Dried blood spot is the feasible matrix for detection of some but not all hepatitis B virus markers of infection

Abstract

Objective

Use of dried blood spots (DBS) for detection of hepatitis B virus (HBV) markers of infection has the potential to facilitate diagnosis of HBV infection especially in resource-limited countries. The aim of this study was to evaluate the feasibility of DBS for detection of various markers of HBV infections.

Results

Fifty-four DBS samples were engineered from well-characterized plasma samples. All DBS samples were tested for HBsAg, total anti-HBc and HBV DNA, 20 of 54 samples were also tested for HBeAg using commercially available assays. HBsAg was detected in 24 of 25 (96%), HBV DNA in 22 of 25 (88%), total anti-HBc in all 9 (100%), and HBeAg in all 7 (100%) DBS samples. The average difference in HBV DNA levels between DBS eluates and corresponding plasma samples was 2.7 log10 IU/mL. Fifteen DBS eluates positive for HBV DNA were sequenced and all of them belonged to HBV genotype A. Thirteen samples which were negative for all HBV markers showed HBeAg false positivity. Therefore, DBS is a reliable sample matrix for detection of HBsAg, total anti-HBc and HBV DNA, but not HBeAg. Further feasibility studies of DBS for diagnostic purposes and epidemiologic studies are warranted.

Introduction

The World Health Organization reported that, in 2015, approximately 257 million people were still living with hepatitis B virus (HBV) infection [1]. HBV causes both acute and chronic infections with potential to induce malignant transformation to hepatocellular carcinoma. Hepatitis B ranked 15th among all causes of human mortality [2, 3]. HBV infection is prevalent in African (6.1%) and Western Pacific (6.2%) populations [1]. The diagnosis for HBV infection is based on detection of various markers which include hepatitis B surface antigen (HBsAg), total hepatitis B core antibody (total anti-HBc), hepatitis B e antigen (HBeAg), and HBV DNA. The sample matrices for these markers are generally serum and plasma samples, which require phlebotomy and subsequent pre-analytic processing of blood samples. Phlebotomy, in addition to being more invasive, labor intensive and incurs higher costs in cold chain during transportation of serum or plasma samples. Dried blood spots (DBS), on the other hand, provide an attractive alternative that obviate these requirements of phlebotomy, cold chain conditions and additional associated costs.

Obtaining DBS is much less invasive than drawing blood, since it only needs finger-pricking for sampling. Additionally, it doesn’t require subsequent centrifugation, refrigeration and large space for shipment and storage, which contribute to high costs. Thus, these advantages of DBS expedite large-scale serologic studies. DBS has been successfully used for detection of various markers of neonatal, congenital, metabolic disorders and infectious diseases [4, 5].

The potential use of DBS, as an alternative sample material, for the detection of HBV markers of infection has been explored [6,7,8,9,10]. The aim of this study was to further evaluate the feasibility of using DBS for detection of HBsAg, total anti-HBc, HBeAg, and HBV DNA.

Main text

Materials and methods

Engineered DBS

Positive and negative control DBS samples were engineered by spotting reconstituted blood consisting of 60% of plasma and 40% of red blood cells (RBC) on filter paper, using the protocol as previously described [4] (Additional file 1: Fig. S1). We prepared engineered DBS and tested them after storing for 1–30 days at − 20 °C. Fifty-four well characterized plasma samples previously tested for all hepatitis markers [11] and O + RBC (Tennessee Blood Services, Memphis, TN, US) were used to create the engineered DBS. The serological profiles of the samples are shown in Table 1. For each of the markers, the panel included 9 samples positive for total anti-HBc, 25 samples positive for HBsAg, 25 samples positive for HBV DNA. Twenty-eight samples negative for all HBV markers were included as negative controls. Among 54 plasma samples, 20 samples including 5 samples with high HBV DNA titer (over 6.4 log10 IU/mL), 3 samples with middle (4.0–6.4 log10 IU/mL), 2 samples with low (2.0–4.0 log10 IU/mL) and 10 samples negative for HBV DNA were selected to test for HBeAg. Of these samples, 7 samples were detected to be positive and 13 samples were negative. For quantitative determination of HBV DNA levels in plasma and corresponding DBS eluates, a standard curve was generated by tenfold serial dilutions of AccuSpan™ HBV DNA Linearity Panel (SeraCare, Milford, MA, US) combined with the RBC (titers ranged from 2.7 log10 IU/mL to 7.7 log10 IU/mL) to make the standard curve spots.

Table 1 Serological profile of 54 test and control samples used for evaluation of DBS for detection of HBV markers of infection

DBS elution

Three spots (containing approximately 50 μL of blood per spot) were collected from each sample by punching disks using a hole-puncher to completely excise the entire blood spot. These disks were transferred directly into an Eppendorf tube and 1.2 mL of phosphate buffer saline (PBS) was added to it for elution resulting in an estimated plasma dilution of 3:40. Elution of the DBS was carried out at 56 °C in a shaker at 700 rpm for 1 h. Eluates were collected in new Eppendorf tubes after the disks had turned white, indicating complete elution of samples. DBS eluates were immediately tested for HBsAg, total anti-HBc, HBeAg and HBV DNA.

Detection of HBV markers

The engineered DBS with corresponding plasma samples were tested for HBsAg and total anti-HBc using chemiluminescent microparticle immunoassay (CMIA) in ARCHITECT i2000SR (04P5325 and 06L2225, Abbott Diagnostics, Abbott Park, Chicago, IL, US), following the manufacturer’s instructions.

HBeAg was tested by enzyme-linked immuno sorbent assay (ELISA) using ETI-EBK PLUS (DiaSorin, Saluggia, VC, Italy) and chemiluminescence assay in VITROS ECi Immunodiagnostic System (Ortho-Clinical Diagnostics, Inc., Rochester, NY, US).

For detection of HBV DNA from plasma and DBS eluate samples, total nucleic acids were extracted from 200 μL of samples using the MagNApure Total Nucleic Acid Isolation Kit (Roche Applied Science, Rochester, NY, US) on the MagNApure LC instrument (Roche Applied Science, Indianapolis, IN, US) with a final elution volume of 50 μL. The dual-labeled Taqman HBV probe and HBV primers were designed from an alignment of representative S-gene sequences of all HBV genotypes as previously described [12]. Detection of HBV DNA by quantitative polymerase chain reaction (qPCR) was performed using 10 μL of extracted nucleic acids in a total reaction volume of 25 μL with Express Supermix (Life Technologies, Grand Island, NY, US). The LC480 system (Roche Applied Science, Rochester, NY, US) was used for amplification, detection, and data analysis. Cycle numbers of amplification were increased to 55 cycles to optimize the condition for getting more PCR products, considering the use of DBS. After uracil-DNA glycosylase activation and incorporation, there was a denaturation step at 95 °C for 2 min. Amplification was conducted by 55 cycles of 95 °C for 10 s, 60 °C for 60 s, and with cooling at 40°C for 30 s. A standard curve was created by using the DBS engineered with tenfold serial dilutions of AccuSpan™ HBV DNA Linearity Panel (SeraCare, Milford, MA, US) as described above.

All HBV DNA positive samples were re-amplified for sequencing by a nested Sybr qPCR targeting the S gene, using primers S1F and S1R in the primary reaction and SNF and SNR in the nested reaction (Additional file 2: Table S1). The first-round PCR conditions included incubation at 94 °C for 6 min, followed by 35 cycles of 94 °C for 45 s, 55 °C for 20 s, 72 °C for 60 s, 72 °C for 7 min. Nested PCR was started at 94 °C for 6 min, followed by 30 cycles of 94 °C for 45 s, 55 °C for 30 s, 72 °C for 60 s, and 72 °C for 7 min. Sequencing reactions were conducted using the BigDye v2.1 chemistry sequencing kit (Life Technologies, Grand Island, NY, US), and products were sequenced using the 3130xl Genetic Analyzer (Life Technologies, Grand Island, NY, US). Sequencing was done in both the forward and reverse directions with internal nested primers. Sequencing PCR involved 25 cycles, each cycle consisting of 96 °C for 10 s, 50 °C for 5 s and 60 °C for 2 min [12]. All PCR programs above include a hold step at 4 °C at the end. Sequence analysis was conducted using SeqMan Pro (DNASTAR, Inc., Madison, WI, US) and phylogenetic tree was created by MEGA 7.0.26 [13].

Reproducibility of serological tests and qPCR

For serological tests, reproducibility was confirmed by testing two HBsAg positive (Signal-to-cut off ratio (S/CO) of 2232.1 and 1762.9), two total anti-HBc positive (S/CO of 10.5 and 10.4) and two both negative samples. HBeAg testing was repeated with all 20 samples which were selected for the test as mentioned above. For molecular test, HBV DNA high, medium, low titer and negative samples were selected: 8.2 log10 IU/mL, 6.2 log10 IU/mL, and 4.4 log10 IU/mL. All Samples were tested in triplicate by the same operator for three days using the same process.

Results

For engineered DBS, sensitivity, specificity and percent agreement between DBS and corresponding plasma were determined by comparing the results of HBsAg, total anti-HBc, HBeAg and HBV DNA in DBS eluates to the results of their corresponding plasma samples (sensitivity (%) = number of positive samples in engineered DBS/number of positive samples in plasma, specificity (%) = number of negative samples in engineered DBS/number of negative samples in plasma, percent agreement (%) = number of samples which concordantly represent positive or negative results in plasma and engineered DBS/number of plasma samples). The sensitivity, specificity and percent agreement of HBsAg detection in DBS was 96.0%, 100% and 98.1%, respectively. One sample that showed discordant result between DBS and corresponding plasma had a lower (6.4) S/CO in corresponding plasma sample (Table 1). Total anti-HBc was detected in all 9 DBS eluates indicating a 100% sensitivity, specificity and percent agreement with corresponding plasma samples (Table 1). HBeAg was detected in all 7 DBS eluates whose corresponding plasma samples were also HBeAg positive, however, HBeAg was also detected in all of 13 samples whose corresponding plasma samples were HBeAg negative (Table 1).

HBV DNA detection showed 88.0% of sensitivity, 100% of specificity and 94.4% of percent agreement (Table 1). The three DBS samples from which HBV DNA was undetectable had 2.47, 2.61 and 2.85 log10 IU/mL HBV DNA (average: 2.64 log10 IU/mL) in their corresponding plasma samples (Table 2). The limit of detection (LOD) of HBV DNA titer in engineered DBS was 2.0 log10 IU/mL and the average of HBV DNA titer difference between DBS and corresponding plasma was 2.7 log10 IU/mL. From engineered DBS and corresponding plasma, HBV DNA was sequenced in 15 of 22 (68.0%) samples. Sequences were concordant between those of DBS and corresponding plasma samples, and all were determined to be genotype A (Fig. 1).

Table 2 Result of HBsAg, total anti-HBc, HBeAg, HBV DNA detection in engineered Dried Blood Spots (DBS) and corresponding plasma samples
Fig. 1
figure 1

Phylogenetic tree of engineered dried blood spots (DBS, hollow bullets) and corresponding plasma (colored bullets); sequences without a bullet represent reference. Hepatitis B virus (HBV) DNA sequences of 15 samples are all concordant between DBS and corresponding plasma and are determined to be genotype A

Reproducibility was confirmed to be 100% in HBsAg, total anti-HBc, HBeAg and HBV DNA testing with DBS.

Discussion

In this study we evaluated the use of DBS samples for detection of HBsAg, HBV DNA, total anti-HBc and HBeAg using well characterized plasma samples and their corresponding engineered DBS samples. Three markers except HBeAg were detected in DBS eluates with 100% concordance for total anti-HBc, 98% for HBsAg and 94% for HBV DNA despite a much smaller sample volume and its dilution during the elution process from DBS. The sensitivity of total anti-HBc in DBS in our study far exceeds the lower sensitivities reported in several previous studies (range 68.0% to 90.5%, average 81.2%) [14,15,16,17,18], although the sample input volume and dilution factors were similar in our study and these studies [16]. It is possible that the protocol for elution of samples from DBS used in our study results in a better yield of the sample and using freshly eluted samples for testing further contributes to a better sensitivity. Detection of HBsAg from DBS also demonstrated excellent performance, which were similar to the report of latest systematic review (98% of sensitivity and 100% of specificity) [7, 9].

The sensitivity of HBV DNA was lower than that of HBsAg and total anti-HBc (88%). This might be explained by loss of HBV DNA titer due to small volume of DBS samples and potential degradation of DNA through multiple steps of DBS processing, such as elution or extraction procedures. The LOD of HBV DNA was 2.0 log10 IU/mL and the average of HBV DNA titer difference between DBS and corresponding plasma was 2.7 log10 IU/mL in this study. According to the latest systematic review on HBV DNA quantification using DBS, the lowest limit of detection was around 3.0 log10 IU/mL [8, 9], which suggests inevitable HBV DNA titer loss in DBS with the current methods. Despite the limitation with the detection of HBV DNA in DBS, as a screening tool for large population studies, the majority of patients currently infected by HBV will be able to be identified. However, unreliable results were observed in HBeAg detection in our study. To date, two studies examined the utility of HBeAg testing with DBS [19, 20]. Mohamed et al. got consistent results of 10 HBeAg positives in both plasma and DBS by immunoassay (No testing platform detail was provided) among 100 HBsAg positive samples [20]. Jackson et al. could detect HBeAg positivity with 80.9% of sensitivity (17/21 samples) by chemiluminescence technology using LIAISON XL (Diasorin, Saluggia, Italy) among 98 HBsAg positive samples [19]. False positivity was not observed in these studies. According to our results, however, HBeAg negative samples concordantly showed false positive in DBS when these were tested by ELISA using ETI-EBK PLUS (DiaSorin) and chemiluminescence technology using VITROS ECi Immunodiagnostic System (Ortho-Clinical Diagnostics, Inc.). The factors responsible for this false positivity for HBeAg remain to be determined.

Use of DBS samples also allows sequencing of viral genome and subsequent determination of viral genotype. Ability to generate sequences from HBV DNA positive DBS eluates is an additional advantage of using DBS.

Conclusion

DBS can be a reliable sample matrix for detecting markers of HBV infection in large population studies, especially in resource-limited countries. Three key markers of HBV infection, HBsAg, total anti-HBc and HBV DNA, were detected in DBS with high accuracy in our study. Further studies are needed for elucidating and evaluating the feasibility of DBS for HBeAg testing and optimization of elution protocols to improve the sensitivity of HBV DNA detection.

Limitations

Our study has some limitations. First, we used only engineered DBS samples. Paired DBS and plasma samples from clinical studies were not available for further validation of DBS. Second, only two platforms, ETI-EBK PLUS and VITROS ECi Immunodiagnostic System, were used for HBeAg testing. Further trials with other platforms are essential to elucidate why DBS samples showed false positivity.

Availability of data and materials

The datasets used and/or analysed during the current study are available from the corresponding author on reasonable request.

Abbreviations

DBS:

Dried blood spot

HBeAg:

Hepatitis B e antigen

HBsAg:

Hepatitis B surface antigens

HBV:

Hepatitis B virus

Total anti-HBc:

Total hepatitis B core antibody

References

  1. WHO. Global Hepatitis Report. 2017. https://apps.who.int/iris/bitstream/handle/10665/255016/9789241565455-eng.pdf;jsessionid=C5E67B8EAF0874328A41092AD68F2F2A?sequence=1. Accessed 17 Aug 2021.

  2. Lozano R, Naghavi M, Foreman K, Lim S, Shibuya K, Aboyans V, et al. Global and regional mortality from 235 causes of death for 20 age groups in 1990 and 2010: a systematic analysis for the Global Burden of Disease Study 2010. Lancet. 2012;380(9859):2095–128.

    Article  Google Scholar 

  3. MacLachlan JH, Cowie BC. Hepatitis B virus epidemiology. Cold Spring Harb Perspect Med. 2015;5(5): a021410.

    Article  Google Scholar 

  4. Tejada-Strop A, Drobeniuc J, Mixson-Hayden T, Forbi JC, Le NT, Li L, et al. Disparate detection outcomes for anti-HCV IgG and HCV RNA in dried blood spots. J Virol Methods. 2015;212:66–70.

    Article  CAS  Google Scholar 

  5. Tuaillon E, Kania D, Pisoni A, Bollore K, Taieb F, Ontsira Ngoyi EN, et al. Dried Blood Spot Tests for the Diagnosis and Therapeutic Monitoring of HIV and Viral Hepatitis B and C. Front Microbiol. 2020;11:373.

    Article  Google Scholar 

  6. Flores GL, Barbosa JR, Cruz HM, Miguel JC, Potsch DV, Pilotto JH, et al. Dried blood spot sampling as an alternative for the improvement of hepatitis B and C diagnosis in key populations. World J Hepatol. 2021;13(4):504–14.

    Article  Google Scholar 

  7. Lange B, Cohn J, Roberts T, Camp J, Chauffour J, Gummadi N, et al. Diagnostic accuracy of serological diagnosis of hepatitis C and B using dried blood spot samples (DBS): two systematic reviews and meta-analyses. BMC Infect Dis. 2017;17(Suppl 1):700.

    Article  Google Scholar 

  8. Lange B, Roberts T, Cohn J, Greenman J, Camp J, Ishizaki A, et al. Diagnostic accuracy of detection and quantification of HBV-DNA and HCV-RNA using dried blood spot (DBS) samples - a systematic review and meta-analysis. BMC Infect Dis. 2017;17(Suppl 1):693.

    Article  Google Scholar 

  9. Villar LM, Bezerra CS, Cruz HM, Portilho MM, Flores GL. Applicability of Oral Fluid and Dried Blood Spot for Hepatitis B Virus Diagnosis. Can J Gastroenterol Hepatol. 2019;2019:5672795.

    Article  Google Scholar 

  10. Yamamoto C, Nagashima S, Isomura M, Ko K, Chuon C, Akita T, et al. Evaluation of the efficiency of dried blood spot-based measurement of hepatitis B and hepatitis C virus seromarkers. Sci Rep. 2020;10(1):3857.

    Article  CAS  Google Scholar 

  11. Mixson-Hayden T, Dawson GJ, Teshale E, Le T, Cheng K, Drobeniuc J, et al. Performance of ARCHITECT HCV core antigen test with specimens from US plasma donors and injecting drug users. J Clin Virol. 2015;66:15–8.

    Article  Google Scholar 

  12. Mixson-Hayden T, Lee D, Ganova-Raeva L, Drobeniuc J, Stauffer WM, Teshale E, et al. Hepatitis B virus and hepatitis C virus infections in United States-bound refugees from Asia and Africa. Am J Trop Med Hyg. 2014;90(6):1014–20.

    Article  CAS  Google Scholar 

  13. Kumar S, Stecher G, Tamura K. MEGA7: Molecular Evolutionary Genetics Analysis Version 70 for Bigger Datasets. Mol Biol Evol. 2016;33(7):1870–4.

    Article  CAS  Google Scholar 

  14. Flores GL, Cruz HM, Potsch DV, May SB, Brandao-Mello CE, Pires MMA, et al. Evaluation of HBsAg and anti-HBc assays in saliva and dried blood spot samples according HIV status. J Virol Methods. 2017;247:32–7.

    Article  CAS  Google Scholar 

  15. Gruner N, Stambouli O, Ross RS. Dried blood spots–preparing and processing for use in immunoassays and in molecular techniques. J Vis Exp. 2015;97: e52619.

    Google Scholar 

  16. Mossner BK, Staugaard B, Jensen J, Lillevang ST, Christensen PB, Holm DK. Dried blood spots, valid screening for viral hepatitis and human immunodeficiency virus in real-life. World J Gastroenterol. 2016;22(33):7604–12.

    Article  Google Scholar 

  17. Tappin DM, Greer K, Cameron S, Kennedy R, Brown AJ, Girdwood RW. Maternal antibody to hepatitis B core antigen detected in dried neonatal blood spot samples. Epidemiol Infect. 1998;121(2):387–90.

    Article  CAS  Google Scholar 

  18. Villar LM, de Oliveira JC, Cruz HM, Yoshida CF, Lampe E, Lewis-Ximenez LL. Assessment of dried blood spot samples as a simple method for detection of hepatitis B virus markers. J Med Virol. 2011;83(9):1522–9.

    Article  CAS  Google Scholar 

  19. Jackson K, Holgate T, Tekoaua R, Nicholson S, Littlejohn M, Locarnini S. Evaluation of dried blood spots for hepatitis B and D serology and nucleic acid testing. J Med Virol. 2019;94(2):642–8.

    Article  Google Scholar 

  20. Mohamed S, Raimondo A, Penaranda G, Camus C, Ouzan D, Ravet S, et al. Dried blood spot sampling for hepatitis B virus serology and molecular testing. PLoS ONE. 2013;8(4): e61077.

    Article  CAS  Google Scholar 

Download references

Acknowledgements

We would like to thank Drs. Jan Drobeniuc, Lixia Li, Joanne Mei and Nicholas Wiese at the Centers for Disease Control and Prevention for providing additional peer review of the manuscript, and their technical assistance.

The findings and conclusions in this report are those of the authors and do not necessarily represent the official position of the Centers for Disease Control and Prevention.

Funding

This research did not receive any specific grant from funding agencies in the public, commercial, or not-for-profit sectors.

Author information

Authors and Affiliations

Authors

Contributions

MK, ATS and SK conceptualized the study. MK, ATS, and PL conducted data curation and formal analysis. MK investigated relevant research papers and data. MK, ATS and TMH designed the methodology. SK administrated the project. MK drafted the original manuscript under supervision of SK and MS. ATS, SK and MS reviewed and edited the manuscript. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Motoji Sawabe.

Ethics declarations

Ethics approval and consent to participate

Not applicable. This study was performed only with engineered DBS. This DBS was created by well characterized plasma samples previously tested for all hepatitis markers [11] and O + RBC (Tennessee Blood Services, Memphis, TN, US), and their corresponding plasma.

Consent for publication

All authors of this manuscript have read and approved the final submitted version and are aware that they are listed as an author on this paper.

Competing interests

The authors declare no conflict of interest.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Supplementary Information

Additional file 1: Figure S1.

Protocol of engineering dried blood spots (DBS) [4]. Use 850 mL of whole blood with O+ red blood cells (RBC) and wash the RBC with 50 mL of 0.9% saline to remove plasma, buffy coat, and anticoagulant. Centrifuge the mixture of whole blood and 0.9% saline at 4,000 rpm in 21 °C for 8 min and discard the supernatant. Repeat the wash procedure three times, and centrifuge for 15 min at the last time. Then confirm the hematocrit of RBC is at least 95%. To reconstitute the whole blood (40% of RBC and 60% of plasma) using the washed RBC, 1 mL of RBC and 1.36 mL of well characterized plasma samples are mixed thoroughly. The reconstituted whole blood is spotted onto filter paper at 50 µL blood per spot.

Additional file 2: Table S1.

Primers and probes of hepatitis B virus (HBV) DNA test and sequencing.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Kikuchi, M., Lindstrom, P., Tejada-Strop, A. et al. Dried blood spot is the feasible matrix for detection of some but not all hepatitis B virus markers of infection. BMC Res Notes 15, 287 (2022). https://doi.org/10.1186/s13104-022-06178-x

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s13104-022-06178-x

Keywords