity, which increases the difficulty of instrument progression into the root canal [24]. Because the WOGL and RBR40 files exhibited similar results in all analyses, it can be concluded that both were equal in safety and predictability during the lower molar distal canal instrumentations. These canals had lower maximum torque medians than the mesial canals, which suggests that molar preparation should begin in the mesial canals where increased instrument stress will occur. The WOGL and RBR40 have a higher calibration, and because of the lower torque generated, they are considered more resistant to preparation and can generate a possible sub-instrumentation of the distal canal using instruments with an initial diameter of 40 and 45 [25, 26].

Further studies should be performed to evaluate the degree of consistency of this method with more than one operator or even with unexperienced operator to test the reliability and reproducibility of the results, considering that measurements can be modified by the operator’s digital pressure and this might be related to the operator’s experience.

Conclusion

Both instruments exhibited higher torque in the apical third and there were no significant differences between the instruments in the distal canal.

Limitations

The limitations of this study can be attributed to the root canal geometry before instrumentation, considering that it is not possible to standardize the internal anatomy of the root canal. Furthermore, the use of an operator-driven instrument has the disadvantage of introducing operator bias.

Abbreviations

WOG:

WaveOne Gold

RB:

Reciproc Blue

WOGP:

WaveOne Gold Primary 25.07

RBR25:

Reciproc Blue R25 25.08

WOGL:

WaveOne Gold Large 45.05

RBR40:

Reciproc Blue R40 40.06

RTTM:

real-time torque monitoring

WL:

working length

References

  1. Keskin C, Inan U, Demiral M, Keles A. Cyclic fatigue resistance of Reciproc Blue, Reciproc, and WaveOne Gold reciprocating instruments. J Endod. 2017;43:1360–3.

    Article  Google Scholar 

  2. De-Deus G, Silva EJ, Vieira VT, Belladona FG, Elias CN, Plotino G, Grande NM. Blue thermomechanical treatment optimizes fatigue resistance and flexibility of the Reciproc files. J Endod. 2017;43:462–6.

    Article  Google Scholar 

  3. Özyürek T. Cyclic fatigue resistance of Reciproc, WaveOne, and WaveOne Gold nickel-titanium instruments. J Endod. 2016;42:1536–9.

    Article  Google Scholar 

  4. Elnaghy AM, Elsaka SE. Evaluation of the mechanical behaviour of PathFile and ProGlider pathfinding nickel-titanium rotary instruments. Int Endod J. 2015;48:894–901.

    Article  CAS  Google Scholar 

  5. Martins RC, Seijo MO, Ferreira EF, Paiva SM, Ribeiro Sobrinho AP. Dental students’ perceptions about the endodontic treatments performed using NiTi rotary instruments and hand stainless steel files. Braz Dent J. 2012;23:729–36.

    Article  Google Scholar 

  6. Ferreira F, Adeodato C, Barbosa I, Aboud L, Scelza P, Zaccaro Scelza M. Movement kinematics and cyclic fatigue of NiTi rotary instruments: a systematic review. Int Endod J. 2017;50:143–52.

    Article  CAS  Google Scholar 

  7. Ungerechts C, Bårdsen A, Fristad I. Instrument fracture in root canals—where, why, when and what? A study from a student clinic. Int Endod J. 2014;47:183–91.

    Article  CAS  Google Scholar 

  8. Goldberg M, Dahan S, Machtou P. Centering ability and influence of experience when using WaveOne single-file technique in simulated canals. Int J Dent. 2012;2012:206321.

    Article  Google Scholar 

  9. Yared GM, Bou Dagher FE, Machtou P. Influence of rotational speed, torque and operator’s proficiency on ProFile failures. Int Endod J. 2001;34:47–53.

    Article  CAS  Google Scholar 

  10. Scelza P, Davidowicz H, Silva LE, Barbosa IB, Scelza MZ. A comparison of two reciprocating instruments using bending stress and cyclic fatigue tests. Braz Oral Res. 2015;2:1–7.

    Google Scholar 

  11. Chang SK, Kim YC, Chang H, Jee KK, Zhu Q, Safavi K, Shon WJ, Bae KS, Spangberg LS, Kum KY. Effect of heat treatment on cyclic fatigue resistance, thermal behavior and microstructures of K3 NiTi rotary instruments. Acta Odontol Scand. 2013;71:1656–62.

    Article  CAS  Google Scholar 

  12. Gündoğar M, Özyürek T. Cyclic fatigue resistance of OneShape, HyFlex EDM, WaveOne gold, and Reciproc blue nickel-titanium instruments. J Endod. 2017;43:1192–6.

    Article  Google Scholar 

  13. Özyürek T, Gündoğar M, Uslu G, Yilmaz K, Staffoli S, Grande NM, Plotino G, Polimeni A. Cyclic fatigue resistances of Hyflex EDM, WaveOne gold, Reciproc blue and 2shape NiTi rotary files in different artificial canals. Odontology. 2018;106:408–13.

    Article  Google Scholar 

  14. Topçuoğlu HS, Demirbuga S, Düzgün S, Topçuoğlu G. Cyclic fatigue resistance of new reciprocating files (Reciproc blue, WaveOne gold, and SmartTrack) in two different curved canals. J Investig Clin Dent. 2018;9:e12344.

    Article  Google Scholar 

  15. Bueno CSP, Oliveira DP, Pelegrine RA, Fontana CE, Rocha DGP, Bueno CEDS. Fracture incidence of WaveOne and Reciproc files during root canal preparation of up to 3 posterior teeth: a prospective clinical study. J Endod. 2017;43:705–8.

    Article  Google Scholar 

  16. Kim HC, Kwak SW, Cheung GS, Ko DH, Chung SM, Lee W. Cyclic fatigue and torsional resistance of two new nickel-titanium instruments used in reciprocation motion: Reciproc versus WaveOne. J Endod. 2012;38:541–4.

    Article  Google Scholar 

  17. Plotino G, Grande NM, Testarelli L, Gambarini G. Cyclic fatigue of Reciproc and WaveOne reciprocating instruments. Int Endod J. 2012;45:614–8.

    Article  CAS  Google Scholar 

  18. Silva EJNL, Hecksher F, Antunes HDS, De-Deus G, Elias CN, Vieira VTL. Torsional fatigue resistance of blue-treated reciprocating instruments. J Endod. 2018;44:1038–41.

    Article  Google Scholar 

  19. Xavier F, Nevares G, Romeiro MK, Gonçalves K, Gominho L, Albuquerque D. Apical extrusion of debris from root canals using reciprocating files associated with two irrigation systems. Int Endod J. 2015;48:661–5.

    Article  CAS  Google Scholar 

  20. Plotino G, Giansiracusa Rubini A, Grande NM, Gamabrini G. Cutting efficiency of Reciproc and WaveOne reciprocating instruments. J Endod. 2014;40:1228–30.

    Article  Google Scholar 

  21. Özyürek T, Yılmaz K, Uslu G. Shaping ability of Reciproc, WaveOne gold, and HyFlex EDM single-file systems in simulated S-shaped canals. J Endod. 2017;43:805–9.

    Article  Google Scholar 

  22. Tzanetakis GN, Kontakiotis EG, Maurikou DV, Marzelou MP. Prevalence and management of instrument fracture in the postgraduate endodontic program at the Dental School of Athens: a five-year retrospective clinical study. J Endod. 2008;34:675–8.

    Article  Google Scholar 

  23. Iqbal MK, Kohli MR, Kim JS. A retrospective clinical study of incidence of root canal instrument separation in an endodontics graduate programme: a PennEndo database study. J Endod. 2006;32:1048–52.

    Article  Google Scholar 

  24. Webber J. Shaping canals with confidence: WaveOne Gold single-file reciprocating system. Roots. 2015;1:34–40.

    Google Scholar 

  25. Weiger R, Bartha T, Kalwitzki M, Löst C. A clinical method to determine the optimal apical preparation size. Part I. Oral Surg Oral Med Oral Pathol Oral Radiol Endodontol. 2006;102:686–91.

    Article  Google Scholar 

  26. Markvart M, Darvann TA, Larsen P, Dalstra M, Kreiborg S, Bjorndal L. Micro-CT analyses of apical enlargement and molar root canal complexity. Int Endod J. 2012;45:273–81.

    Article  CAS  Google Scholar 

Download references

Authors’ contributions

FB designed the study, revised the manuscript critically for important intellectual content. JVF contributed to acquisition of data and interpretation. FSFT designed the study and helped to draft the manuscript. MCLG helped to draft the manuscript and performed the data analysis. JFM helped to draft the manuscript and performed the literature review. MDS designed the study and revised the manuscript critically for important intellectual content. All authors read and approved the final manuscript.

Acknowledgements

None.

Competing interests

The authors declare that they have no competing interests.

Availability of data and materials

All the data supporting the findings is contained within the manuscript.

Consent for publication

Not applicable.

Ethics approval and consent to participate

The study protocol has been approved by the Research Ethics Committee of the Universidade Positivo (opinion 2,077,489). All donors signed an informed consent form and a donation term.

Funding

This study was supported by grants from the Conselho Nacional de Desenvolvimento Científico e Tecnológico (National Council for Scientific and Technological Development-CNPq) under Grant 118599/2017-8. The role of the funding was to was to supply the acquisition of the necessary materials for the research.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Flares Baratto-Filho.

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Baratto-Filho, F., de Freitas, J.V., Gabardo, M.C.L. et al. Analysis of reciprocating system real-time torque variation during root canal preparation. BMC Res Notes 12, 47 (2019). https://doi.org/10.1186/s13104-019-4080-z

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s13104-019-4080-z

Keywords