Simulated-use validation of a sponge ATP method for determining the adequacy of manual cleaning of endoscope channels

Abstract

Background

The objective of this study was to validate the relative light unit (RLU) cut-off of adequate cleaning of flexible colonoscopes for an ATP (adenosine tri-phosphate) test kit that used a sponge channel collection method.

Methods

This was a simulated-use study. The instrument channel segment of a flexible colonoscope was soiled with ATS (artificial test soil) containing approximately 8 Log10 Enterococcus faecalis and Pseudomonas aeruginosa/mL. Full cleaning, partial cleaning and no cleaning were evaluated for ATP, protein and bacterial residuals. Channel samples were collected using a sponge device to assess residual RLUs. Parallel colonoscopes inoculated and cleaned in the same manner were sampled using the flush method to quantitatively assess protein and bacterial residuals. The protein and viable count benchmarks for adequate cleaning were <6.4 ug/cm2 and <4 Log10 cfu/cm2.

Results

The negative controls for the instrument channel, over the course of the study remained low with on average 14 RLUs, 0.04 ug/cm2 protein and 0.025 Log10 cfu/cm2. Partial cleaning resulted in an average of 6601 RLUs, 3.99 ug/cm2, 5.25 Log10 cfu/cm2 E. faecalis and 4.48 Log10 cfu/cm2 P. aeruginosa. After full cleaning, the average RLU was 29 (range 7–71 RLUs) and the average protein, E. faecalis and P. aeruginosa residuals were 0.23 ug/cm2, 0.79 and 1.61 Log10 cfu/cm2, respectively.

Conclusions

The validated cut-off for acceptable manual cleaning was set at ≤100 RLUs for the sponge collected channel ATP test kit.

Background

Cleaning of flexible endoscopes is still predominately a manual process that is fraught with errors [14]. As reviewed by Ofstead et al. [5], human factors play a significant role in the efficacy of the manual cleaning of flexible endoscopes and only 1.4 % of all flexible GI endoscopes reprocessed that they evaluated had all the steps properly performed. The recent outbreaks of carbapenem-resistant enterobacteriaceae (CRE) associated with improperly reprocessed flexible duodenoscopes [1, 6] has led to questions regarding how endoscopy clinics can ensure their endoscope reprocessing is adequate. Culture of fully reprocessed ERCP duodenoscopes has been suggested as a means of detection of specific pathogens such as CRE [6], however, this would require quarantine of the endoscopes until the culture results are back (usually 48 h). Culture to determine if ERCP endoscopes harbor CRE is an excellent initial assessment. However, there needs to be a quality system approach to endoscope reprocessing to ensure on an ongoing basis that the basic reprocessing steps are being performed properly.

Recently, the development of rapid audit tools (e.g. detection of residual organic residues [79], or adenosine tri-phosphate (ATP) residues [1014]) has allowed users to assess the efficacy of their manual cleaning such that improperly cleaned flexible endoscope channels can be re-cleaned prior to going to the high level disinfection (HLD) step. This process of rapidly auditing the cleaning and repeating it, if needed, ensures that an inadequately cleaned endoscope will not be used on a subsequent patient. This reduces the risk of improperly reprocessed flexible endoscopes being responsible for transmission of infectious diseases. Testing for residual ATP is one of the most widely published methods to rapidly audit cleaning compliance of flexible endoscopes [8, 10, 11, 1315].

There have been a number of publications related to using ATP to monitor channel cleaning where the sample methods consists of a flush or flush-brush-flush of the channel [911, 1315]. Since the sensitivity of different manufacturer’s ATP test kits varies [12], it is crucial that each manufacturer validate the “cut-off” level of ATP for adequate cleaning for their specific test kit. There have been reports validating the cut-off for some ATP test kits [13], but there are no published reports of validation of an ATP test kit that uses a “sponge” sample collection method to assess endoscope channel cleaning.

The objective of this study was to assess the efficacy of sample collection using a moistened sponge passed through the colonoscope instrument channel and to determine the appropriate RLU cut-off for this test kit that correlates with adequate cleaning as defined by protein and bacterial markers.

Methods

No research or ethics approval was needed, as this was a totally in vitro study.

Flexible endoscopes

Two Fujinon colonoscopes Model EC-530HL (Fujinon, Saitama City, Japan) were used for this study (colonoscope #1 and colonoscope #2). These colonoscopes were provided by Ruhof for the purposes of this study and were returned to Ruhof at the completion of the study. Colonoscope #1 and colonoscope #2 were alternated during the testing so they were used equally throughout the experimental testing protocol. Only instrument channel testing was performed (i.e. soiling and sample collection were only done for the instrument channel segment). After each experiment all channels and all surfaces of the colonoscope were cleaned and then the colonoscope was sterilized as per the endoscope manufacturer’s instructions. The STERIS System 1 (STERIS Corp., Mentor, OH) was used for sterilization of the colonoscopes (peracetic acid was the sterilant). Post-sterilization, all colonoscopes channels were flushed with 70 % ethanol and forced air was used to dry all channels prior to storage.

Colonoscope cleaning methods tested

Full-cleaning

Leak testing was performed using a LT-7 hand-held leak tester (Fujinon) while the colonoscope was fully immersed. Full cleaning was performed as per the colonoscope manufacturer’s instructions using Ruhof Endozime® Bio-Clean enzymatic detergent (Ruhof, Mineola, New York, USA) at a use-dilution of 6 mL/L and a total contact time of 2 min at room temperature. While fully immersed, the colonoscope was wiped with a lint-free cloth, the outlets of the instrument channel port and the air–water port were brushed with short brushes (Ruhof) and all channels were brushed three times with a double-ended channel cleaning brush (Ruhof). Subsequently, all channels were flushed with detergent manually using the CA-510 cleaning adaptor for G5 series Fujinon colonoscopes. A total of 90 mLs of detergent was flushed through each channel. The colonoscope was then transferred and immersed in a basin of tap water, the exterior wiped with a lint-free cloth and a total of 90 mLs of water was flushed through each channel using the CA-510 adaptor.

Partial cleaning

The partial clean consisted of a 50 mL flush of sterile reverse osmosis (sRO) water through the instrument channel segment (no brushing of this channel).

Microorganisms

Enterococcus faecalis ATCC 29212 and Pseudomonas aeruginosa ATCC 27853 were stored at −70 °C as stock cultures and were sub-cultured three times on Tryptic Soy agar containing 5 % (v/v) sterile sheep blood (BA) prior to being used in experimental protocols. This consecutive 3 day sub-culturing protocol was used to ensure the bacteria were actively replicating prior to being used in the experimental protocol.

Test soil

For soiling the colonoscope channel, the organic challenge consisted of ATS (artificial test soil) containing ~108 cfu/mL of Enterococcus faecalis ATCC 29212 and Pseudomonas aeruginosa ATCC 27853. The ATS was freshly prepared and does contain ATP (derived from serum and blood in the test soil), blood, serum and thickening agents (US Patent No. 6447990). The ATS used for this study contained 1,390,000 RLUs/mL (average of five replicates).

Colonoscope inoculation

The ATS containing E. faecalis and P. aeruginosa was drawn up into a 30cc syringe that was attached to the instrument channel port on the colonoscope. The test soil was flushed through the instrument channel segment and excess soil flushed out with air. The inoculated channel was allowed to dry passively for 1 h. After drying, a 30cc syringe containing 30 mL sRO water was attached to the instrument channel port and flushed through the instrument channel segment (similar to a bedside flush used during patient-procedures). This pre-rinse was done to reduce the channel soiling to ensure the RLU level did not exceed 9999 since the Ruhof luminometer has a maximum RLU reading of 9999 (i.e. any test sample with RLUs ≥9999 will all give the same reading).

Sample harvesting of endoscope channel

The ATP test kit manufacturer’s instructions indicated that once the sponge has been passed through the endoscope channel, the entire sponge should be cut off and placed into the ATP test device. This test method would not allow quantitation of protein or viable count from that sponge sample. As such we developed a sponge elution method so that ATP, protein quantitation and viable count of the sponge sample could be determined. The three sample methods included: Sponge sample (as described by the manufacturer of this ATP test kit), sponge elution sample (done for research purposes only) and the fluid flush sample (done for research purposes to determine the protein and viable count in the colonoscope channel).

Sponge sample (ATP test only)

The Ruhof ATP test swab (ATP Complete), the Ruhof Test Instrusponge™ Channel Testing Sponge and the Ruhof ATP Complete hand-held chemiluminometer (Ruhof) were used for this study. Collection of the sample from a colonoscope channel consisted of passing the Instrusponge once through the instrument channel segment. The sponge was then cut off and placed into the ATP Test tube, which was activated and then placed into the hand-held ATP luminometer, and the RLUs determined as per the manufacturer’s instructions.

Sponge elution sample (ATP, protein, viable count testing)

The ATP test kit evaluated in this study uses a sponge to collect channel samples. Instead of bristles on the channel brush there is an adsorbent sponge that fits into the channel snuggly thereby creating friction against the inner channel surface. We wanted to determine at what RLU cutoff does the ATP test have protein and viable counts that meet the benchmarks for adequate cleaning. Since the sponge sample method was intended only to be used for ATP determination we evaluated this elution method to determine if ATP, protein and viable count could be reliably determined from the sponge eluate (i.e. could all the test parameters we wanted to compare be done on one sponge sample?). The sponge provided in the ATP test kit was moistened with sRO water and passed through the instrument channel segment. Once the sponge emerged from the distal tip, it was aseptically cut off and placed in 20 mL of sRO water. The sample was eluted by mixing well using a vortex mixer. The eluted fluid was used for viable count, ATP and protein determinations. For ATP measurement of the sponge elution sample, the sponge in the eluted sample was aseptically removed and excess fluid allowed to drip off the sponge, which was then transferred into the ATP test tube and then tested for ATP, as described for the ATP test sponge sample.

Flush sample (ATP, protein viable count testing)

A 30cc sterile syringe containing 20 mL sRO water was attached to the instrument channel port and the fluid flushed through the instrument channel segment and collected in a sterile container. This fluid was used for viable count, ATP and protein determinations. For the ATP measurement of the flush sample, the sponge was inserted into the sample and allowed to absorb fluid (each sponge absorbs 0.041 mLs of fluid). The ATP sponge was then cut off and transferred into the ATP test tube and then tested for ATP as described for the ATP test sponge sample.

Assay test methods

For each testing protocol there were between 3–5 replicates performed (i.e. N = 3 or N = 5 where samples were collected from sequential testing using colonoscope #1 and colonoscope #2). Table 1 summarizes the testing protocol for each type of endoscope sample collected.

Table 1 Tests performed on each colonoscope channel sample type

Viable bacterial count (for flush or sponge elution fluid samples)

Quantitation of viable bacteria was performed using standard serial 1:10 dilutions where 0.1 mLs of each dilution was spread over the surface of a BBL™CHROMagar™ Orientation media plate (Becton Dickenson, Orville, ON, Canada). The limit of detection for the viable count assay was 10 cfu/mL.

Protein assay (for flush or sponge elution fluid samples)

The samples collected were assayed for protein using the QuantiPro BCA assay kit, which includes an internal bovine serum albumin protein standard, and is a quantitative assay based on bicinchoninic acid (Sigma, St Louis, Missouri). The protein assay was performed as per the manufacturer’s instructions and had a limit of detection of 0.5 μg/mL.

Benchmarks for adequate manual cleaning

The manual cleaning benchmarks for flexible endoscope channels that were established by Alfa et al. [7] and supported by Pineau et al. [8] were used. If manual cleaning of the endoscope channel has been adequate then there should be ≤6.4 µg/cm2 of protein and ≤4 Log10 cfu/cm2 of bioburden.

Results

The data in Table 2 show that the protein residuals collected on the sponge are not well eluted as the μg/cm2 for protein collected by the sponge was much lower than that collected using a channel flush method. Also the residual ATP detected in the channel by the direct sponge method was higher than the RLU levels detected by the flush method. Based on these findings, the remainder of the simulated-use testing focused on using parallel soiled colonoscopes where the protein and viable count residuals post-treatment were determined using the flush collection method and the ATP was determined using the direct sponge collection method on a second endoscope channel inoculated at the same time.

Table 2 Comparison of residual ATP detected in a soiled colonoscope instrument channel evaluated using various sample collection methods

To validate the appropriate RLU cutoff of this sponge ATP channel collection method, parallel soiling of the colonoscope instrument channel segment was used (Table 3). The negative controls from fully reprocessed colonoscope instrument channels (i.e. post-cleaning and sterilization) ranged from 1 to 39 RLUs over the course of the study. This indicated that even with repeated ATS soiling, cleaning and sterilization that the level of RLUs in the instrument channel returned to similar low values after full reprocessing was performed. This negative control data was captured and presented separately (i.e. not subtracted from the positive, partial clean or full-clean data) to ensure the data for “full cleaning” was not affected by repeated round of soiling and testing. After full manual cleaning of the colonoscope channel, the ATP level detected by the sponge method was on average 29.0 ± 36.39 RLUs (range 7–71 RLUs).

Table 3 Comparison of residual ATP detected by the direct sponge method versus level of residual protein and viable count in colonoscope instrument channels

Discussion

Most traditional ATP test kits use swabs to sample surfaces and use fluid to collect samples from endoscope channels [7, 9, 10, 1315]. Our report is the first to experimentally validate an appropriate RLU cut-off for the sponge-based ATP test kit. We recommend a cutoff of 100 RLUs for this sponge-based ATP test kit. Although the cut-off could have been set at 71 RLUs, as this was highest RLU found post-cleaning, we believe that there needs to be some margin allowed for variability in the ATP assay. The 100 RLU cut-off ensures that the maximum RLU level detected post cleaning during simulated-use testing was well below this cutoff. Furthermore, this cutoff represented two standard deviations higher than the mean RLU providing a reasonable margin to accommodate the variability in low RLU levels post-cleaning. Table 3 data indicated that a cut-off of ≤100 RLUs for adequate cleaning would ensure that the protein and bioburden levels in this channel would be well below the established cut-off of ≤6.4 µg/cm2 protein and ≤4 Log10 cfu/cm2 bioburden.

The 100 RLU cut-off for this sponge channel-sample method is different from the 200 RLU cut-off established for the flush channel-sample method previously reported by our lab for a different ATP monitoring test kit [13, 14]. This likely reflects differences in both the channel sample method and the hand-held ATP detection luminometer parameters. Differences in sensitivity and luminometer characteristics of different manufacturer’s of ATP detection kits have been clearly outlined by Sciortino et al. [12] and Aiken et al. [10]. Data from the current study and from our previously published report [13] confirm the importance of each manufacturer clearly validating the appropriate “cut-off” for clean that relates to the specific adaptation of their ATP test kit (i.e. the cut-off for “clean” may be different for the same test kit for different applications such as environmental monitoring versus monitoring of manual cleaning for flexible endoscope channels and it may also be different for kits manufactured by different companies).

We did evaluate a similar sponge collection method previously [13] however, the ATP test in that evaluation was done only on the eluted sponge. Our current study data indicated that the direct ATP sponge collection method “concentrates” the residual material collected from the channel (i.e. is not diluted by eluting in fluid) thereby ensuring the highest detectable ATP signal. The only disadvantage is that other markers for cleaning adequacy cannot be assessed on the same sponge that was used for ATP testing.

Our simulated-use negative control data do confirm that for colonoscopes the level of ATP in the instrument channel was returned to a reproducibly low level for all fully reprocessed colonoscopes for the two different manufacturers ATP test kits that we have evaluated. For the flush only elution method used for Alfa et al. [13], the ATP level for the instrument channel of fully reprocessed colonoscopes was 25.5 RLUs (standard deviation of 21.4 RLUs) compared to 14 RLUs (standard deviation of 21.7 RLUs) for the ATP sponge collection method used in the current study.

It should be noted that ATP test methods are excellent for monitoring manual cleaning adequacy for flexible GI endoscopes, however this test cannot replace culture for detection of specific pathogens present in low levels post-HLD. It requires from 102–103 cfu of bacteria per sponge sample to detect 1 RLU (i.e. 104–105 cfu are needed to detect 100 RLU). Furthermore, if ATP was detected post-disinfection/sterilization it would not be possible to differentiate whether the ATP residuals were from bacteria or from other organic residuals such as patient-secretions. Sites that want to assess their ERCP endoscopes for carbapenem-resistant Enterobacteriaceae or other specific types of bacteria will need to perform culture. Further studies are needed to determine whether the ATP test methods could detect biofilm accumulation within endoscope channels.

A limitation of this study was that we did not have the history of the two colonoscopes used for our simulated-use studies so do not know if their usage history was similar. In addition, only the instrument channel segment of the suction channel from colonoscopes was evaluated and no other channels or endoscope types were assessed. Currently there are no sponges that fit into the other smaller channels nor could this method be used for the un-sealed elevator guide-wire channel as the sponge could not fit into this channel. For the instrument channels of gastrointestinal (GI) flexible endoscopes that have similar internal channel diameters (e.g. 3.7–4.0 mm diameter) we would expect the sponge to provide similar sample collection efficacy from the channel surface results regardless of the endoscope type (i.e. colonoscope versus duodenoscope versus gastroscope). In addition, only the one specific type of channel sponge was evaluated. Despite these limitations, the sponge channel collection method provides an easy to perform, rapid and sensitive method for routine monitoring of the cleaning adequacy for flexible GI endoscopes. The frequency of testing the adequacy of manual cleaning (i.e. every endoscope every time versus a portion of endoscopes tested per week) still requires further studies.

In summary, for the instrument channel segment of the suction-biopsy channel, the sponge ATP test method works well and our the simulated-use data confirmed that when the endoscope manufacturer’s cleaning instructions are followed that the ≤100 RLUs cut-off for this ATP test kit should be reliably achieved. Further studies are warranted to determine how well this 100 RLU cleaning cut-off works in a busy endoscopy clinic.

References

  1. Kovaleva J, Peters FT, van der Mei HC, Degener JE. Transmission of infection by flexible gastrointestinal endoscopy and bronchoscopy. Clin Microbiol Rev. 2013;26(2):231–54.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  2. Quality Assurance In Endoscopy Committee ASGE, Petersen BT, Chennat J, et al. Multisociety guideline on reprocessing flexible gastrointestinal endoscopes: 2011. Gastrointest Endosc. 2011;73(6):1075–84.

    Article  Google Scholar 

  3. Beilenhoff U, Neumann CS, Rey JF, et al. ESGE-ESGENA guideline for quality assurance in reprocessing: microbiological surveillance testing in endoscopy. Endoscopy. 2007;39(2):175–81.

    Article  CAS  PubMed  Google Scholar 

  4. AAMI TIR30. Association for the advancement of medical instrumentation. A compendium of processes, materials, test methods and acceptance criteria for cleaning reusable medical devices. AAMI TIR30 Technical Information Report 2011.

  5. Ofstead CL, Wetzler HP, Snyder AK, Horton RA. Endoscope reprocessing methods: a prospective study on the impact of human factors and automation. Gastroenterol Nurs. 2010;33(4):304–11.

    Article  PubMed  Google Scholar 

  6. Epstein L, Hunter JC, Arwady MA, et al. New Delhi metallo-beta-lactamase-producing carbapenem-resistant Escherichia coli associated with exposure to duodenoscopes. JAMA. 2014;312(14):1447–55.

    Article  PubMed  Google Scholar 

  7. Alfa MJ, Olson N, Murray BL. Comparison of clinically relevant benchmarks and channel sampling methods used to assess manual cleaning compliance for flexible gastrointestinal endoscopes. Am J Infect Control. 2014;42(1):e1–5.

    Article  PubMed  Google Scholar 

  8. Pineau L, Philippe E. Evaluation of endoscope cleanliness after reprocessing: a clinical-use study. Zent Steril. 2013;1:22–7.

    Google Scholar 

  9. Visrodia KH, Ofstead CL, Yellin HL, Wetzler HP, Tosh PK, Baron TH. The use of rapid indicators for the detection of organic residues on clinically used gastrointestinal endoscopes with and without visually apparent debris. Infect Control Hosp Epidemiol. 2014;35(8):987–94.

    Article  PubMed  Google Scholar 

  10. Aiken ZA, Wilson M, Pratten J. Evaluation of ATP bioluminescence assays for potential use in a hospital setting. Infect Control Hosp Epidemiol. 2011;32(5):507–9.

    Article  PubMed  Google Scholar 

  11. Obee PC, Griffith CJ, Cooper RA, Cooke RP, Bennion NE, Lewis M. Real-time monitoring in managing the decontamination of flexible gastrointestinal endoscopes. Am J Infect Control. 2005;33(4):202–6.

    Article  PubMed  Google Scholar 

  12. Sciortino CV, Giles RA. Validation and comparison of three adenosine triphosphate luminometers for monitoring hospital surface sanitization: a Rosetta Stone for adenosine triphosphate testing. Am J Infect Control. 2012;40(8):e233–9.

    Article  CAS  PubMed  Google Scholar 

  13. Alfa MJ, Fatima I, Olson N. Validation of adenosine triphosphate to audit manual cleaning of flexible endoscope channels. Am J Infect Control. 2013;41(3):245–8.

    Article  PubMed  Google Scholar 

  14. Alfa MJ, Fatima I, Olson N. The adenosine triphosphate test is a rapid and reliable audit tool to assess manual cleaning adequacy of flexible endoscope channels. Am J Infect Control. 2013;41(3):249–53.

    Article  CAS  PubMed  Google Scholar 

  15. Fushimi R, Takashina M, Yoshikawa H, et al. Comparison of adenosine triphosphate, microbiological load, and residual protein as indicators for assessing the cleanliness of flexible gastrointestinal endoscopes. Am J Infect Control. 2012;41(2):161–4.

    Article  PubMed  Google Scholar 

Download references

Authors’ contributions

MA; conceived of the testing protocol and design, analyzed the data and wrote the manuscript. NO; wrote the experimental testing protocol, tabulated and entered the data into an excel database, and provided critical review of the manuscript. Both authors read and approved the final manuscript.

Acknowledgements

The funds for this study as well as the colonoscopes, ATP test tubes, sponges and luminometer were provided by Ruhof.

Availability of data and materials

The dataset(s) supporting the conclusions of this article are included within the article and its additional file(s).

Competing interests

No potential competing interest for NO. MA was an invited speaker at a Brazil conference and was sponsored by Ruhof. MA acts as a consultant, has received research funding, or acts as an advisory board member for the following: Ruhof, Healthmark Industries, 3M, Olympus, STERIS, J&J, KARL STORZ, Sealed Air, One Life and Ofstead & Associates.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Michelle J. Alfa.

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Alfa, M.J., Olson, N. Simulated-use validation of a sponge ATP method for determining the adequacy of manual cleaning of endoscope channels. BMC Res Notes 9, 258 (2016). https://doi.org/10.1186/s13104-016-2066-7

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s13104-016-2066-7

Keywords